User Manual

38 • Maintenance Model GX-2001 O
p
erator’s Manual
Replacing the Scrubber Filters
1. Verify that the Model GX-2001 is off.
2. Unscrew the two screws that secure the sensor retainer and sensor
cover retainer to the Model GX-2001 and remove them.
3. Using a small at blade screwdriver, gently pry each side of the sensor
cover retainer away from the sensor retainer.
4. Remove the sensor cover to gain access to the scrubbers.
5. The charcoal lter disk is located in the CO sensor diffusion port. One
lter disk is used. Two H2S lter disks are also used in the CO sensor
diffusion port. They are installed over the charcoal lter disk.
Two H2S removal lter disks are also located in the combustible gas
diffusion port.
6. When the appropriate scrubber lters have been replaced, reinstall the
sensor cover.
7. Reattach the sensor cover retainer in its original position. It should snap
into place onto the sensor retainer.
8. Replace the two screws that secure the sensor retainer and sensor
cover retainer to the Model GX-2001, which you removed in step 2
above.
H2SRemoval Filter
Disk
02
RKI
POWER
H2 S
MODE DISP AIR
CO
H2S Removal Filter Disk
Sensor Cover
HCGX-2001
Sensor Cover Retainer
Charcoal Filter Disk
Figure 9: Removing the Sensor Retainer and Replacing the Scrubber Filters