96 • Detail Settings Button
from Table 5 on page 102 or through testing as described in
“Obtaining a Relative Response Factor” on page 101. A valid
character for the Factor is a value between 0.01 and 25.00.
There are limitations for the full scale and increment values that
depend on the factor for both the high range and the low range.
Table 2 and Table 3 below list these limitations.
Table 2: High Range PID
Factor
Full Scale
(ppm)
Increment
0.25-0.36 1000 1
0.37-0.49 1500 1
0.50-0.61 2000 1
0.62-0.74 2500 1
0.75-0.99 3000 2
1.00-1.24 4000 2
1.25-2.49 5000 2
2.50-3.74 10000 10
3.75-4.99 15000 10
5.00-7.49 20000 10
7.50-9.99 30000 20
10.00-
12.49
40000 20
12.50 50000 2500
Table 3: Low Range PID
Factor
Full Scale
(ppm)
Increment
0.20-0.29 10.00 0.01
0.30-0.39 15.00 0.01
0.40-0.49 20.00 0.01