Owner's manual

39 Calibration Mode 10,000 ppm Gas Tracer Operator’s Manual
5. Hold down the AIR button. The unit will prompt you to continue
holding the AIR button.
6. When the unit prompts you to release the AIR button, release it.
The instrument will perform a fresh air adjustment on the ppm
sensor and return you to the AIR CAL menu item.
Performing a Dry Fresh Air Adjustment
This section describes how to perform a fresh air adjustment on the
PPM combustible gas sensor using the DRY setting. Use the DRY
setting with a zero air calibration cylinder for a biannual calibration.
NOTE: It is recommended that you perform a fresh air adjust on the
ppm sensor using a zero air cylinder and the DRY setting
followed by a fresh air adjustment using the WET setting once
every 6 months to maintain the accuracy of the ppm
combustible gas reading. See “Performing a Wet Fresh Air
Adjustment” on page 37 for instructions to perform a WET
fresh air adjustment.
1. After entering Cal Mode, the AIR CAL menu item will be
displayed.
AIr
CAL
MAINTENANCE
HOLD AIR
MAINTENANCE
AIR CAL