User Manual
Diffusion Ea
g
le Instruction Manual Maintenance • 56
6. Remove the sensor.
7. Install the replacement sensor in reverse order. When plugging in the wire
connector into the circuit board, make sure to line up the tabs on the end
of the wire connector with the notches on the connector of the circuit
board. Be very careful not to bend the pins.
NOTE: Verify that you installed the H
2
S or CO sensor in the correct
position. If the sensors are not installed in the correct position, the
top of the Remote Detector will not close correctly when it is
replaced.