202 • Parts List EAGLE 2 Operator’s Manual
the ESM-01 chamber.
21. Install the two sensor bracket screws.
22. Confirm that the main PCB is seated in its slots and that its bottom edge
is resting on the bottom of the bottom case. If the main PCB is not seated
properly, it may be damaged when the top case is re-installed.
23. Make sure that the top case gasket is fully seated in its groove and
carefully put the top case back on the bottom case. If you have any
difficulty mating the top and bottom cases, inspect the placement of the
main PCB and the placement of the top case gasket.
24. Turn the EAGLE 2 upside down and tighten the three case screws to
secure the top case to the bottom case.
25. Calibrate the ESM-01 channel as described in “ESM-01 Calibration” on
page 181.
Parts List
Table 19: ESM-01 EAGLE 2 Parts List
Part Number Description
33-7120RK H
2
S scrubber disk, for ESM-01DH-D-HCN sensor
33-7121RK H
2
S scrubber disk, for ESM-01DH-F-SO2 sensor
47-5015RK ESM-01 sensor cable
ESM-01DH-ASH3 Arsine sensor, 0-1.50 ppm
ESM-01R-NH3 Ammonia sensor, 0-75.0 ppm
ESM-01DH-D-HCN Hydrogen cyanide sensor, 0-15.0 ppm
ESM-01DH-PH3 Phosphine sensor, 0-1.00 ppm
ESM-01DH-F-SO2 Sulfur dioxide sensor, 0-6.00 ppm
ESM-K01-CL2 Chlorine sensor, 0-3.00 ppm
81-0142RK-02 Calibration cylinder, 5-gas (SO
2
, CH
4
, O
2
, H
2
S, CO), 58-liter aluminum
81-0142RK-04 Calibration cylinder, 5-gas (SO
2
, CH
4
, O
2
, H
2
S, CO), 34-liter aluminum
81-0170RK-02 Calibration cylinder, 5 ppm SO
2
in nitrogen, 58 liter
81-0170RK-04 Calibration cylinder, 5 ppm SO
2
in nitrogen, 34 liter
81-0175RK-02 Calibration cylinder, 10 ppm NH
3
in nitrogen, 58 liter
81-0175RK-04 Calibration cylinder, 10 ppm NH
3
in nitrogen, 34 liter