Manual

Viewing Data in the Instrument Information Screen • 23
Gas - Lists each channel’s gas in the order it appears on
the Eagle 2 screen.
Calib. Date - Shows the date and time of the most recent
successful calibration for each channel.
Before — shows the gas response prior to calibration.
After — shows the gas response after calibration.
A. Cal. — lists the auto-calibration setting for each channel
of the Eagle 2. If a Eagle 2 passes its calibration, the
“After” column should match the “A. Cal.” column. If the
Eagle 2 fails calibration on any of its channels, those
channels will retain the previous calibration information.
NOTE: If a unit is calibrated using Single Calibration in the
Eagle 2’s Calibration Mode (see the Eagle 2 Operator’s
Manual) it is possible for the “After” reading to be
different from the “A.Cal” setting if the unit was set to a
level different than the “A.Cal” setting.
Cal. Due (Days) - Shows how many days remain before
calibration is required for each gas.
The Warning and Alarm Point Frame is displayed in the lower
half of the window.
Gas - Lists each channel’s gas in the order it appears on
the Eagle 2 screen.
Warning - Shows the warning setpoint.
Alarm - Shows the alarm setpoint.
STEL - Shows the STEL (short term exposure limit)
setpoint for applicable gases.
TWA - Shows the TWA (time-weighted average) setpoint
for applicable gases.
All values in the Warning and Alarm Point Frame can be
changed using the Set window.