Gas Tracer Operator’s Manual Calibration Mode 69
1. While in the calibration menu, use the AIR or (SHIFT) button
to navigate to the ONE CAL menu option.
2. Press and release the POWER ENTER button to display the single
calibration menu. The %LEL sensor will be displayed first.
3. Use the AIR or (SHIFT) button to display the sensor you want
to calibrate (in this example the combustible gas % LEL sensor). If
you have a 5 sensor unit, the %vol sensor will be one of the choices.
It is shown below.
4. Press and release the POWER ENTER button. The single
calibration screen displays for the sensor you selected. The gas
reading flashes.