88 Calibration Mode Gas Tracer Operator’s Manual
NOTE: Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.
10.When AUTO CAL in the BUMP--SET menu item in Maintenance
Mode is set to ON:
If the ppm combustible sensor passes the bump test,
The following screen appears:
To view the bump test gas reading press the AIR or
(SHIFT) button. You can scroll between the bump test results
and the bump test gas reading with the AIR or (SHIFT)
button.
To return to the BUMP screen in PPM CAL, press POWER
ENTER.
Disconnect the humidifier from the probe.
NOTE: Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.
P
ppm
MAINTENANCE
CH4
BUMP
1000
ppm
MAINTENANCE
CH4
BUMP
MAINTENANCE
BUMP