Manual
Gas Tracer Operator’s Manual Using Maintenance Mode • 151
1. Use the AIR▲ or (SHIFT)▼ buttons to display the ALM SLNC
menu item.
2. Press and release the POWER ENTER button. The screen will
indicate the current setting.
3. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting
and return to the main menu.
Updating the Data Log Interval Setting
This setting indicates how often the Gas Tracer saves readings to the
data logger. The following interval times can be selected: 6 minutes, 5
minutes, 3 minutes, 1 minute, 30 seconds, 20 seconds, or 10 seconds.
The factory setting is 5 minutes (300 seconds).
1. Use the AIR▲ or (SHIFT)▼ buttons to display the LG INTVL
menu item.
MAINTENANCE
ALM SLNC
On
MAINTENANCE
ALM SLNC
MAINTENANCE
LG INTVL