User Manual

Diffusion Ea
g
le Instruction Manual Setup Mode 35
Updating the full-scale setting
1. Press the SHIFT/ button to place the prompt in the second line, then
press the POWER/ENTER button. The full-scale setting ashes.
The maximum full-scale setting for a toxic gas channel is 1000 PPM; the
minimum setting is 1.00 PPM. The default setting is 10.0 PPM.
2. Press the AIR/
and SHIFT/ buttons to display the desired full-scale
setting, then press the POWER/ENTER button to enter the setting. The
prompt ashes.
NOTE: The display increment setting automatically updates its default
setting as you change the full-scale setting.
Updating the display increment setting
1. Press the SHIFT/ button to place the prompt in the third line, then press
the POWER/ENTER button. The display increment setting ashes.
The minimum display increment setting is 0.1 PPM; the maximum
display increment setting is 2.5 PPM.
2. Press the AIR/
and SHIFT/ buttons to display the desired display
increment setting, then press the POWER/ENTER button to enter the
setting. The prompt ashes.
Returning to the main menu
1. Press the SHIFT/ button. The ESCAPE message displays. (Press the
AIR/
button to return to the previous screen.)
2. Press the POWER/ENTER button to save the new settings. The OTHER
GAS SET message displays, then the Gas Combinations menu displays.
3. To exit the Gas Combinations menu, press the SHIFT/
button until the
prompt is next to Channel 4, then press the SHIFT/
button again. The
ESCAPE message displays.
4. Press the POWER/ENTER button. The message SAVING DATA
displays, then the main menu displays.