142 • PID Relative Response Feature EAGLE 2 Operator’s Manual
Table 15: High Range Relative Response Gas List
The last choice in each list, PID, can be defined by the user and loaded in the
EAGLE 2 using the Eagle 2 Maintenance Data Loader Program. See the
Eagle 2 Maintenance Data Loader Program Operator’s Manual.
3. Use the AIR ▲ YES or RANGE
▼ SHIFT button to move the cursor next
to the desired gas.
4. Press and release POWER ENTER RESET. The PID channel will be
configured to the selected gas and the EAGLE 2 will proceed to the
STEL Screen. This configuration will be in force until either a different
gas is selected in Display Mode or the unit is turned off.
PID 0-50.00 OFF OFF OFF OFF
Target Gas Detection
Range (ppm)
Alarm 1 Factory
Setting (ppm)
Alarm 2 Factory
Setting (ppm)
STEL
(ppm)
TWA
(ppm)
Acetone 0-1000 500 750 750 500
Benzene 0-1000 50 250 OFF OFF
Diesel Fuel
NO 1
0-1500 200 300 OFF 200
Ethanol 0-15000 1000 1500 OFF 1000
Gasoline 0-2000 300 500 500 300
Isobutylene 0-2000 400 600 60 42
Isopropanol 0-5000 200 400 400 200
JP-5 Fuel 0-1000 140 210 OFF 14
Methyl Ethyl
Ketone
0-1500 200 300 300 200
Toluene 0-1000 50 150 150 50
N-Hexane 0-5000 500 1000 1000 500
Propylene 0-2500 500 750 OFF 500
Styrene 0-500 20 40 40 20
Tetrachloro-
ethylene
0-1000 25 100 100 25
Trichloro-
ethylene
0-1000 50 100 100 50
Vinyl Chloride 0-4000 100 500 OFF OFF
PID 0-2000 OFF OFF OFF OFF
Table 14: Low Range Relative Response Gas List