10,000 ppm Gas Tracer Operator’s Manual Calibrating and Bump Testing the PPM Com-
instrument is not in a fresh air environment, connect a gas bag with
zero air to the rigid tube on the probe using the 6 inch humidifier
tube and open the clamp allowing the instrument to draw zero air
for 2 minutes.
6. Hold down the AIR button. The unit will prompt you to continue
holding the AIR button.
7. When the unit prompts you to release the AIR button, release it.
The instrument will perform a fresh air adjustment on the ppm
sensor and return you to the AIR CAL menu item. If a gas bag with
zero air was used, disconnect the humidifier from the probe.
NOTE: Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.
Performing a Dry Fresh Air Adjustment
This section describes how to perform a fresh air adjustment on the
PPM combustible gas sensor using the DRY setting. Use the DRY
setting with a gas bag filled with zero air for a biannual calibration.
NOTE: It is recommended that you perform a fresh air adjust on the
ppm sensor using a gas bag filled with zero air and the DRY
setting followed by a fresh air adjustment using the WET
setting once every 6 months to maintain the accuracy of the
ppm combustible gas reading. See “Performing a Wet Fresh
Air Adjustment” on page 100 for instructions to perform a
WET fresh air adjustment.
AIr
CAL
MAINTENANCE
HOLD AIR