User Manual

6 • Description Model GD-K11D Operator’s Manual
Status lights
The display board includes ve status lights. Table 3 lists each light and its function.
Program buttons
The display board includes ve buttons. The MODE,
(PEAK),
, and SET (AL.P)
buttons are directly below the display screen. The TEST button is the red button near the
bottom right of the display board. Table 4 lists each button and its function.
Table 3: Model GD-K11D Status Lights
Light Status
PW/TR
(power)
live power is connected to GD-K11D and the
power switch is ON (solid light)
fail condition (ashing light)
AL
(alarm)
gas alarm condition (solid light)
over range condition (solid light)
ZERO GD-K11D is prepared for adjustment of zero
setting (ashing light)
GD-K11D accepted zero setting (solid light)
SPAN GD-K11D is prepared for adjustment of span
setting (ashing light)
GD-K11D accepted span setting (solid light)
SKIP GD-11D is in maintenance mode (ashing light)
Table 4: Model GD-K11D Program Buttons
Button Function
MODE Enters maintenance mode
(PEAK)
Increases/changes the setting on the display
Displays the peak reading
Decreases/changes the setting on the display
SET (AL.P) Prepares a setting to be changed
Enters a new setting
Displays the alarm point
TEST Enters alarm test mode