210 Bump Testing Gas Tracer Operator’s Manual
To return to the BUMP screen in Calibration Mode, press
POWER ENTER at any time. If any channel failed the
calibration, the buzzer will sound a double pulsing tone until
you press and release the POWER ENTER button.
Disconnect the tubing from the probe.
10.With the BUMP menu option displayed, press the (SHIFT) button
until the NORMAL menu option is displayed.
11.Press and release the POWER ENTER button to return to
Measuring Mode.
49
12.0
5
%LEL
%
ppm
MAINTENANCE
CH4
O2
/CAL
CO
MAINTENANCE
BUMP