Manual

8 Maintenance 8-2. Gas calibration method
GD-70D - 66 -
<Span Adjustment "2-2">
This is used to perform the span adjustment. For the oxygen deficiency alarm specification (OSU - 0 - 25
vol%), this is the same as "1-1".
Span Adjustment Display
PW A1 A2 F LCD
2-2.SPAN
Press the SET key.
2-2
SPAN
MAINTENANCE
Gas Introduction
Introduce a gas, and then press the SET
key when the reading is stabilized.
* In case of less than 10%FS, this is not
switched to the next menu.
2.0ppm
SPAN GAS
MAINTENANCE
8.0ppm
SPAN VAL
MAINTENANCE
Span Adjustment
Because the reading will be fixed, select
the introduced gas concentration by
pressing the or key. Select the
number, and then press the SET key.
(20.9 vol% in the oxygen deficiency alarm
specification)
Span Adjustment Completed
8.0ppm
SPAN OK
MAINTENANCE
Adjustment result record
To record the adjustment result, press the
SET key.
(Press the MODE key to cancel this menu.)
8.0ppm
SPAN Y/N
MAINTENANCE
Adjustment Completed
The menu returns to 2-2.SPAN
automatically.
8.0ppm
SPAN END
MAINTENANCE
Return to 2-2.SPAN
* When the span adjustment fails
8.0ppm
SPAN NG
MAINTENANCE
Return to 2-2.SPAN
CAUTION
Do not press the
SET key before the
readin
g
is stabilized.
CAUTION
Exhaust gas must be collected
in the exhaust bag or discharged
through the exhaust line.