Owner's manual
3 • Specifications 10,000 ppm Gas Tracer Operator’s Manual
Specifications
Table 1: 10,000 ppm Gas Tracer Specifications
Target Gas Combustible Gas; calibrated to methane
Range
(Increment)
0-9990 ppm
(10 ppm)
NOTE: Detection range is 0-9990 ppm, but displayed range can be
0-1000, 0-2000, 0-4000 or 0-9990 ppm.
Sampling Method Sample Draw
Response Time T90 Within 30 Seconds
Accuracy ± 30 ppm or ± 5% of reading (whichever is greater)
Display Digital LCD Display
Gas Alarms There are no gas alarms in the 10,000 ppm Gas Tracer.
Operating
Temperature &
Humidity
-20°C to 50
°
C/Below 85% RH (Without Condensation)
Safety/
Regulatory
C US
pending
CSA classified, “C/US”, as Intrinsically Safe. Exia. Class I, Groups A,
B, C, & D. Temperature Code T3C.
Power Supply • Three AA size alkaline batteries standard
• Lithium Ion Battery Pack (3.7 VDC Nominal) optional, Direct
Charging
Continuous
Operating Hours
@ 25 °C
• Alkaline Batteries: 15 Hours (Non Alarm Operation, Fully Charged)
• Lithium Ion Battery Pack: 10 Hours (Non alarm Operation, Fully
Charged)
Case High-impact Plastic with protective rubber layer, RF Shielded, Dust
and Weather Proof
Standard
Accessories
• 10 foot hose
• 10 inch probe
• Flexible 4 inch tapered rubber nozzle
Optional
Accessories
• Rechargeable Lithium-Ion Battery Pack
• 115 VAC Charging Station
• 12 VDC Charging Station
• Data Logger Management Program Software
• IrDA/USB Cable for Downloading Data to a Computer With Data
Logger Management Program (not needed if computer has an
infrared port)
• SDM-2102 Automatic Calibration Station & Software
• Various Probes (see “Parts List” on page 95)