30 • Viewing, Printing, Exporting, and Deleting Data in the Data Window
After specifying the file name, file type, and file location click 
the Save button to save the file to the specified location.
 12.To go back and view other data, click the Return button in 
the upper right corner of the Data Window or the Data 
control button on the right side of the program window and 
select the data you want to view.
Event Data
The Eagle 2 not only saves trend files which include logged data 
at scheduled times, but also saves the 100 most recent alarm 
events and the 100 most recent trouble events. When an 
instrument is downloaded, the Eagle 2 Data Logger 
Management Program will retrieve these events from an 
instrument and save them in alarm event files and trouble event 
files for each instrument that is downloaded.
Alarm event files save the time, instrument channel, gas, and 
alarm type of every gas alarm event that occurs on a particular 
instrument. Warning (low alarm), Alarm (high alarm), STEL, 
TWA, and overscale events are saved.
Trouble event files note the time, instrument channel, whether 
the event is an instrument system failure or sensor failure and 
the specific type of failure. Calibration failures, dead battery 
alarms, and sensor failures are among the trouble events that 
are saved.