Manual

3 How to Operate 3-2. How to start the detector
- 31 - GD-70D
<Start-up Procedures (approximately 25 seconds for system
check of the detector and alarm deactivation)>
Power On
PW A1 A2 F
LCD
Initial Clear
──────
WARM UP
Gas
Specifications
Display
15.0ppm
SIH4
Detection Mode
0.0ppm
SIH4
WARNING
When Oxygen (OSU) is selected, approximately AIR
(normal oxygen concentration) (20.9 vol%) is output after
the detector is started in the atmosphere.
For a specification where an alarm is triggered by the AIR
(normal oxygen concentration) output (e.g., 0 - 5 vol%),
be careful of an alarm issue after the detector is started.
CAUTION
Do not turn off the detector during the initial clear. The detector is reading the sensor memory during
the initial clear.
If a new sensor unit is installed or the sensor unit is replaced after the detector is started, the sensor
unit must be warmed up for a specified period which is determined depending on the type of the
sensor unit.
When the semiconductor type sensor unit (SGU) is selected, the suggested warm-up completion
time is displayed in the maintenance mode.
During the warm-up, the alarm activation and output signals are unstable. Provide a prior notification
to the related sections so that they can prepare for false abnormalities.
Because the pyrolyzer unit (PLU-70) must be warmed up for one hour, please warm it up with the
detector.
After the warm-up is completed, verify that the reading on the flow rate indicator corresponds to the
specified flow rate, and then perform a gas calibration.
PW:POWER
A1:ALM1
A2:ALM2
F:FAULT
:Lamp on
:Lamp off