5−4−2.External output action
4∼20mA output version
(1)Signal transmission method : Electric current transmission (not isolated)
(2)Transmission path : CVVS
(3)Transmission distance : Below 1Km
(4)Max loop load : 300Ω at 24VDC
(5)Status signal levels
Detection mode : 4∼20mA(depends on gas concentration)
Gas alarm : 4∼20mA(depends on gas concentration)
Initial clear : 2.5mA(Fix)
Maintenance mode : 2.5mA(Fix)
Alarm test : 4∼20mA(depends on gas concentration)
Trouble alarm : 0.5mA(Fix)
Point skip : 2.5mA(Fix)
(6)Power interruption : 0mA
Relations between gas concentration and output are shown below;
CAUTION
420mA output has been adjusted. Even if re-adjustment is required after
installation, do not operate without our permission.
Please contact with our nearest agent or RIKEN KEIKI if it is required.
2.5mA
4m
20m
22m
Full scale
Gas concentration
Output
0
6% of full scale
(
Zero su
pp
ression level
)
Detection mode
Maintenance mode