Owner manual

228 • Infrared CO
2
Calibration EAGLE 2 Operator’s Manual
5. The Calibration Mode Screen displays with the cursor next to AUTO
CALIBRATION.
6. Move the cursor to the PERFORM AIR ADJUST menu item by using
the RANGE SHIFT button.
7. Remove the black caps from the ends of the carbon dioxide scrubber. Be
sure to grab the scrubber by the EAGLE 2 fitting so that the fitting and
tubing do not come off.
8. Attach the CO
2
scrubber to the inlet fitting of the EAGLE 2 and allow the
instrument to draw gas for 1 minute before performing an air adjust.
NOTE: Attaching the CO
2
scrubber to the inlet fitting eliminates the
background CO
2
found in fresh air and allows the EAGLE 2 to
obtain an accurate zero reading.
9. Press and release the POWER ENTER RESET button. The following
screen appears.
10. Press and release the AIR YES button to continue.
If you do not want to continue, press the DISPLAY ADJUST NO button
and the unit will return to the Calibration Mode Screen.
CALIBRATION MODE
> AUTO CALIBRATION
SINGLE CALIBRATION
PERFORM AIR ADJUST
NORMAL OPERATION
CALIBRATION MODE
AUTO CALIBRATION
SINGLE CALIBRATION
> PERFORM AIR ADJUST
NORMAL OPERATION
PERFORM
AIR ADJUST?