EAGLE 2 Operator’s Manual Part Number: 71-0154RK Revision: S Released: 7/31/14 www.rkiinstruments.
WARNING Read and understand this instruction manual before operating instrument. Improper use of the gas monitor could result in bodily harm or death. Periodic calibration and maintenance of the gas monitor is essential for proper operation and correct readings. Please calibrate and maintain this instrument regularly! Frequency of calibration depends upon the type of use you have and the sensor types.
Table of Contents Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . About the EAGLE 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring Mode, Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Monitoring an Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using Optional Sample Hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using Exhaust Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Combustible Gas Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Monitoring Combustible Gas in the PPM or %Volume Ranges . .
Chapter 5: Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Replacing or Recharging the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . Replacing the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Updating the Alarm Point Settings . . . . . . . . . . . . . . . . . . . . . . . . . Updating the Alarm Latching Setting . . . . . . . . . . . . . . . . . . . . . . . Updating the Alarm Silence Setting . . . . . . . . . . . . . . . . . . . . . . . . Turning the User/Station ID Function On or Off . . . . . . . . . . . . . . Updating the Autocal Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Updating the Backlight Delay Setting . . . . . . . . . . . . . . . . . . . . . .
PID Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Start Up and Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PID Relative Response Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PID Sensor Relative Response Screen in Display Mode . . . . . . . . . PID Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TC Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Replacing the TC Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Configuring the TC Gas in Setup Mode . . . . . . . . . . . . . . . . . . . . . . . . . . Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Infrared Hydrocarbon Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Infrared Sub PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Start Up and Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 %LEL/2.0-30.0 %vol Autoranging . . . . . . . . . . . . . . . . . . . . Catalytic (LEL) Sensor Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Infrared Hydrocarbon Calibration . . . . . . . . . . . . . . .
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 289 291 291 292 Appendix N: Using the EAGLE 2 in Inert Mode. . . .
Chapter 1: Introduction Overview This chapter briefly describes the EAGLE 2 gas monitor. This chapter also describes the EAGLE 2 Operator’s Manual (this document). Table 1 at the end of this chapter lists the specifications for the EAGLE 2.
• Data logging functions (when used in Normal Mode) • Alarm trend data (when used in Normal Mode) • STEL and TWA (when used in Normal Mode) and over range alarms • Peak readings (when used in Normal Mode) • Built-in time function • Lunch break feature • RF shielded high impact plastic case • CSA classified for Class I, Division I, Groups A, B, C, and D hazardous atmospheres WARNING: The Model EAGLE 2 detects oxygen deficiency, elevated levels of oxygen, combustible gases, carbon monoxide, and hydrogen sulfi
Specifications Table 1: Standard Sensor Specifications Combustible Gas, Methane (CH4) Calibration Standard Oxygen (O2) Hydrogen Sulfide (H2S) Carbon Monoxide (CO) Detection Range 0 - 100 %LEL 0 - 40 volume% 0 - 100.0 ppm 0 - 500 ppm Reading Increment 1 %LEL 0.1 volume % 0.5 ppm 1 ppm Alarm 1 Factory Setting 10 %LEL* 19.5 volume %* 5.0 ppm* 25 ppm* Alarm 2 Factory Setting 50 %LEL 23.5 volume % 30.0 ppm 50 ppm STEL Alarm n/a n/a 5.0 ppm 200 ppm TWA Alarm n/a n/a 1.
Safety/ Regulatory 186718 CSA classified as Intrinsically Safe. Exia. Class I, Groups A, B, C, & D. Temperature Code T3C.
• Chapter 3 describes the operation of the EAGLE 2. • Chapter 4 describes Calibration Mode which allows you to calibrate the EAGLE 2’s active channels. • Chapter 5 describes the EAGLE 2’s maintenance requirements and procedures. • Appendix A describes calibration of the EAGLE 2 using a sample bag. • Appendix B describes Setup Mode which allows you to configure different parameters of the EAGLE 2.
Chapter 2: Description Overview This chapter describes the EAGLE 2 instrument and accessories. Instrument Description The EAGLE 2 includes the case, sensors, LCD, control buttons, printed circuit boards, alarm LEDs, infrared communication port, buzzer, battery case and batteries, and flow system.
Top Case Handle Strap Case Screw, 3X (Hidden From view) Bottom Case Case Foot, 6X Figure 2: Component Location, Side View Buzzer Inlet Fitting Exhaust Fitting Charging Jack Battery Case Thumbscrew Infrared Port Figure 3: Component Location, Front & Back EAGLE 2 Operator’s Manual Instrument Description • 7
Case The EAGLE 2’s sturdy, high-impact plastic case is radio frequency (RF) resistant and is suitable for use in many environmental conditions, indoors and out. The case is dust proof and water resistant. It’s two main components, the top case and bottom case, are held together with three screws located on the bottom case. The interface between the top case and bottom case is gasketed. A sturdy, well balanced handle on the top case allows you to hold the instrument comfortably.
Oxygen Sensor The O2 sensor is a galvanic type of sensor. A membrane behind the openings on the sensor face allows gas to diffuse into the sensor at a rate proportional to the partial pressure of oxygen. The oxygen reacts in the sensor and produces a voltage proportional to the concentration of oxygen. The voltage is measured by the EAGLE 2’s circuitry, converted to a measurement of gas concentration, and displayed on the LCD.
Control Buttons Four control buttons are located below the LCD. They are, from left to right, POWER ENTER RESET, DISPLAY ADJUST NO, AIR ▲ YES, and RANGE ▼ SHIFT.
Printed Circuit Boards (PCBs) Display PCB Main PCB I.S. Barrier PCB (not visble) Figure 4: EAGLE 2 PCBs The EAGLE 2’s PCBs analyze, record, control, store, and display the information collected. The main PCB and I.S. barrier PCB are located in the bottom case. The display PCB is located in the top case. The display PCB and I.S. barrier PCB are not user serviceable and are not involved in any user performed maintenance.
H2S Connector (H2S) Display connector CO Connector (CO) LEL Connector (HC) I.S. Barrier Connector Oxygen Connector Pins (OXY) DP Switch Connector Pump Connector Float Probe Connector IR Communication Sensor/Transmitter Figure 5: Main PCB The main PCB is located on the right side of the bottom case. It slides into guiding grooves at the bottom, front, and rear of the bottom case. It is held in place by the top case.
infrared port or an IrDA/USB cable connected to a USB port can be used to download data saved by the EAGLE 2 to a computer using the Eagle 2 Data Logger Management Program. See the Data Logger Management Program operator’s manual for data logging and downloading instructions. Buzzer A solid-state electronic buzzer is located on the front of the top case. It is a panel mounting type of buzzer and is water resistant and sealed to the inside of the top case with an O-ring.
Flow System The EAGLE 2 flow system consists of the inlet fitting, hydrophobic filter, pump, internal tubing, differential pressure (DP) switch, sensor chamber, charcoal filter, and exhaust fitting. Inlet Fitting The inlet fitting is on the right front (when viewed from the front) of the bottom case. It is a nickel plated brass quick connect fitting. It mates with either the sample hose or with the hydrophobic probe.
the O-ring compression force. Charcoal Filter The charcoal filter is located in the front of the flow chamber next to the CO sensor. It contains activated charcoal. The CO sensor will respond if exposed to H2S and certain hydrocarbon gases. The charcoal filter scrubs these gases out of the sample to avoid false CO readings. If false or elevated CO readings are noticed, especially in the presence of H2S, change the charcoal filter.
CAUTION: Sample hose lengths of more than 125 feet are not recommended for the EAGLE 2 because of flow rate reduction and increased response time. Consult RKI Instruments, Inc. for hose lengths longer than 125 feet. The probe includes a replaceable particle filter and hydrophobic filter disk that prevent particulates and water from entering the EAGLE 2’s flow system.
Rechargeable Ni-MH Batteries Rechargeable Ni-MH batteries are available for the EAGLE 2. A fully charged set of Ni-MH batteries will power the EAGLE 2 for 18 hours. The batteries will last for a minimum of 500 charge cycles. See “General Parts List” on page 84 for ordering information.
DC Charger An optional DC powered charger is available with a vehicle plug 12 VDC adapter. It uses the same charging module as the standard AC charger. Vehicle Plug DC Adapter Charging Module Figure 8: EAGLE 2 DC Charger AC/DC Charger A charger is also available that includes both the AC adapter and the 12 VDC vehicle plug adapter. The charging module is the same as the one used for the AC charger and the DC charger.
• 30 inch stainless steel probe Figure 10: 30 Inch Stainless Steel Probe This probe is physically the same as the 30 inch aluminum probe and is intended for applications where a high level of corrosion resistance is required in the long probe tube. • 4 foot stainless steel probe Figure 11: 4 Foot Stainless Steel Probe This probe is designed for areas where it is necessary to put the probe tip in areas that are out of reach for even the 30 inch probes.
• 10 inch probe with dust filter Figure 13: 10 Inch Probe With Dust Filter This probe is designed for use where drawing water or moisture into the EAGLE 2 is not a concern. Instead of a hydrophobic filter, a cotton dust filter is used. • 32 inch telescoping probe with dust filter This probe is designed for use where it is necessary to put the probe tip in areas not accessible with the 10 inch probe with dust filter and applications where the probe tube must be collapsible for storage.
External Dilution Fittings Two external dilution fittings are available for the EAGLE 2, a 1:1 dilution fitting and a 3:1 dilution fitting. They are designed to mate with the inlet fitting and accept a sample hose or probe. The fittings are made with brass and nickel plated brass and are appropriate for use with the four standard gases. The 1:1 fitting is normally used when it is necessary to introduce air into a sample that has no oxygen or a very low level of oxygen, such as a nitrogen purged sample.
Chapter 3: Operation Overview This chapter explains how to use the EAGLE 2 to perform confined space entry monitoring or general area monitoring in Normal Mode. There are three operational modes in Normal Mode: Measuring Mode, Display Mode, and Calibration Mode. While in Normal Mode, the unit is normally operating in Measuring Mode. Display Mode and Calibration Mode are accessible from Measuring Mode. Display Mode is described in this chapter.
4. The LCD will show the following screen for about ten seconds. GAS MONITOR EAGLE 2 5. The Battery Voltage Screen appears for a few seconds. BATTERY MIN: 4.3 VOLTS BATTERY NOW: 5.2 VOLTS 6. The Active Gases Screen appears for a few seconds indicating which channels are active and their target gas. ACTIVE GASES CH4 H2S OXY CO 7. If LUNCH BREAK is turned on (see “Updating the Lunch Break Setting” on page 129), the Resume Measurement Screen appears.
from the last time the EAGLE 2 was used. See “Time in Operation Screen” on page 51 for more information about how the EAGLE 2 tracks the operating time. The short-term exposure limit (STEL) reading is reset each time the EAGLE 2 is turned on. • To reset the accumulation of these measurements, press and release the DISPLAY ADJUST NO button before the countdown reaches 0. 8.
• If the unit is due for calibration and CAL PAST DUE ACT is set to CONFIRM TO CAL, then the following screen displays and the buzzer sounds in a double pulsing pattern. CALIBRATION DATE IS PAST DUE PERFORM CALIBRATION? To perform a calibration, press and release the AIR ▲ YES button. The EAGLE 2 will enter Calibration Mode and the LCD will show the Calibration Mode main menu. See “Chapter 4: Calibration Mode” on page 55 for instructions to calibrate the EAGLE 2.
will again appear in the startup sequence. • If the unit is due for calibration and CAL PAST DUE ACT is set to NOTIFICATION ONLY, then the following alert screen displays and the buzzer sounds in a double pulsing pattern. CALIBRATION DATE IS PAST DUE Press and release the POWER ENTER RESET button to acknowledge the alert and continue with the startup sequence. 10. The Date/Time Screen appears for a few seconds. 9/12/2008 15:00:00 11.
If you wish to continue, press and release the POWER ENTER RESET button to acknowledge the failure. The gas reading for the failed sensor will be replaced by “XXX”. Replace the failed sensor as soon as possible. 13. The EAGLE 2 is now monitoring for gas in Measuring Mode. The Normal Operation Screen appears displaying the current gas reading for each target gas. CH4 0%LEL OXY 20.9vol% H2S 0.
Using the Battery Charger for Continuous Operation The battery charger can be used with an AC adapter or a vehicle plug DC adapter to run the EAGLE 2 in continuous operation instead of charging the batteries. Batteries do not need to be installed in the EAGLE 2 but if there are batteries installed, they must be NiMH batteries and they must have a charge.
7. Insert the module’s round plug into the EAGLE 2’s charging jack as shown in Figure 15 below. Figure 15: Connecting the EAGLE 2 to the Charger 8. See “Chapter 3: Operation” on page 22 for instructions for start-up and operation of the EAGLE 2. 9. While the charging module is powering the EAGLE 2, its amber LED will be off and its green LED will be on. Measuring Mode, Normal Operation When the EAGLE 2 completes its startup sequence, it is in Measuring Mode.
Monitoring an Area 1. Start up the EAGLE 2 as described above in “Start Up” on page 22. It is now in Measuring Mode. CH4 0%LEL OXY 20.9vol% H2S 0.0ppm CO 0ppm 2. Take the EAGLE 2 to the monitoring area. Put the probe tip in the area to be monitored. NOTE: If you notice that water has entered the flow system through the probe, replace the probe’s hydrophobic filter and inspect the O-ring for filter particles.
The chart below illustrates how response time is affected by the sample hose length. Table 4: EAGLE 2 Response Time vs.
Table 5: Full Response Mode Conversion Factors (Methane Calibration) Target Gas LEL Factor PPM Factor Acetone 1.40 0.70 Benzene 1.75 Butyl Acrylate Target Gas LEL Factor PPM Factor Isobutane 1.61 0.58 0.42 Isopropanol 2.22 0.89 3.95 1.34 Methane 1.00 1.00 Butyl Acetate 3.38 0.88 Methanol 1.23 1.48 2-Butyl Alcohol 1.94 0.66 Methyl Acetate 1.37 0.85 1-Butyl Alcohol 2.65 0.74 Methyl Acrylate 1.10 0.62 Cyclohexane 1.82 0.47 Methyl Ethyl Ketone 2.53 0.71 Cumene 3.
CAUTION: Any rapid increase in the combustible gas reading on the catalytic combustible channel followed by a declining or erratic reading may indicate a gas concentration above the LEL which may be hazardous. • Some gases such as silicone vapors, chlorinated hydrocarbons, and sulphur compounds can contaminate the detection elements inside the combustible sensor damaging the sensor and result in reduced response to combustible gas. Make every effort to avoid these gases.
If the Catalytic Units menu item in Setup Mode is set to CHANGE OK, the standard factory setting, then you can change the catalytic combustible channel’s units by pressing and releasing RANGE ▼ SHIFT. CH4 0%LEL OXY 20.9vol% H2S 0.0ppm CO 0ppm RANGE SHIFT CH4 0.00vol% OXY 20.9vol% H2S 0.0ppm CO 0ppm RANGE RANGE SHIFT SHIFT CH4 0ppm OXY 20.9vol% H2S 0.
The %LEL and %volume unit readings will still increase up to 100% LEL and 15 %volume respectively, which are equivalent to 150,000 ppm. The catalytic combustible sensor is slightly affected by humidity. This is not apparent when the EAGLE 2 is used for %LEL or %volume detection, but because of the high sensitivity in the ppm range, significant humidity changes can affect the ppm reading, especially in the lower part of the range.
Alarm Indications The EAGLE 2 will sound an alarm and flash the LED arrays when one of the target gas concentrations rises above the Low Alarm level, or in the case of oxygen falls below the Low Alarm level, for that gas. The EAGLE 2 also sounds an alarm and flashes the LED arrays when one of the target gas concentrations rises above the High Alarm level and when the STEL and TWA alarm levels are reached for CO and H2S.
Table 6: Alarm Types and Indications Alarm Type Visual Indications Audible Indication Low Flow • The display indicates FAIL LOW FLOW LEVEL • Alarm LED arrays flash in a double pulsing pattern once per second • Backlight turns on Double pulsing tone once per second Low Battery Warning • BATT appears vertically along the left side of LCD None Dead Battery Alarm • Gas readings replaced by RECHARGE INSTRUMENT BATTERIES • Alarm LED arrays flash in a double pulsing pattern once per second Double pulsin
are the factory settings. When the EAGLE 2 goes into gas alarm, press and release any button to silence the buzzer. If the gas concentration was still above the alarm level when the button was pressed, the LED arrays continue to flash, and the EAGLE 2 continues to display the current alarm level. The gas reading must fall below (or rise above for an oxygen low alarm) an alarm setting before you can reset the alarm.
Responding to Over Range Alarms WARNING: An over range condition may indicate an extreme combustible gas, toxic gas, or oxygen concentration. Confirm a normal condition with a different EAGLE 2 or with another gas detecting device. 1. Determine which channel is in alarm. 2. Follow your established procedure for an extreme gas condition. 3. Reset the alarm using the POWER ENTER RESET button once the alarm condition has cleared. 4.
3. If the sensor failure continues, replace the sensor as described in “Replacing a Sensor” on page 80. 4. If the sensor failure condition continues after you have replaced the sensor, contact RKI Instruments, Inc. for further instructions. Display Mode Two other operating modes are accessible when the EAGLE 2 is in Measuring Mode. They are Display Mode and Calibration Mode. This section describes using the EAGLE 2 in Display Mode.
NO button. • To enter an item when a question is asked, press and release the AIR ▲ YES button. • To change a flashing parameter, use either the AIR ▲ YES button or RANGE ▼ SHIFT button. NOTE: Each screen displays for 20 seconds. If you do not press a button within 20 seconds, the EAGLE 2 automatically returns to Measuring Mode. Peak Screen The peak screen displays the highest (lowest for oxygen) concentrations detected since the EAGLE 2 was turned on.
Battery Voltage Screen The Battery Voltage Screen displays the minimum operating voltage and the current battery voltage. Fully charged alkaline batteries typically indicate 6.0 volts; fully charged Ni-MH batteries typically indicate 5.2 volts. This screen also displays during the startup sequence. BATTERY MIN: 4.3 VOLTS BATTERY NOW: 5.2 VOLTS Gas Display Screen The Gas Display Screen gives you the option to select how the active channels are displayed.
If you do not want to change the setting, press and release the DISPLAY ADJUST NO button and the unit will return to the Gas Display Screen. Catalytic (LEL) Sensor Screen This screen appears only when either a TC sensor or an infrared combustible sensor is installed in an EAGLE 2 along with a catalytic combustible LEL sensor. See “Appendix F: TC Sensors”, “Appendix H: Infrared Methane Sensor”, or “Appendix I: Infrared Hydrocarbon Sensor” for a description of this screen and instructions to use it.
catalytic combustible channel. SELECT RELATIVE RESPONSE TO CALIBRATED GAS FOR CATALYTIC SENSOR You can select from a list of gases whose response relative to the configured gas, normally methane, is programmed into the EAGLE 2’s memory. This includes several pre-defined gases and 5 gases that can be entered into the EAGLE 2 in the field using the Eagle 2 Maintenance Data Loader Program.
The following is the complete list of factory defined gases.
STEL Screen The STEL Screen displays the short term exposure limit (STEL) readings for H2S and CO only. The STEL reading is the average reading over the last 15 minutes. S H2S 0.0ppm T CO 0ppm E L TWA Screen The TWA Screen displays the time weighted average (TWA) readings for H2S and CO only. T H2S 0.0ppm W CO 0ppm A The TWA reading is the average reading over the last 8 hours. If 8 hours have not elapsed since the last time the TWA reading was cleared, the average is still calculated over 8 hours.
AIR ▲ YES button. The following screen appears showing Channel 1 alarm points. CHANNEL 1 CH4 0- 100 %LEL LO ALRM: 10 %LEL HI ALRM: 10 %LEL LO ALRM: 5000 ppm HI ALRM:25000 ppm 2. If the channel description is too long to fit across the screen, press and release the POWER ENTER RESET button to view the remainder of the displayed channel’s description. Press and release the POWER ENTER RESET button again to return to the previous screen.
4. Press and release the AIR ▲ YES button to scroll through screens that display the rest of the active channels’ alarm settings. CHANNEL 1 CH4 0- 100 %LEL LO ALRM: 10 %LEL HI ALRM: 50 %LEL LO ALRM: 5000 ppm HI ALRM:25000 ppm AIR YES CHANNEL 2 OXY 0- 40.0 vol% LO ALRM: 19.5 vol% HI ALRM: 23.5 vol% AIR AIR YES YES CHANNEL 4 CO 0- 500 ppm LO ALRM: 25 ppm HI ALRM: 50 ppm STEL ALRM: 200 ppm TWA ALRM: 25 ppm AIR YES CHANNEL 3 H2S 0-100.0 ppm LO ALRM: 5.0 ppm HI ALRM: 30.0 ppm STEL ALRM: 5.
user IDs stored in the EAGLE 2, see the Eagle 2 Maintenance Data Loader Program Operator’s Manual. To select a different user ID: 1. With the Select User ID Screen displayed, press and release the AIR ▲ YES button. A screen appears that includes the current user ID which is indicated by the cursor next to it. Bruce > John Mike ---------------------------------------------The user IDs are displayed in groups of six. The previous group of six is displayed when the cursor is moved up past the top of the LCD.
Select Station ID Screen This screen displays only if USER/STATION ID in the Setup Mode menu is set to ON (see “Turning the User/Station ID Function On or Off” on page 118). Use this screen to select a station ID from the station ID list in the EAGLE 2’s memory. The current station ID is displayed. A station ID can be up to 16 characters long. The EAGLE 2 can store up to 128 station IDs.
screen if the cursor is moved up from the first station ID or to the next screen if the cursor is moved down from the last station ID. Any of the station IDs in the list that have not been changed from the factory setting will be shown as dashes (-). 2. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to move the cursor up and down the screen and scroll through the available station IDs to find the desired station ID. 3.
Date/Time Screen The Date/Time Screen displays the current date and time. 9/26/2009 09:08:35 Data Logging Screen CAUTION: Once you clear the data logger, you cannot retrieve any data previously stored in the data logger. The Data Logging screen displays the time remaining until the data logger memory is full and asks if you want to clear the data logger memory.
3. If you want to clear the data logger memory, press and release the AIR ▲ YES button. The following screen appears asking you to reconfirm that you want to clear the data logger memory. LOG MEMORY ARE YOU SURE YOU WANT TO PERMANENTLY ERASE DATA FROM MEMORY? 4. If you want to clear the data logger memory, press and release the AIR ▲ YES button. The unit will display the following screen as it clears the data. PLEASE WAIT ....... 5.
Data Logger Management Program and a computer with an infrared port or a USB port that runs one of the following operating systems: Windows 2000, Windows XP, or Windows Vista. If your computer has an infrared port, then no additional accessories are needed to download data from the EAGLE 2. If your computer does not have an infrared port but does have a USB port, a USB/IrDA adapter cable can be used to download data from the EAGLE 2 using the USB port.
Chapter 4: Calibration Mode Overview This section describes the EAGLE 2 in Calibration Mode. In Calibration Mode, you can move through a menu of screens to do the following: • Perform a span adjustment on all channels simultaneously using auto calibration • Perform a span adjustment on one channel at a time using single calibration • Perform a fresh air (zero) adjustment NOTE: You can set up the EAGLE 2 to alert you during the startup sequence when calibration is due.
Calibration Supplies and Equipment To calibrate the EAGLE 2, you will need: • Known calibrating samples of the gases being detected. The combustible and toxic gas samples should have concentrations between 10 and 50% of the full scale value. For example, if you are calibrating the catalytic combustible gas channel, your calibration cylinder should have a combustible gas concentration between 10% LEL and 50% LEL. An oxygen-free source, such as 100% nitrogen is recommended for setting the oxygen zero.
cylinder. This chapter also includes instructions for calibrating one channel at a time using single calibration. Entering Calibration Mode To enter Calibration Mode, do the following: 1. Find a fresh-air environment. This is an environment free of toxic or combustible gases and of normal oxygen content (20.9%). 2. While in Measuring Mode, press and hold the RANGE ▼ SHIFT button, then press the DISPLAY ADJUST NO button and release both buttons. 3.
Calibrating Using the Auto Calibration Method This method allows you to calibrate the CH4 (%LEL sensor), oxygen, H2S, and CO sensors simultaneously. It is designed for use with the RKI 4-gas calibration cylinder and is the quickest and easiest method to calibrate the EAGLE 2. Setting the Fresh Air Reading 1. While in the Calibration Mode Screen, move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT button.
4. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen. ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION Performing a Span Adjustment in Auto Calibration 1. Install the demand flow regulator onto the calibration cylinder. 2. Connect the sample tubing to the demand flow regulator. 3.
4. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 5. Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4-gas calibration cylinder.
channel. The calibration gas value begins to flash. 9. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to adjust the calibration gas setting to the desired value. NOTE: The calibration gas value cannot be set lower than the low alarm setting. If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting, enter Setup Mode and change the low alarm setting. See “Updating the Alarm Point Settings” on page 115 for instructions.
changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again. 15. When you make your selection and press the desired button, the unit returns to the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL 16. Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing. CAL IN PROCESS CH4 0 %LEL OXY 20.9 vol% H2S 0.
19. If all channels passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If any of the sensors cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and lists the sensor(s) that failed to calibrate. In the example below, the oxygen and H2S channels failed calibration. The other sensors calibrated normally.
Calibrating Using the Single Calibration Method Single Calibration allows you to calibrate one channel at a time. This is useful if you only want to calibrate one or two channels. Setting the Fresh Air Reading 1. While in the Calibration Mode Screen, move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT button. CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION 2. Press and release the POWER ENTER RESET button. The following screen appears.
4. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen. ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION Performing a Span Adjustment in Single Calibration 1. Install the demand flow regulator onto the calibration cylinder. 2. Connect the sample tubing to the demand flow regulator.
4. Move the cursor next to the SINGLE CALIBRATION menu item by using the AIR ▲ YES button. CALIBRATION MODE AUTO CALIBRATION > SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 5. Press and release the POWER ENTER RESET button. The Select Sensor Screen appears with the cursor flashing. SELECT SENSOR TO CALIBRATE >ESCAPE CH4 OXY H2S CO 6. Move the cursor next to the sensor you want to calibrate with the AIR ▲ YES and RANGE ▼ SHIFT buttons.
7. Press and release the POWER ENTER RESET button to proceed to the Single Calibration Gas Value Screen for the selected channel. The calibration gas value is flashing. SINGLE CALIBRATION CH4 50 %LEL UP/DOWN TO ADJUST CALIBRATION VALUE ENTER WHEN DONE 8. If necessary, adjust the calibration gas value to match the cylinder concentration with the AIR ▲ YES and RANGE ▼ SHIFT buttons. NOTE: The calibration gas value cannot be set lower than the low alarm setting.
response level to the applied gas. This calculated range is independent of the calibration gas value that was entered in Step 8 and Step 9 above. The adjustment range is included on the result screen to indicate the condition of the sensor. If the calibration gas value is in the adjustment range, the span adjustment will pass. If the calibration gas value is out of the adjustment range, the span adjustment will fail. If the span adjustment is successful, the following screens display.
CAUTION: When calibrating the oxygen sensor, verify the concentration of oxygen listed on the cylinder’s label. For oxygen-free samples (100% nitrogen for example), set the oxygen calibration value to 0.0%. 15. After the last channel is calibrated, disconnect the calibration tubing from the probe, then unscrew the demand flow regulator from the calibration cylinder. 16. With the Select Sensor Screen displayed, place the cursor next to ESCAPE using the AIR ▲ YES button.
Chapter 5: Maintenance Overview This chapter describes troubleshooting procedures for the EAGLE 2. It also includes procedures for replacing and recharging the batteries and replacing various consumable parts. WARNING: RKI Instruments, Inc. recommends that service, calibration, and repair of RKI instruments be performed by personnel properly trained for this work. Replacing sensors and other parts with original equipment does not affect the intrinsic safety of the instrument.
Table 9: Troubleshooting the EAGLE 2 Symptoms Probable Causes Recommended Action • The unit indicates flow failure and does not recover when POWER ENTER RESET is pressed and released. • The probe tube is clogged. • The hydrophobic filter disk in the probe is dirty. • The sample hose has a kink or obstruction. • The internal hydrophobic filter is dirty. • The pump is malfunctioning. 1. Inspect the probe tube for any obstructions. 2.
Replacing or Recharging the Batteries WARNING: To prevent ignition of a hazardous atmosphere, batteries must only be changed or charged in an area known to be nonhazardous. Replace or charge the batteries when the EAGLE 2 indicates that it is in low battery warning. When in low battery warning, BATT appears vertically along the left side the LCD. CH4 0%LEL B OXY 20.9vol% A H2S 0.
Figure 16: Removing the Battery Case 3. Pull the battery case away from the bottom case. The thumbscrew is captive and will not fall out. 4. Carefully remove the old batteries. Verify that the battery compartment and electrical contacts are clean. 5. Carefully install the new C-size batteries. Follow the battery diagram inside the battery case. Make sure the batteries are pushed in all the way. Figure 17: Installing the Batteries 6.
Recharging the Ni-MH Batteries CAUTION: Use with Ni-MH battery p/n 49-1330RK. Charge only with RKI charger model 49-2175RK, 49-2176RK, or 49-2177RK. Use of other rechargeable batteries or chargers or charging of other rechargeable batteries in the EAGLE 2 will void the warranty. WARNING: Do not plug the charger cable into a battery pack that contains alkaline batteries. Do not attempt to charge alkaline batteries. The charging module can either be used with an AC adapter or a vehicle plug DC adapter.
6. While the batteries are charging, the green indicator LED will be off and the amber one will be on. 7. The charging module has an internal timeout feature set at 9.5 hours. A full charge should be reached in less than 9.5 hours. When a full charge has been reached, both the green and yellow LEDs will be on. 8. If charging should fail, the green indicator LED will be off and the amber one will be blinking. Table 10 summarizes the battery charger conditions.
O-ring (Inside) Hydrophobic Filter Disk Particle Filter, Not Included in PID Probe O-ring Probe Body Figure 19: Replacing the Particle Filter and Hydrophobic Filter Disk 2. Remove the white hydrophobic filter disk from the top of the particle filter or from the probe body. 3. Remove the particle filter from the probe body. 4. Clean the inside of the probe body if necessary. 5. Hold the probe half that has the plastic tube fitting and the probe tube with the fitting and tube facing down. 6.
7. Place the new filter disk flat on top of the particle filter. Make sure it is centered over the particle filter. 8. Carefully screw the other half of the probe body onto the half with the particle filter and filter disk while keeping the probe oriented vertically to keep the disk centered. 9. When you feel the O-ring being compressed, grasp both ends of the probe and tighten them together very firmly to ensure a seal. 10. To test the seal, do the following.
6. Locate the hydrophobic filter. It is over the oxygen sensor. Note which side of the hydrophobic filter has the RKI logo and part number. This is the inlet side and should be facing toward the front of the EAGLE 2. RKI Logo & Part Number on This Side Hydrophobic Filter Grey Connecting Tube, 2X Bottom Case Shown Without Tubing For Clarity Figure 20: Replacing the Hydrophobic Filter 7. Pull the grey connecting tubes off of each end of the filter and remove it. 8.
Replacing the Charcoal Filter 1. Verify that the EAGLE 2 is off. 2. Place the EAGLE 2 upside down on a flat surface or hold it upside down. 3. Unscrew the three case screws until they disengage from the top case. They are captive screws so they will not fall off of the bottom case. 4. Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case. Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB. 5.
the flow chamber and push it in until it bottoms out. 9. Insert the charcoal filter endcap into the end of the charcoal filter and push it in until it bottoms out. 10. Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case. If the main PCB is not seated properly, then it may be damaged when the top case is re-installed. 11. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case.
5. Lay the top case down next to the bottom case to allow access to the flow system. 6. Locate the sensor you want to replace and remove it from the flow chamber. H2S Sensor LEL Sensor Cable LEL Sensor Bracket CO Sensor Oxygen Sensor Bracket LEL Sensor Oxygen Sensor Main PCB LEL Connector H2S Connector CO Connector Oxygen Connector Flow Block Figure 22: Replacing a Sensor 7. To remove the LEL sensor, do the following: • Unscrew and remove the two screws that hold down the LEL sensor bracket.
• If the sensor stayed in the flow chamber, grasp the sensor and pull it out of the flow chamber. 8. To remove the oxygen sensor, do the following: • Unscrew the two screws that hold the oxygen sensor bracket a few turns so that you can rotate and remove the oxygen sensor bracket. Make sure to note the routing of the oxygen sensor cable to the main PCB so that you can route the replacement sensor cable the same way.
• Insert the LEL sensor into the LEL sensor chamber in the flow chamber. • Line up the holes in the LEL sensor bracket with the two standoffs on either side of the LEL sensor chamber. • Install the two sensor bracket screws tightening them a little at a time alternately to push the sensor into its chamber evenly. 12.
General Parts List Table 11 lists part numbers for the EAGLE 2’s replacement parts and accessories.
Table 11: General Parts List (cont.
Appendix A: Calibrating with a Sample Bag Overview The EAGLE 2 can be calibrated with a gas bag calibration kit instead of a demand flow regulator kit. This appendix describes how to use a sample bag calibration kit to calibrate the EAGLE 2. A parts list at the end of this appendix lists spare parts for the calibration kit.
To calibrate the EAGLE 2, you will need: • Known calibrating samples of the gases being detected. The combustible and toxic gas samples should have concentrations between 10 and 50% of the full scale value. For example, if you are calibrating the catalytic combustible gas channel, your calibration cylinder should have a combustible gas concentration between 10% LEL and 50% LEL. An oxygen-free source, such as 100% nitrogen is recommended for setting the oxygen zero.
3. If the unit prompts you for the password, enter it by using the AIR ▲ YES and RANGE ▼ SHIFT buttons to select each password number and then pressing and releasing POWER ENTER RESET to enter the number and move on to the next one. 4. The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION.
2. Press and release the POWER ENTER RESET button. The following screen appears. PERFORM AIR ADJUST? 3. Press and release the AIR ▲ YES button to continue. If you do not want to continue, press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen. 4. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen.
2. Install the probe on the EAGLE 2 inlet fitting. Make sure the probe is complete with internal O-ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration. See Figure 19, “Replacing the Particle Filter and Hydrophobic Filter Disk” on page 76 for an illustration of the internal parts of the probe. 3. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button.
5. To adjust the values on the screen, hold down the RANGE ▼ SHIFT button, then press the DISPLAY ADJUST NO button and release both. The following screen appears with the cursor next to CH4. ADJUST AUTO CALIBRATION VALUES > CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm 6. Place the cursor next to the channel whose gas value you want to change using the AIR ▲ YES and RANGE ▼ SHIFT buttons. 7. Press and release the POWER ENTER RESET button to select the channel. The calibration gas value begins to flash.
12. Press and release the POWER ENTER RESET button. The following screen appears. DO YOU WANT TO STORE NEW VALUE(S) IN MEMORY FOR FUTURE CALIBRATIONS? PRESS YES OR NO 13. If you select YES by pressing and releasing the AIR ▲ YES button, the changes that you made will be saved in the EAGLE 2’s memory as the new auto calibration gas values.
17. Fill the gas collection bag by screwing the fixed flow regulator or dispensing valve onto the calibration cylinder and turning the knob counterclockwise. 18. Allow the gas to dispense until the gas collection bag is a little over half full. 19. Turn the knob clockwise to stop the gas flow, clamp down the hose clamp and remove the regulator or dispensing valve from the cylinder. 20. Disconnect the tubing from the regulator or dispensing valve. 21. Open the hose clamp on the gas bag tubing. 22.
If any of the sensors cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and lists the sensor(s) that failed to calibrate. In the example below, the oxygen and H2S channels failed calibration. The other sensors calibrated normally. FAIL SENSOR < > < > ENTER TO CONTINUE The buzzer and LED arrays activate in a double pulsing pattern. Press and release the POWER ENTER RESET button to reset the alarm and return to the Calibration Mode Screen.
2. Press and release the POWER ENTER RESET button. The following screen appears. PERFORM AIR ADJUST? 3. Press and release the AIR ▲ YES button to continue. If you do not want to continue, press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen. 4. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen.
Performing a Span Adjustment in Single Calibration 1. Slide the tubing clamp onto the tubing and connect the tubing to the sample bag’s inlet. Leave the clamp unclamped for now. 2. Connect the other end of the tubing to the regulator’s or dispensing valve’s hose barb fitting. 3. Install the probe on the EAGLE 2 inlet fitting. Make sure the probe is complete with internal O-ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration.
and press and release POWER ENTER RESET to return to the Calibration Mode Screen. If you do want to continue with the calibration, proceed with the next step. 7. Press and release the POWER ENTER RESET button to proceed to the Single Calibration Gas Value Screen for the selected channel. The calibration gas value is flashing. SINGLE CALIBRATION CH4 50 %LEL UP/DOWN TO ADJUST CALIBRATION VALUE ENTER WHEN DONE 8.
12. Turn the knob clockwise to stop the gas flow, clamp down the hose clamp and remove the regulator or dispensing valve from the cylinder. 13. Disconnect the tubing from the regulator or dispensing valve. 14. Open the hose clamp on the gas bag tubing. 15. Connect the tubing from the gas bag to the rigid tube on the probe. Allow the gas to flow for one minute.
In the example above, the EAGLE 2 could have adjusted the reading as low as 11 %LEL and as high as 43 %LEL. Since the calibration gas value entered was 50 %LEL, the unit failed the span adjustment. The buzzer and alarm LED arrays activate in a double pulsing pattern. Press and release the POWER ENTER RESET button to reset the alarm and return to the Select Sensor Screen. Attempt to calibrate again. If the failure continues, investigate the cause.
Parts List Table 12: Sample Bag Calibration Kit Spare Parts Part Number Description 06-1248RK-03 Calibration kit tubing,3 foot length 81-0090RK-01 Calibration cylinder, 3-gas mix, LEL/Oxygen/CO, 34 liter steel 81-0154RK-04 Calibration cylinder, 4-gas mix, LEL/Oxygen/CO/H2S, 34 liter aluminum 81-1001RK Dispensing valve, for 17/34 liter steel cylinders 81-1051RK-60 Regulator with gauge and knob, 34 liter aluminum/58 liter/103 liter cylinders, 6 LPM 81-1126RK Gas bag with clamp and hose barb, 9” x
Appendix B: Setup Mode Overview This appendix describes the EAGLE 2 in Setup Mode.
• turn the confirmation alert feature on or off • turn the password feature on or off and set the password • reset the instrument parameters to their default settings • turn the lunch break function on or off • turn the span factor on or off • select the language • return to normal operation The EAGLE 2 is factory-set to suit most applications. Update settings in Setup Mode only if required for your specific application. The description of each item below indicates the factory setting for each item.
• Press the DISPLAY ADJUST NO button while in a screen where you are entering or updating parameters to exit the screen without saving any changes. Using Setup Mode WARNING: The EAGLE 2 is not in operation as a gas detector while in Setup Mode. 1. Take the EAGLE 2 to a non-hazardous location and turn it off if it is on. 2. Press and hold the AIR ▲ YES and RANGE ▼ SHIFT buttons, then press and hold the POWER ENTER RESET button. When you hear a beep, release the buttons. 3.
5. If the unit prompts you for the password, enter it by using the AIR ▲ YES and RANGE ▼ SHIFT buttons to select each password number and then pressing and releasing the POWER ENTER RESET button to enter it and move on to the next number until all of the numbers are entered. The main menu displays. It displays six menu items at a time. >SET DATE & TIME SET DATE FORMAT SET BATTERY TYPE CONFIGURE CHANNELS CONFIGURE GASES CATALYTIC UNITS 6.
2. Press and release POWER ENTER RESET. The Set Date Format screen appears with the current setting flashing. SET DATE FORMAT UP/DOWN THEN ENTER MM/DD/YYYY 3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting. 4. Press and release POWER ENTER RESET to save the setting and return to the main menu. Setting the Battery Type This menu item allows you to select between alkaline and Ni-MH batteries.
configuration. CAUTION: Before changing the channel configuration, confirm that the correct sensors and electronic hardware are installed in the EAGLE 2 and that its construction and flow system are appropriate for the installed sensors. Operation of the EAGLE 2 with a flow system or construction not compatible with the installed sensors will result in inaccurate readings. Consult RKI Instruments, Inc. if you cannot confirm either of these items. 1.
4. Press and release POWER ENTER RESET. The available configuration options are shown. In the example below, channel 3 has been selected for configuration. RANGE CHANNEL 3 (H2S) CHANNEL 3 (H2S) SHIFT > CAT (CH4) OXY (OXY) H2S (H2S) CO (CO) AIR > OP1 (---) OP2 (---) OP3 (---) NOT USED YES 5. Use AIR ▲ YES and RANGE ▼ SHIFT to move the cursor next to the desired configuration for the selected channel. In the example below, the cursor has been moved next to the NOT USED selection.
8. Use RANGE ▼ SHIFT to move the cursor next to the END menu item. NOTE: If you want to exit to the main menu without saving any channel configuration changes, press and release DISPLAY ADJUST NO. 9. Press and release POWER ENTER RESET to save the changes and return to the main menu. Configuring the Combustible Gas This menu item allows you to configure the gas for a catalytic, TC (thermal conductivity), or PID (photo ionization detector) sensor.
LEL is shown in terms of ppm. The available choices are on two screens. RANGE CAT(0-50000ppm CH4) 50000 ppm CH4 11000 ppm HEX 40000 ppm H2 > 18000 ppm IBU SHIFT AIR CAT(0-50000ppm CH4) > 50000 ppm GAS 50000 ppm CAT 10000 ppm - - - * END YES All of the gases except for the one with the asterisk (*) next to it are predefined. The gas with the asterisk next to it is user defined. The user defined gas can be used if the desired gas is not one of the pre-defined gases.
WARNING: Do not configure the catalytic sensor gas to hydrogen if you are monitoring for general hydrocarbons. Only use this selection if you are monitoring exclusively for hydrogen or if you do not want to see a significant response to other combustible gases. • 18000 ppm IBU This selection is for isobutane. The LEL for isobutane is 1.8%, or 18,000 ppm.
• 10000 ppm - - - * This selection is a user defined selection. The factory setting is 10000 ppm - - - *, with the asterisk (*) indicating that it is user defined. If this selection has been updated in the field, it will appear differently, but the asterisk will always remain next to the menu item to indicate it is user defined.
NOTE: If you define a gas whose LEL is above 50,000 ppm, the %LEL reading in Measuring Mode will reflect the defined ppm ratio, but the ppm reading in Measuring Mode will not indicate above 50,000 ppm.
Mode. 13. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired catalytic sensor voltage, 1.30 volts or 2.40 volts. 14. Press and release POWER ENTER RESET. The confirmation screen appears. In the example below, the user defined gas has been selected and defined as propane with the gas name set to PRO. CHANGE TO PRO ? PRESS YES OR NO 15. If you want to accept the gas configuration change, press and release AIR ▲ YES. The unit will return to the Configure Gases screen.
2. Press and release POWER ENTER RESET. The Catalytic Units screen appears with the current setting flashing at the bottom of the screen. CATALYTIC UNITS UP/DOWN THEN ENTER CHANGE OK 3. Use AIR ▲ YES or RANGE ▼ SHIFT to scroll through the choices, CHANGE OK, vol% ONLY, %LEL ONLY, and ppm ONLY. 4. When the desired setting is on the screen, press and release POWER ENTER RESET to save the setting and return to the main menu.
Updating the Alarm Point Settings This menu item allows you to update one or more alarm points (the reading at which the EAGLE 2 recognizes the alarm condition). 1. From the main menu, place the cursor next to ALARM POINTS. 2. Press and release POWER ENTER RESET. The Change Alarm Point Settings Screen appears and all detection channels are displayed. CHANGE ALARM POINT SETTINGS > 1: CH4 2: OXY 3: H2S 4: CO END 3. Move the cursor next to the channel of the alarm point you want to update.
5. Press and release POWER ENTER RESET. The alarm point or alarm operation (oxygen only) will begin to flash. 6. Use AIR ▲ YES and RANGE ▼ SHIFT to adjust the alarm point or alarm operation (oxygen only) to the desired setting. Keep the following in mind: • The low alarm cannot be set higher than the high alarm and the high alarm cannot be set lower than the low alarm. • Any alarm setting can be turned off by adjusting it to its lowest setting. The setting will be displayed as OFF.
1. From the main menu, place the cursor next to ALARM LATCHING. 2. Press and release POWER ENTER RESET. The Alarm Latching Screen appears. ALARM LATCHING UP/DOWN THEN ENTER LATCHING 3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting. 4. Press and release POWER ENTER RESET to save the setting and return to the main menu. Updating the Alarm Silence Setting With ALARM SILENCE set to ON (factory setting), pressing and releasing any button silences the buzzer when the EAGLE 2 is in alarm.
3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting. 4. Press and release POWER ENTER RESET to save the setting and return to the main menu. Turning the User/Station ID Function On or Off With USER/STATION ID set to ON, the ID Screen displays during start up and the Select User ID Screen and Select Station ID Screen appear in Display Mode. The ID’s can be selected in Display Mode.
2. Press and release POWER ENTER RESET. The Adjust Auto Calibration Values Screen appears. The auto calibration value for each channel is shown. ADJUST AUTO CALIBRATION VALUES > CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm 3. Use AIR ▲ YES or RANGE ▼ SHIFT to place the cursor next to the auto calibration value you want to change. 4. Press and release POWER ENTER RESET. The auto calibration value begins to flash indicating it can be adjusted. 5.
3. Use AIR ▲ YES and RANGE ▼ SHIFT to adjust the time to the desired setting. 4. Press and release POWER ENTER RESET to save the setting and return to the main menu. Updating the Auto Fresh Air Setting This setting allows you to configure the EAGLE 2 so that a fresh air adjustment takes place automatically as part of the instrument startup sequence. If AUTO FRESH AIR ADJ is set to ON, the EAGLE 2 performs a fresh air adjustment at the end of the startup sequence before entering Normal Operation.
2. Press and release POWER ENTER RESET. The Data Log Interval Screen appears. DATA LOG INTERVAL UP/DOWN THEN ENTER 30 SECONDS 3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting. 4. Press and release POWER ENTER RESET to save the setting and return to the main menu. Updating the Data Log Overwrite Setting With DATA LOG OVERWRITE set to ON (factory setting), the EAGLE 2 writes over the oldest data with new data when the data logger memory is full.
shows the remaining log time and does not give you the opportunity to clear the logged data. 1. From the main menu, place the cursor next to DATA LOG MEMORY. 2. Press and release POWER ENTER RESET. The Prompt to Clear Data Log Memory? Screen appears. PROMPT TO CLEAR DATA LOG MEMORY? UP/DOWN THEN ENTER ON 3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting. 4. Press and release POWER ENTER RESET to save the setting and return to the main menu.
1. From the main menu, place the cursor next to CAL REMINDER. 2. Press and release POWER ENTER RESET. The Calibration Reminder Screen appears. CALIBRATION REMINDER UP/DOWN THEN ENTER ON 3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting. 4. Press and release POWER ENTER RESET to save the setting and return to the main menu. Updating the Calibration Past Due Action Setting This item defines what indication is given during start up when calibration is due and CAL REMINDER is set to ON.
2. Press and release POWER ENTER RESET. The Calibration Past Due Action Screen appears. CALIBRATION PAST DUE ACTION UP/DOWN THEN ENTER CONFIRM TO CAL 3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting. 4. Press and release POWER ENTER RESET to save the setting and return to the main menu. Updating the Calibration Interval This setting defines the amount of time between calibrations. The time can be set in 1 day increments. The minimum setting is 1 day and the maximum setting is 365 days.
With LC/BH MODE SELECT set to LEAK CHECK ONLY, the Mode Select Screen appears when the unit is turned on. You are able to select from Normal Mode and Leak Check Mode. With LC/BH MODE SELECT set to OFF, the Mode Select Screen does not appear when the unit is turned on and the unit goes into Normal Mode after the start up sequence. This setting is factory set to OFF when a unit is shipped unless the instrument is ordered for bar hole measurement or leak checking use.
4. Press and release POWER ENTER RESET to save the setting and return to the main menu. Zero Follower Settings The ZERO FOLLOWER setting is not intended for field adjustment. The default setting for most target gases is ON. The default setting for carbon dioxide channels and some configurations of non-standard toxic gas channels is OFF. The oxygen channel does not support this feature. Zero Suppression Settings The ZERO SUPRESSION setting is not intended for field adjustment.
Turning the Password Function On or Off With CHANGE PASSWORD set to ON, the EAGLE 2 prompts you for a password when you enter Calibration Mode or Setup Mode. NOTE: If a calibration is due and CAL REMINDER is set to ON and CAL PAST DUE ACT is set to MUST CALIBRATE, no password will be required to perform a calibration during startup. With CHANGE PASSWORD set to OFF (factory setting), no password is required to enter Calibration Mode or Setup Mode. 1.
password is saved and you return to the main menu. Restoring the Default Settings Each of the EAGLE 2 setup parameters, such as the auto calibration values, zero and span settings, or parameters in Setup Mode, has a default setting. For the items in Setup Mode, the default settings are the same as the standard factory settings. If you want to return the EAGLE 2 to its default configuration, it is possible to do so by using the Default Settings menu item in Setup Mode.
4. Press and release AIR ▲ YES. A screen appears asking you to confirm that you want to restore the default configuration. ARE YOU SURE YOU WANT TO SET DEFAULT CONFIGURATION? 5. Press and release AIR ▲ YES. The screen will indicate that the default configuration has been restored and return to the main menu. Updating the Lunch Break Setting With LUNCH BREAK set to OFF (factory setting), the EAGLE 2 automatically starts new TWA and PEAK reading collection and resets the time in operation at startup.
1. From the main menu, place the cursor next to SPAN FACTOR. 2. Press and release POWER ENTER RESET. The Span Factor Screen appears. SPAN FACTOR DISPLAY UP/DOWN THEN ENTER ON 3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting. 4. Press and release POWER ENTER RESET to save the setting and return to the main menu. Updating the Language Setting This setting allows you to select the language for the EAGLE 2’s user interface.
Exiting Setup Mode 1. From the main menu, place the cursor in front of NORMAL OPERATION at the bottom of the menu. 2. Press and release POWER ENTER RESET. 3. A screen appears that asks if you want to save the changes you have made. SAVE ALL CHANGES IN MEMORY? NOTE: If you entered Setup Mode and did not make any changes, the above screen will still appear. In this case, press and release DISPLAY ADJUST NO to proceed to exit Setup Mode and begin the EAGLE 2’s startup sequence. 4.
Appendix C: Sub PCBs Overview An EAGLE 2 that has one or more of the standard four sensors, catalytic LEL, oxygen, H2S, and CO, and no optional sensors does not have any sub PCBs installed. The sub PCBs are used to add circuitry to the EAGLE 2 that supports various optional sensors. An EAGLE 2 has provisions to install up to three sub PCBs.
retained in place with a screw/flat washer/lock washer. Any sub PCB can be installed in any of the sub PCB positions. A sensor that is supported by a sub PCB connects to that sub PCB with a cable. There are four types of sub PCBs, each supporting one type of optional sensor: • The PID sub PCB supports a low range or a high range PID (photo ionization detector) sensor. • The ESM-01 sub PCB supports any of the ESM-01 family toxic sensors.
CONFIGURE CHANNELS LCD screen below illustrates this situation. CONFIGURE CHANNELS > CH1: CH2: CH3: CH4: CAT (CH4) OXY (OXY) H2S (H2S) OP1 (---) The “---” to the right of OP1 indicates that no sub PCB is installed in position SUB1. See “Configuring the Channels” on page 105 for a complete description of the CONFIGURE CHANNELS Setup Mode menu item. Sub PCBs and CONFIGURE GASES The CONFIGURE GASES menu item in Setup Mode allows you to configure the gas for a catalytic, TC, or PID sensor that is installed.
• The currently configured gas The catalytic sensor is always defined and configured even if it is not setup as active in CONFIGURE CHANNELS. If any of the Sub PCBs are not installed, then the corresponding sensor is undefined and the next two fields are filled with dashes (---).
Appendix D: PID Sensors Overview The PID (photo ionization detector) sensors are used for applications where high sensitivity is needed to monitor ppm levels of VOCs (volatile organic compounds). This appendix describes the EAGLE 2’s PID sensors and includes instructions to use an EAGLE 2 that has a PID sensor installed. It also includes instructions to maintain and replace a PID sensor.
Top Case Not Shown Charcoal Filter PID Sensor Pump Standard Flow Chamber CO Sensor Oxygen Sensor Oxygen Sensor H2S Sensor Bracket LEL Sensor and Sensor Bracket Figure 25: Typical PID Sensor Location PID Sensor & Sensor Adapter The PID sensor is a cylindrical sensor with a diffusion opening on the front and 3 pins on the back. It is plugged into a sensor adapter with a 5 wire cable that terminates in a 5-position connector.
use the probe intended for PID use that does not have a particle filter installed and that you install the probe directly to the inlet fitting when monitoring for gas with the PID sensor. The figure below shows an exploded drawing of the PID probe. Figure 26: PID Probe Start Up and Normal Operation For instructions to startup and use an EAGLE 2 that includes a PID sensor, reference “Start Up” on page 22, “Measuring Mode, Normal Operation” on page 29, and “Measuring Mode, Alarms” on page 35.
state, the alternating numbers will be replaced by a gas reading. If the sensor doesn’t exit its warm up state after 5 minutes, see “Troubleshooting” on page 160. • If using a probe, be sure to use the PID probe that does not have a particle filter installed. If it is necessary to use a sample hose for any reason, you must use the teflon lined hose that is supplied with the PID instrument. A 5 foot hose is supplied as standard. 10, 15, and 20 foot hoses are also available.
isobutylene (IBL), you can select isopropyl alcohol (IPA) from a gas list accessible from the PID Relative Response Screen in Display Mode so that the PID channel responds to gas as if it were calibrated to isopropyl alcohol. The EAGLE 2 will clear the gas configuration change when it is turned off and will return to the programmed configuration when it is turned on again.
2. With the PID Sensor Relative Response Screen displayed, press and release AIR ▲ YES. A list of gases will appear on the screen with EXIT at the top of the list. >EXIT ACETONE BENZENE DIESEL FUEL NO 1 ETHANOL GASOLINE There are multiple screens of gases. The following is the complete list of factory defined gases along with their detection ranges, low alarm, high alarm, STEL, and TWA settings. Table 14 is a list of the low range values and Table 15 is a list of the high range values.
Table 14: Low Range Relative Response Gas List PID 0-50.
PID Calibration A PID channel can be calibrated using the auto calibration method or the single calibration method. To calibrate a PID channel using the single calibration method, see “Calibrating Using the Single Calibration Method” on page 64 and follow the instructions for calibrating a single channel. NOTE: The factory setting for the Span Factor menu item in Setup Mode is OFF for an EAGLE 2 with a PID channel.
5. The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 6. Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT button. CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION 7. Press and release the POWER ENTER RESET button. The following screen appears. PERFORM AIR ADJUST? 8.
9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen. ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION 10. Install the demand flow regulator onto the multi-gas calibration cylinder. 11. Connect the sample tubing to the demand flow regulator. 12.
13. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14. Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4-gas calibration cylinder.
channel. The calibration gas value begins to flash. 18. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to adjust the calibration gas setting to the desired value. NOTE: The calibration gas value cannot be set lower than the low alarm setting. If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting, enter Setup Mode and change the low alarm setting. See “Updating the Alarm Point Settings” on page 115 for instructions.
changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again. 24. When you make your selection and press the desired button, the unit returns to the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL 25. Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing. CAL IN PROCESS CH4 0 %LEL OXY 20.9 vol% H2S 0.
28. If all channels passed calibration, the following screen sequence occurs. AUTO CALIBRATION DONE REMOVE GAS AUTO CALIBRATION END CAL GAS VALUE IBL 100 ppm ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and lists the sensor(s) that failed to calibrate. In the example below, the oxygen and H2S channels failed calibration. The other sensors calibrated normally.
31. If you want to change the PID channel’s calibration gas value, follow Step 5 - Step 24 above beginning with the PID Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen. CAL GAS VALUE IBL 100 ppm ENTER TO BEGIN CAL 32. With the PID Calibration Gas Value Screen displayed, press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the PID channel with CAL IN PROCESS flashing.
36. If the PID channel passed calibration, the following screen sequence occurs. AUTO CALIBRATION DONE REMOVE GAS AUTO CALIBRATION END CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the PID channel cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and indicates that the PID sensor failed to calibrate. FAILED SENSOR ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern.
Calibrating with a 5-Gas Cylinder 1. See “Calibration Supplies and Equipment” on page 56 for a description of the necessary calibration supplies. See Table 17 on page 178 for available cylinders. Make sure your calibration cylinder is appropriate for the PID detection range. 2. Find a fresh-air environment. This is an environment free of toxic or combustible gases and of normal oxygen content (20.9%). 3.
If you do not want to continue, press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen. 9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen. ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION 10.
13. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14. Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4-gas calibration cylinder.
channel. The calibration gas value begins to flash. 18. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to adjust the calibration gas setting to the desired value. NOTE: The calibration gas value cannot be set lower than the low alarm setting. If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting, enter Setup Mode and change the low alarm setting. See “Updating the Alarm Point Settings” on page 115 for instructions.
changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again. 24. When you make your selection and press the desired button, the unit returns to the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL 25. Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing. CAL IN PROCESS CH4 0 %LEL OXY 20.9 vol% H2S 0.
28. If all channels passed calibration, the following screen sequence occurs. AUTO CALIBRATION DONE REMOVE GAS AUTO CALIBRATION END CAL GAS VALUE IBL 100 ppm ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and lists the sensor(s) that failed to calibrate. In the example below, the oxygen and H2S channels failed calibration. The other sensors calibrated normally.
29. If you want to change the PID channel’s calibration gas value, follow Step 5 - Step 24 above beginning with the PID Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen. CAL GAS VALUE IBL 100 ppm ENTER TO BEGIN CAL 30. With the PID Calibration Gas Value Screen displayed, press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the PID channel with CAL IN PROCESS flashing.
33. If the PID channel passed calibration, the following screen sequence occurs. AUTO CALIBRATION DONE REMOVE GAS AUTO CALIBRATION END CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the PID channel cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and indicates that the PID sensor failed to calibrate. FAILED SENSOR ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern.
Maintenance The PID sensor includes user serviceable parts. They are the lamp and the electrode stack. The following sections include troubleshooting instructions as well as procedures for cleaning the lamp, replacing the lamp, replacing the electrode stack, and replacing the PID sensor. Troubleshooting The troubleshooting table describes symptoms, probable causes, and recommended action for problems you may encounter with the PID sensor.
• a small vial of aluminum oxide powder • 40 cotton swabs • 10 finger cots Perform the following procedure to clean the PID lamp: 1. Verify that the EAGLE 2 is off. 2. Place the EAGLE 2 upside down on a flat surface or hold it upside down. 3. Unscrew the three case screws until they disengage from the top case. They are captive screws so they will not fall off of the bottom case. 4. Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case.
10. Hold the PID sensor steady on the working surface with one hand and using the other hand, locate the tabs on the electrode stack removal tool and insert them into the slots on the side of the PID sensor near the face. Figure 27: Using Removal Tool 11. Squeeze the removal tool to push the tabs into the sensor slots until the electrode stack and lamp are released. 12. Carefully lift the PID sensor body away from the electrode stack and lamp.
16. Use this cotton swab to polish the PID lamp window. Use a circular motion, applying light pressure to clean the lamp window. Do not touch the lamp window with your fingers. Figure 29: Polishing the Electrode Lamp Window 17. Continue polishing until you can hear a squeaking sound made by the cotton swab moving over the window surface. This usually occurs after about 15 seconds of polishing. 18. Remove the residual powder from the lamp window with a clean cotton swab.
20. Hold the electrode stack between the thumb and forefinger of one hand and place the window end of the lamp inside the O-ring seal in the electrode stack with the other hand as shown below. Figure 30: Reinstalling the Electrode Lamp 21. Twisting the lamp slightly during insertion will help to ensure the lamp window is snug against the stack's front electrode. The lamp should be supported by the O-ring. 22.
28. Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case. If the main PCB is not seated properly, it may be damaged when the top case is re-installed. 29. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case. If you have any difficulty mating the top and bottom cases, inspect the placement of the main PCB and the placement of the top case gasket. 30.
using the other hand, locate the tabs on the electrode stack removal tool and insert them into the slots on the side of the PID sensor near the face. Figure 31: Using Removal Tool 11. Squeeze the removal tool to push the tabs into the sensor slots until the electrode stack and lamp are released. 12. Carefully lift the PID sensor body away from the electrode stack and lamp. If the lamp remains lodged in the sensor body, carefully remove it with tweezers. 13.
Figure 32: Reinstalling the Electrode Lamp 16. Twisting the lamp slightly during insertion will help to ensure the lamp window is snug against the stack's front electrode. The lamp should be supported by the O-ring. 17. Continuing to hold the electrode stack between your forefinger and thumb, carefully insert the lamp into the lamp cavity in the sensor ensuring that the lamp remains in position. 18.
24. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case. If you have any difficulty mating the top and bottom cases, inspect the placement of the main PCB and the placement of the top case gasket. 25. Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case. 26. Calibrate the PID channel as described in “PID Calibration” on page 143.
NOTE: At this point it is recommended that the finger cots be used on the fingers handling the lamp. Finger cots are included with the lamp cleaning kit. 10. Hold the PID sensor steady on the working surface with one hand and using the other hand, locate the tabs on the electrode stack removal tool and insert them into the slots on the side of the PID sensor near the face. Figure 33: Using Removal Tool 11.
Figure 34: Reinstalling Electrode Lamp 16. Twisting the lamp slightly during insertion will help to ensure the lamp window is snug against the stack's front electrode. The lamp should be supported by the O-ring. 17. Continuing to hold the electrode stack between your forefinger and thumb, carefully insert the lamp into the lamp cavity in the sensor ensuring that the lamp remains in position. 18.
is resting on the bottom of the bottom case. If the main PCB is not seated properly, it may be damaged when the top case is re-installed. 24. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case. If you have any difficulty mating the top and bottom cases, inspect the placement of the main PCB and the placement of the top case gasket. 25. Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case.
12. Insert the sensor adapter into the PID flow chamber and push it in until it bottoms out. 13. Plug the PID cable connector into the PID sub PCB on the main PCB. 14. Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case. If the main PCB is not seated properly, it may be damaged when the top case is re-installed. 15. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case.
4. The “S” will then disappear and the following screen will appear for a few seconds. GAS MONITOR EAGLE 2 5. If the unit prompts you for the password, enter it by using AIR ▲ YES and RANGE ▼ SHIFT to select each password number and then pressing and releasing the POWER ENTER RESET button to enter it and move on to the next number until all of the numbers are entered. The main menu displays. It displays six menu items at a time.
8. Use RANGE ▼ SHIFT to move the cursor down the menu to the PID sensor. CONFIGURE GASES CAT : CH4 (CAT) > OP1 : IBL (PID) OP2 : --- (---) OP3 : --- (---) 9. To change the PID sensor gas configuration, press and release POWER ENTER RESET. 10. A screen appears that indicates the PID type, low range or high range (flashing), and the detection range for the currently configured gas. In the example below, the PID sensor is currently configured as a high range sensor for 0 - 2,000 ppm isobutylene.
Table 16: PID Gas Configurations Detection Range Full Scale Low Range/High Range Gas Name (LCD Abbreviation) Acetone (ACT) 30.00 ppm/1000 ppm Benzene (BNZ 25.00 ppm/1000 ppm Diesel Fuel No. 1 (DSL) 40.00 ppm/1500 ppm Ethanol (ETA) 400.0 ppm/15000 ppm Gasoline (GSL) 50.00 ppm/2000 ppm Isobutylene (IBL) Standard Factory Gas Configuration 50.00 ppm/2000 ppm Isopropanol (IPA) 200.0 ppm/5000 ppm JP-5 Fuel (JP5) 30.00 ppm/1000 ppm Methyl Ethyl Ketone (MEK) 40.00 ppm/1500 ppm Toluene (TOL) 25.
13. Press and release the POWER ENTER RESET button. The LCD will ask for confirmation that you want to change the configured gas. CHANGE TO IPA ? PRESS YES OR NO 14. To change the gas configuration, press and release AIR ▲ YES. The unit will return to the Configure Gases Screen reflecting the new gas configuration. CONFIGURE GASES CAT : CH4 (CAT) > OP1 : IPA (PID) OP2 : --- (---) OP3 : --- (---) If you do not want to change the gas configuration, press and release DISPLAY ADJUST NO.
19. A screen appears that asks if you want to save the changes you have made. SAVE ALL CHANGES IN MEMORY? NOTE: If you entered Setup Mode and did not make any changes, the above screen will still appear. In this case, press and release DISPLAY ADJUST NO to proceed to exit Setup Mode and begin the EAGLE 2’s startup sequence. 20. If you do not want to save the changes, press and release DISPLAY ADJUST NO. The unit will begin its startup sequence without saving the changes.
Parts List Table 17: PID EAGLE 2 Spare Parts Part Number Description 33-0560RK Electrode stack, 0 - 50 ppm, 2 pack 33-0561RK Electrode stack, 0 - 2,000 ppm, 2 pack 51-1500RK Replacement lamp, 0 - 50 ppm 51-1501RK Replacement lamp, 0 - 2,000 ppm 57-2005RK Adapter assembly, for PID sensor 61-0300RK-01 PID sensor, 0 - 50 ppm VOC 61-0300RK-02 PID sensor, 0 - 2,000 ppm VOC 80-0131RK-20 PID probe, particle filter removed 80-0605RK Teflon lined sample hose, 5 feet 80-0610RK Teflon lined sample
Appendix E: ESM-01 Toxic Sensors Overview The ESM-01 sensors are used to monitor levels of a variety of toxic gases. This appendix describes the EAGLE 2’s ESM-01 sensors and includes instructions to use an EAGLE 2 that has one or more ESM-01 sensors installed. It also includes instructions to replace an ESM-01 sensor. Table 18: ESM-01 Sensor Specifications Target Gas Detection Range (ppm) Reading Increment Alarm 1 Factory Setting Alarm 2 Factory Setting STEL (ppm) TWA (ppm) Ammonia (NH3) 0 - 75.
See “Replacing the ESM-01 Sensor or Changing Sensor Type” on page 199 for instructions to replace or change an ESM-01 sensor. Top Case Not Shown Charcoal Filter ESM-01 Sensor Pump Standard Flow Chamber CO Sensor Oxygen Sensor Oxygen Sensor Bracket H2S Sensor LEL Sensor and Sensor Bracket Figure 35: Typical ESM-01 Sensor Location The ESM-01 sensor is installed in a single sensor flow chamber which is located in the area next to the standard 4-sensor flow chamber.
ESM-01 Sub PCB The ESM-01 sub PCB is a circuit board that is installed on the main PCB in one of the 3 sub PCB positions when an ESM-01 sensor is used with the EAGLE 2. The ESM-01 sensor connects to the sub PCB with a 12-position connector. The sub PCB plugs into the main PCB and is held in place with a screw/flat washer/lock washer. There are no user serviceable parts on the ESM-01 sub PCB.
described below in “Calibrating with a 5-Gas Cylinder” on page 191, uses a multi-gas cylinder that contains the target gas for all installed sensors. Follow the set of instructions appropriate for your cylinder configuration. NOTE: If your instrument has more than one ESM-01 sensor, you will need a calibration cylinder for each sensor and follow the instructions in “Calibrating with a 4-Gas Cylinder and an ESM-01 Cylinder” on page 182.
6. Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT button. CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION 7. Press and release the POWER ENTER RESET button. The following screen appears. PERFORM AIR ADJUST? 8. Press and release the AIR ▲ YES button to continue. If you do not want to continue, press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen.
9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen. ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION 10. Install the demand flow regulator onto the multi-gas calibration cylinder. 11. Connect the sample tubing to the demand flow regulator. 12.
13. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14. Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4-gas calibration cylinder.
channel. The calibration gas value begins to flash. 18. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to adjust the calibration gas setting to the desired value. NOTE: The calibration gas value cannot be set lower than the low alarm setting. If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting, enter Setup Mode and change the low alarm setting. See “Updating the Alarm Point Settings” on page 115 for instructions.
changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again. 24. When you make your selection and press the desired button, the unit returns to the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL 25. Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing. CAL IN PROCESS CH4 0 %LEL OXY 20.9 vol% H2S 0.
28. If all channels passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CAL GAS VALUE NH3 25 ppm ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and lists the sensor(s) that failed to calibrate. In the example below, the oxygen and H2S channels failed calibration. The other sensors calibrated normally.
31. If you want to change the ESM-01 channel’s calibration gas value, follow Step 15 - Step 24 above beginning with the ESM-01 Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen. CAL GAS VALUE NH3 25 ppm ENTER TO BEGIN CAL 32. With the ESM-01 Calibration Gas Value Screen displayed, press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the ESM-01 channel with CAL IN PROCESS flashing.
36. If the ESM-01 channel passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the ESM-01 channel cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and indicates that the ESM-01 sensor failed to calibrate. FAILED SENSOR ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern.
Calibrating with a 5-Gas Cylinder The calibration instructions below show a 5 channel instrument which has the four standard channels, LEL/oxygen/H2S/CO, and an ESM-01 channel for sulfur dioxide (SO2). 1. See “Calibration Supplies and Equipment” on page 56 for a description of the necessary calibration supplies. See Table 19 on page 202 for available cylinders. Make sure your calibration cylinder is appropriate for the ESM-01 detection range. 2. Find a fresh-air environment.
7. Press and release the POWER ENTER RESET button. The following screen appears. PERFORM AIR ADJUST? 8. Press and release the AIR ▲ YES button to continue. If you do not want to continue, press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen. 9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen.
the probe are tightened firmly together to avoid leaks that can affect the calibration. See Figure 19 on page 76 for an illustration of the internal parts of the probe. 13. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14. Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.
16. Place the cursor next to the channel whose gas value you want to change using the AIR ▲ YES and RANGE ▼ SHIFT buttons. 17. Press and release the POWER ENTER RESET button to select the channel. The calibration gas value begins to flash. 18. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to adjust the calibration gas setting to the desired value. NOTE: The calibration gas value cannot be set lower than the low alarm setting.
If you select NO by pressing and releasing the DISPLAY ADJUST NO button, the changes you made will be used for any calibrations performed during the current operating session only. The EAGLE 2 will delete the changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again. 24. When you make your selection and press the desired button, the unit returns to the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.
28. If all channels passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CAL GAS VALUE SO2 5 ppm ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and lists the sensor(s) that failed to calibrate. In the example below, the oxygen and H2S channels failed calibration. The other sensors calibrated normally.
29. If you want to change the ESM-01 channel’s calibration gas value, follow Step 15 - Step 24 above beginning with the ESM-01 Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen. CAL GAS VALUE SO2 5 ppm ENTER TO BEGIN CAL 30. With the ESM-01 Calibration Gas Value Screen displayed, press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the ESM-01 channel with CAL IN PROCESS flashing.
33. If the ESM-01 channel passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the ESM-01 channel cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and indicates that the ESM-01 sensor failed to calibrate. FAILED SENSOR ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern.
Maintenance The SO2 and HCN ESM-01 sensors are the only ESM-01 sensors that include user serviceable parts. This section includes a procedure for replacing an ESM-01 sensor and for replacing the H2S scrubber in the SO2 and HCN sensors. When replacing a sensor, you may either replace it with another of the same sensor or you may install a different ESM-01 sensor. If a different one is installed, the EAGLE 2 will load the sensor parameters and configure the ESM-01 channel for the new sensor.
the ESM-01 chamber. 14. Install the two sensor bracket screws. 15. Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case. If the main PCB is not seated properly, it may be damaged when the top case is re-installed. 16. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case.
in the sensor body by a threaded collar on the connector end of the sensor. Unscrew the collar from the sensor body. Sensor Body Groove H2S Scrubber Electrolyte Reservoir Assembly Collar Figure 36: ESM-01 Sensor Component Location 12. Grasp the connector and lift the electrolyte reservoir assembly out of the sensor body. 13. Turn the body upside down to remove the H2S scrubber. If it does not come out, push on it from the top of the upside down body. 14. Discard the old H2S scrubber. 15.
the ESM-01 chamber. 21. Install the two sensor bracket screws. 22. Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case. If the main PCB is not seated properly, it may be damaged when the top case is re-installed. 23. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case.
Part Number Description 81-0185RK-02 Calibration cylinder, 0.
Appendix F: TC Sensors Overview The TC sensors are used to monitor high levels of combustible gas. This appendix describes the EAGLE 2’s TC sensor and includes instructions to use an EAGLE 2 that has a TC sensor installed. It also includes instructions to calibrate and replace a TC sensor. Table 20: TC Sensor Specifications Target Gas Detection Range Reading Increment Alarm 1 Factory Setting Alarm 2 Factory Setting STEL TWA Methane (CH4) 0 - 100.0 %vol 0.
Top Case Not Shown TC Sensor Charcoal Filter Standard Flow Chamber CO Sensor Pump Oxygen Sensor Oxygen Sensor Bracket H2S Sensor LEL Sensor and Sensor Bracket Figure 37: Typical TC Sensor Location TC Sensor The TC sensor’s appearance is exactly the same as that of the LEL sensor. Its housing includes a sintered metal flame arrestor on one end that allows gas to diffuse into the sensor. On the other end, five pins extend from the sensor.
TC sensor cable connects to the sub PCB. The sub PCB plugs into the main PCB and is held in place with a screw/flat washer/lock washer. There are no user serviceable parts on the TC sub PCB. Start Up and Normal Operation For instructions to startup and use an EAGLE 2 that includes a TC sensor, reference “Start Up” on page 22, “Measuring Mode, Normal Operation” on page 29, and “Measuring Mode, Alarms” on page 35.
1. Use the DISPLAY ADJUST NO button to enter Display Mode and scroll to the Catalytic (LEL) Sensor screen. The current setting will be flashing. The screen below indicates the warning that appears when a TC sensor and a catalytic LEL sensor are both installed. Since the TC sensor only reads in %volume, if the catalytic LEL sensor is disabled, there will be no alarms for the LEL range. CAT (LEL) SENSOR * * * WARNING * * * NO LEL ALARMS IF CAT (LEL) IS DISABLED ENABLED 2.
appropriate for the TC detection range. If the EAGLE 2 is intended for use in a landfill, RKI Instruments, Inc. recommends using the carbon dioxide/methane mix calibration cylinder because it is representative of gases present in a landfill. NOTE: If your instrument has more than one TC sensor, you will need a calibration cylinder for each sensor. 2. Find a fresh-air environment. This is an environment free of toxic or combustible gases and of normal oxygen content (20.9%). 3.
7. Press and release the POWER ENTER RESET button. The following screen appears. PERFORM AIR ADJUST? 8. Press and release the AIR ▲ YES button to continue. If you do not want to continue, press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen. 9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen.
complete with internal O-ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration. See Figure 19, “Replacing the Particle Filter and Hydrophobic Filter Disk” on page 76 for an illustration of the internal parts of the probe. 13. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14.
15. To adjust the values on the screen, press and hold the RANGE ▼ SHIFT button, then press the DISPLAY ADJUST NO button and release both. The following screen appears with the cursor next to CH4. ADJUST AUTO CALIBRATION VALUES > CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm 16. Place the cursor next to the channel whose gas value you want to change using the AIR ▲ YES and RANGE ▼ SHIFT buttons. 17. Press and release the POWER ENTER RESET button to select the channel.
22. Press and release the POWER ENTER RESET button. The following screen appears. DO YOU WANT TO STORE NEW VALUE(S) IN MEMORY FOR FUTURE CALIBRATIONS? PRESS YES OR NO 23. If you select YES by pressing and releasing the AIR ▲ YES button, the changes that you made will be saved in the EAGLE 2’s memory as the new auto calibration gas values.
27. Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the programmed values. 28. If all channels passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CAL GAS VALUE CH4 50.0 %vol. ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and lists the sensor(s) that failed to calibrate.
31. If you want to change the TC channel’s calibration gas value, follow Step 15 - Step 24 above beginning with the TC Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen. CAL GAS VALUE CH4 50.0 %vol. ENTER TO BEGIN CAL 32. With the TC Calibration Gas Value Screen displayed, press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the TC channel with CAL IN PROCESS flashing. CAL IN PROCESS CH4 0.0 %vol.
36. If the TC channel passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the TC channel cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and indicates that the TC sensor failed to calibrate. FAILED SENSOR ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern.
Maintenance The TC sensor does not include any user serviceable parts. This section includes a procedure for replacing the TC sensor. Replacing the TC Sensor 1. Verify that the EAGLE 2 is off. 2. Place the EAGLE 2 upside down on a flat surface or hold it upside down. 3. Unscrew the three case screws until they disengage from the top case. They are captive screws so they will not fall off of the bottom case. 4. Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case.
properly, it may be damaged when the top case is reinstalled. 16. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case. If you have any difficulty mating the top and bottom cases, inspect the placement of the main PCB and the placement of the top case gasket. 17. Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case. 18. Calibrate the TC channel as described in “TC Calibration” on page 207.
4. The “S” will then disappear and the following screen will appear for a few seconds. GAS MONITOR EAGLE 2 5. If the unit prompts you for the password, enter it by using AIR ▲ YES and RANGE ▼ SHIFT to select each password number and then pressing and releasing the POWER ENTER RESET button to enter it and move on to the next number until all of the numbers are entered. The main menu displays. It displays six menu items at a time.
8. Use RANGE ▼ SHIFT to move the cursor down the menu to the TC sensor. CONFIGURE GASES CAT : CH4 (CAT) > OP1 : CH4 (TC) OP2 : --- (---) OP3 : --- (---) 9. To change the TC sensor gas configuration, press and release POWER ENTER RESET. 10. A screen appears that indicates the detection range for the currently configured gas. In the example below, the TC sensor is currently configured for 0 - 100 %volume CH4. The screen also shows the gas configuration choices for the TC sensor.
WARNING: Do not use the user defined gas configuration without consulting RKI Instruments, Inc. 14. Enter the gas name. There are 3 characters available for the gas name. The factory setting of TC uses only 2 characters. Use AIR ▲ YES and RANGE ▼ SHIFT to display the desired character, then press POWER ENTER RESET to enter the displayed character and move to the next character. Repeat until all three characters are entered. When the last character is entered, the full scale value will be flashing. 15.
21. Press and release POWER ENTER RESET. 22. A screen appears that asks if you want to save the changes you have made. SAVE ALL CHANGES IN MEMORY? NOTE: If you entered Setup Mode and did not make any changes, the above screen will still appear. In this case, press and release DISPLAY ADJUST NO to proceed to exit Setup Mode and begin the EAGLE 2’s startup sequence. 23. If you do not want to save the changes, press and release DISPLAY ADJUST NO.
Parts List Table 21: TC EAGLE 2 Parts List Part Number Description 47-5010RK TC/LEL sensor cable TE-7568 TC sensor 81-0013RK-01 Calibration cylinder, 50% vol CH4 in N2, 34 liter steel 81-0013RK-05 Calibration cylinder, 50% vol CH4 in N2, 58 liter steel 81-0023RK-01 Calibration cylinder, H2, 8% volume in nitrogen, 34 liter steel 81-0024RK-01 Calibration cylinder, H2, 100% volume, 34 liter steel 81-0025RK-01 Calibration cylinder, 35% CO2/50% CH4, balance N2, 34 liter steel, intended for landfil
Appendix G: Infrared Carbon Dioxide Sensors Overview The infrared CO2 sensors are used to monitor levels of carbon dioxide. This appendix describes the EAGLE 2’s infrared CO2 sensors and includes instructions to use an EAGLE 2 that has one or more infrared CO2 sensors installed. It also includes instructions to replace an infrared CO2 sensor. Table 22: Infrared CO2 Sensor Specifications Range Increment Alarm 1 Factory Setting Alarm 2 Factory Setting STEL (ppm) TWA (ppm) 0-5.00 %vol 0.02 %vol 0.
Top Case Not Shown CO Sensor Infrared CO2 Sensor Charcoal Filter Standard Flow Chamber Oxygen Sensor Bracket Pump H2S Sensor Oxygen Sensor CO2 Scrubber LEL Sensor and Sensor Bracket Figure 38: Typical Infrared CO2 Sensor Location The infrared CO2 sensor is installed in a single sensor flow chamber which is located in the area next to the standard 4-sensor flow chamber. This area can accommodate up to three single sensor flow chambers.
one of the 3 sub PCB positions when an infrared CO2 sensor is used with the EAGLE 2. The infrared CO2 sensor connects to the sub PCB with a 12position connector. The sub PCB plugs into the main PCB and is held in place with a screw/flat washer/lock washer. There are no user serviceable parts on the infrared sub PCB. CO2 Scrubber A carbon dioxide scrubber is mounted to the exterior side of EAGLE 2s that are factory-shipped with carbon dioxide sensors for the ranges of 0-5 %vol and 0-10,000 ppm.
Start Up and Normal Operation For instructions to startup and use an EAGLE 2 that includes an infrared CO2 sensor, reference “Start Up” on page 22, “Measuring Mode, Normal Operation” on page 29, and “Measuring Mode, Alarms” on page 35. Follow these instructions keeping the following special considerations in mind: • A background level of CO2 exists in fresh air. The low range sensors will display a reading in fresh air. Table 23 below indicates typical gas readings in fresh air.
5. Put the caps back on the scrubber. Infrared CO2 Calibration An infrared CO2 channel can be calibrated using the auto calibration method or the single calibration method. To calibrate an infrared CO2 channel using the single calibration method, see “Calibrating Using the Single Calibration Method” on page 64 and follow the instructions for calibrating a single channel. If your instrument is a multi-channel instrument that includes one or more infrared CO2 channels, RKI Instruments, Inc.
5. The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 6. Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT button. CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION 7. Remove the black caps from the ends of the carbon dioxide scrubber.
11. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen. ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION 12. Remove the CO2 scrubber from the EAGLE 2 inlet fitting. 13. Put the caps back on the scrubber. 14. Install the demand flow regulator onto the multi-gas calibration cylinder.
17. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 18. Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4-gas calibration cylinder.
channel. The calibration gas value begins to flash. 22. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to adjust the calibration gas setting to the desired value. NOTE: The calibration gas value cannot be set lower than the low alarm setting. If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting, enter Setup Mode and change the low alarm setting. See “Updating the Alarm Point Settings” on page 115 for instructions.
changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again. 28. When you make your selection and press the desired button, the unit returns to the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL 29. Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing. CAL IN PROCESS CH4 0 %LEL OXY 20.9 vol% H2S 0.
32. If all channels passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CAL GAS VALUE CO2 2.50 vol% ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and lists the sensor(s) that failed to calibrate. In the example below, the oxygen and H2S channels failed calibration. The other sensors calibrated normally.
35. If you want to change the infrared CO2 channel’s calibration gas value, follow Step 17 - Step 26 above beginning with the infrared CO2 Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen. CAL GAS VALUE CO2 2.50 vol% ENTER TO BEGIN CAL 36. With the infrared CO2 Calibration Gas Value Screen displayed, press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the infrared CO2 channel with CAL IN PROCESS flashing.
adjustment for the infrared CO2 channel to the programmed value. 40. If the infrared CO2 channel passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the infrared CO2 channel cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and indicates that the infrared CO2 sensor failed to calibrate.
Maintenance This section includes a procedure for replacing an infrared CO2 sensor. Replacing the IR CO2 Sensor or Changing Sensor Type 1. Verify that the EAGLE 2 is off. 2. Place the EAGLE 2 upside down on a flat surface or hold it upside down. 3. Unscrew the three case screws until they disengage from the top case. They are captive screws so they will not fall off of the bottom case. 4. Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case.
properly, it may be damaged when the top case is re-installed. 16. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case. If you have any difficulty mating the top and bottom cases, inspect the placement of the main PCB and the placement of the top case gasket. 17. Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case. 18.
Appendix H: Infrared Methane Sensor Overview This appendix describes the EAGLE 2’s infrared methane sensors and includes instructions to use an EAGLE 2 that has an infrared methane sensor installed. It also includes instructions to replace an infrared methane sensor.
• propane • toluene The infrared methane sensor is known to not or to poorly respond to the following combustible gases: • acetylene • hydrogen • styrene Description Table 26 and Table 27 above list the available infrared methane sensor configurations. The infrared methane sensor is a smart sensor that stores the target gas.
other two sensor chamber locations depending on the particular version of the EAGLE 2. Some infrared methane instrument configurations do not include the 4-sensor flow chamber. Infrared Methane Sensor The infrared methane sensor is a cylindrical sensor with a diffusion opening on the front and a 12 pin circular connector on the back. A 12 wire cable plugs into the back of the infrared methane sensor with a circular connector that includes a locking lever.
sensor has its own protective shut off, exposure to high levels of combustible gas can still stress the catalytic LEL sensor. The catalytic LEL sensor can be enabled or disabled in the Catalytic (LEL) Sensor screen in Display Mode. The default setting is enabled. To change the setting, do the following: NOTE: The Catalytic (LEL) Sensor setting is reset when the EAGLE 2 is turned off. When the EAGLE 2 is turned on, this setting is always ENABLED. 1.
maximum accuracy in the %vol range, the sensor may be calibrated to 50 %vol. To use the auto calibration method to calibrate a multi-channel instrument that includes an infrared methane channel, do the following: 1. See “Calibration Supplies and Equipment” on page 56 for a description of the necessary calibration supplies. In addition to an appropriate multigas cylinder that is used to calibrate any active standard channels, you will also need a cylinder to calibrate the infrared methane channel.
7. Press and release the POWER ENTER RESET button. The following screen appears. PERFORM AIR ADJUST? 8. Press and release the AIR ▲ YES button to continue. If you do not want to continue, press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen. 9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen.
the probe are tightened firmly together to avoid leaks that can affect the calibration. See Figure 19, “Replacing the Particle Filter and Hydrophobic Filter Disk” on page 76 for an illustration of the internal parts of the probe. 13. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14. Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen.
15. To adjust the values on the screen, press and hold the RANGE ▼ SHIFT button, then press the DISPLAY ADJUST NO button and release both. The following screen appears with the cursor next to CH4. ADJUST AUTO CALIBRATION VALUES > CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm 16. Place the cursor next to the channel whose gas value you want to change using the AIR ▲ YES and RANGE ▼ SHIFT buttons. 17. Press and release the POWER ENTER RESET button to select the channel.
22. Press and release the POWER ENTER RESET button. The following screen appears. DO YOU WANT TO STORE NEW VALUE(S) IN MEMORY FOR FUTURE CALIBRATIONS? PRESS YES OR NO 23. If you select YES by pressing and releasing the AIR ▲ YES button, the changes that you made will be saved in the EAGLE 2’s memory as the new auto calibration gas values.
27. Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the programmed values. 28. If all channels passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CAL GAS VALUE CH4 50 %LEL ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and lists the sensor(s) that failed to calibrate.
31. If you want to change the infrared methane channel’s calibration gas value, follow Step 15 - Step 24 above beginning with the infrared methane Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen. CAL GAS VALUE CH4 50 %LEL ENTER TO BEGIN CAL 32.
36. If the infrared methane channel passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the infrared methane channel cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and indicates that the infrared methane sensor failed to calibrate.
Maintenance This section includes a procedure to replace an infrared methane sensor. Replacing the IR Methane Sensor 1. Verify that the EAGLE 2 is off. 2. Place the EAGLE 2 upside down on a flat surface or hold it upside down. 3. Unscrew the three case screws until they disengage from the top case. They are captive screws so they will not fall off of the bottom case. 4. Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case.
16. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case. If you have any difficulty mating the top and bottom cases, inspect the placement of the main PCB and the placement of the top case gasket. 17. Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case. 18. Calibrate the infrared methane channel as described in “Infrared Methane Calibration” on page 241.
Appendix I: Infrared Hydrocarbon Sensor Overview This appendix describes the EAGLE 2’s infrared hydrocarbon sensor and includes instructions to use an EAGLE 2 that has an infrared hydrocarbon sensor installed. It also includes instructions to replace an infrared hydrocarbon sensor.
Top Case Not Shown Charcoal Filter Infrared HC Sensor Pump Standard Flow Chamber CO Sensor Oxygen Sensor Oxygen Sensor Bracket H2S Sensor LEL Sensor and Sensor Bracket Figure 41: Typical Infrared Hydrocarbon Sensor Location Infrared Hydrocarbon Sensor The infrared hydrocarbon sensor is a cylindrical sensor with a diffusion opening on the front and a 12 pin circular connector on the back.
Start Up and Normal Operation For instructions to startup and use an EAGLE 2 that includes an infrared hydrocarbon sensor, reference “Start Up” on page 22, “Measuring Mode, Normal Operation” on page 29, and “Measuring Mode, Alarms” on page 35. 0-100 %LEL/2.0-30.0 %vol Autoranging The infrared hydrocarbon sensor can be factory set to detect gas in a 0-100 %LEL configuration or an autoranging configuration. The autoranging configuration detects gas on a 0-100 %LEL and a 2.0-30.0 %vol scale.
If set to DISABLED, the gas reading for the catalytic LEL channel will be replaced by dashes (---). 3. Press and release POWER ENTER RESET to save the setting and return to the main menu. 4. Use the DISPLAY ADJUST NO button to scroll through the rest of Display Mode and enter Normal Operation. Infrared Hydrocarbon Calibration An infrared hydrocarbon channel can be calibrated using the auto calibration method or the single calibration method.
5. The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 6. Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT button. CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION 7. Press and release the POWER ENTER RESET button. The following screen appears. PERFORM AIR ADJUST? 8.
9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds, then indicate that the operation is complete before returning to the Calibration Mode Screen. ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION > PERFORM AIR ADJUST NORMAL OPERATION 10. Install the demand flow regulator onto the multi-gas calibration cylinder. 11. Connect the sample tubing to the demand flow regulator. 12.
13. Move the cursor next to the AUTO CALIBRATION menu item by using the AIR ▲ YES button. CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14. Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4-gas calibration cylinder.
channel. The calibration gas value begins to flash. 18. Use the AIR ▲ YES and RANGE ▼ SHIFT buttons to adjust the calibration gas setting to the desired value. NOTE: The calibration gas value cannot be set lower than the low alarm setting. If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting, enter Setup Mode and change the low alarm setting. See “Updating the Alarm Point Settings” on page 115 for instructions.
changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again. 24. When you make your selection and press the desired button, the unit returns to the Calibration Gas Values Screen. CAL GAS VALUES CH4 50 %LEL OXY 12.0 vol% H2S 25.0 ppm CO 50 ppm ENTER TO BEGIN CAL 25. Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing. CAL IN PROCESS CH4 0 %LEL OXY 20.9 vol% H2S 0.
28. If all channels passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CAL GAS VALUE HC 50 %LEL ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and lists the sensor(s) that failed to calibrate. In the example below, the oxygen and H2S channels failed calibration. The other sensors calibrated normally.
31. If you want to change the infrared hydrocarbon channel’s calibration gas value, follow Step 15 - Step 24 above beginning with the infrared hydrocarbon Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen. CAL GAS VALUE HC 50 %LEL ENTER TO BEGIN CAL 32.
36. If the infrared hydrocarbon channel passed calibration, the following screen sequence occurs. AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END CALIBRATION MODE > AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the infrared hydrocarbon channel cannot be adjusted to the proper value, a screen displays that indicates a calibration failure and indicates that the infrared hydrocarbon sensor failed to calibrate.
Maintenance This section includes a procedure to replace an infrared hydrocarbon sensor. Replacing the IR Hydrocarbon Sensor 1. Verify that the EAGLE 2 is off. 2. Place the EAGLE 2 upside down on a flat surface or hold it upside down. 3. Unscrew the three case screws until they disengage from the top case. They are captive screws so they will not fall off of the bottom case. 4. Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case.
16. Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case. If you have any difficulty mating the top and bottom cases, inspect the placement of the main PCB and the placement of the top case gasket. 17. Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case. 18. Calibrate the infrared hydrocarbon channel as described in “Infrared Hydrocarbon Calibration” on page 255.
Appendix J: Methane Elimination Mode Overview For applications where methane is an interfering gas, you can set the EAGLE 2 to eliminate most response to methane using methane elimination mode. CAUTION: The EAGLE 2 catalytic combustible sensor experiences a significant gas response drop when it is changed from full response mode to methane elimination mode. The EAGLE 2 must be calibrated in both full response mode and in methane elimination mode to account for this.
6. You will note that if methane elimination is enabled, the Catalytic Sensor Relative Response screen no longer appears even if the relative response feature in Setup Mode is turned on. While methane elimination is enabled, the Catalytic Sensor Relative Response screen will not appear in Display Mode even if the relative response feature in Setup Mode is enabled. 7. Continue to press the DISPLAY button until the Measuring Mode screen appears. HEX 0%LEL OXY 20.9vol% H2S 0.
Target Gas LEL Factor PPM Factor Target Gas LEL Factor PPM Factor 1-Butyl Alcohol 1.71 2.18 Methyl Acrylate 0.83 2.11 Cyclohexane 1.06 1.25 Methyl Ethyl Keytone 1.04 1.32 Cumene 1.74 1.42 Methyl Isobutyl Keytone 1.30 1.42 Ethylene Dichloride 2.04 11.50 Mixed Xylenes 1.36 1.36 Ethyl Alcohol 0.57 1.71 Nonane 1.66 1.21 Ethyl Chloride 0.59 2.04 Pentane 0.54 0.74 Ethyl Acrylate 1.32 1.68 Propane Low response Low response Hexane 1.00 1.00 Styrene 1.74 1.
indicating that the unit is measuring in full response mode. 15. Allow 2 minutes for the combustible sensor to stabilize before monitoring the target gas. 16. Perform a demand zero. See “Performing a Demand Zero” on page 27 for instructions. 17. Monitor for the target gas. NOTE: The EAGLE 2 retains the methane elimination mode setting when it is turned off. The methane elimination mode setting, ENABLED or DISABLED, will remain in effect until it is changed in Display Mode.
Appendix K: Using the EAGLE 2 in Bar Hole Mode Overview This chapter explains how to operate the EAGLE 2 in Bar Hole Mode. Bar Hole Mode is used to perform consistent checks of bar holes when tracking down underground gas leaks. When the EAGLE 2 is in Bar Hole Mode, only the combustible and oxygen sensors are displayed.
NORMAL MODE and LEAK CHECK MODE will be the only choices displayed. In this case, see “Turning On the EAGLE 2” on page 22 or “Turning On the EAGLE 2, Leak Check Mode” on page 279. 1. Connect the sample hose to the EAGLE 2’s quick connect inlet fitting. 2. Connect the bar hole probe to the sample hose’s quick connect fitting. 3. Press and briefly hold down the POWER ENTER RESET button. Release the button when you hear a beep. 4. The LCD will show the following screen for about ten seconds.
NOTE: If no button is pressed for 20 seconds, the unit will proceed into whichever mode has the cursor in front of it. 9. The Active Gases Screen appears for a few seconds indicating which channels are active and their target gas. ACTIVE GASES CH4 H2S CH4 OXY CO 10. The gas alarm setpoints are displayed by three screens in sequence: the Low Alarm Screen, High Alarm Screen, and STEL/TWA Alarm Screen. Each screen remains on the LCD for three seconds. A CH4 10 %LEL L L OXY 19.5 vol% O A H2S 10.
• If the unit is due for calibration and CAL PAST DUE ACT is set to CONFIRM TO CAL, then the following screen displays and the buzzer sounds in a double pulsing pattern. CALIBRATION DATE IS PAST DUE PERFORM CALIBRATION? To perform a calibration, press and release the AIR ▲ YES button. The EAGLE 2 will enter Calibration Mode and the LCD will show the Calibration Mode main menu. See “Chapter 4: Calibration Mode” on page 55 for instructions to calibrate the EAGLE 2.
NOTIFICATION ONLY, then the following alert screen displays and the buzzer sounds in a double pulsing pattern. CALIBRATION DATE IS PAST DUE Press and release the POWER ENTER RESET button to acknowledge the alert and continue with the startup sequence. 12. The Date/Time Screen appears for a few seconds. 9/12/2008 15:00:00 13. If USER/STATION ID is turned on (see “Turning the User/Station ID Function On or Off” on page 118), the ID Screen appears for a few seconds.
If one of the combustible sensors, LEL or TC, fails, it is not possible to enter Bar Hole Mode. Press and release the POWER ENTER RESET button to acknowledge the failure and return to the Mode Select Screen. Replace the failed sensor as soon as possible. If the oxygen sensor fails, press and release the POWER ENTER RESET button to acknowledge the failure and continue to Bar Hole Mode. The gas reading for the oxygen sensor will be replaced by “XXX”. Replace the failed sensor as soon as possible.
3. Press and hold the AIR ▲ YES button. The display will indicate that a demand zero is taking place and prompt you to hold the AIR ▲ YES button. 4. Continue to hold the AIR ▲ YES button until the display prompts you to release it. The EAGLE 2 will set the fresh air reading for the CH4 and oxygen channels. Start up is complete and the unit is now ready for bar hole testing.
3. Insert the probe into the bar hole and press and release the ENTER button. The pump will turn on, the display will indicate “MEASURING...” below the oxygen reading, and the sample period will begin with the sample period counting down in seconds in the lower right corner of the display. The CH4 channel will be displayed in vol%. BAR HOLE MODE CH4 0 vol% OXY 20.9 vol% MEASURING... 30 SEC 4.
• Press and release the DISPLAY ADJUST NO button. The display will now indicate “FRESH AIR PURGE . . .” below the oxygen reading and the purge time will begin counting down from 30 seconds in the lower right corner of the display. BAR HOLE MODE CH4 0 vol% OXY 20.9 vol% FRESH AIR PURGE... 30 SEC • After 15 seconds of sampling, if the CH4 reading is less than 5 vol%, the CH4 channel will automatically begin displaying in the units you selected earlier.
Appendix L: Using the EAGLE 2 in Leak Check Mode Overview This chapter explains how to operate the EAGLE 2 in Leak Check Mode. Leak Check Mode is used to pinpoint small leaks of combustible gas from valves, flanges, connections, and other potential leak points. When the EAGLE 2 is in Leak Check Mode, only the catalytic combustible sensor is active.
4. The LCD will show the following screen for about ten seconds. GAS MONITOR EAGLE 2 5. The Battery Voltage Screen appears for a few seconds. BATTERY MIN: 4.3 VOLTS BATTERY NOW: 5.2 VOLTS 6. The Mode Select Screen displays. > NORMAL MODE BAR HOLE MODE LEAK CHECK MODE 7. The cursor will flash in front of NORMAL MODE. Use the RANGE ▼ SHIFT button to move the cursor next to LEAK CHECK MODE. 8.
10. The gas alarm setpoints are displayed by three screens in sequence: the Low Alarm Screen, High Alarm Screen, and STEL/TWA Alarm Screen. Each screen remains on the LCD for three seconds. A CH4 50 %LEL HL OXY 23.5 vol% I A H2S 30.0 ppm GR CO 50 ppm HM CH4 50 vol% S A CH4 10 %LEL L L OXY 19.5 vol% O A H2S 10.0 ppm WR CO 25 ppm M CH4 10 vol% S ALARMS STEL & TWA H2S(ppm) 15.0 10.0 CO (ppm) 200 25 CH4(vol%) OFF OFF 11.
done with the calibration and exit Calibration Mode, the unit will begin the startup sequence. If the calibration was successful, the screen above will not appear again until the unit is due for calibration. If the calibration was not successful, the screen above will again appear in the startup sequence. To continue without performing a calibration, press and release the DISPLAY ADJUST NO button.
12. The Date/Time Screen appears for a few seconds. 9/12/2008 15:00:00 13. If USER/STATION ID is turned on (see “Turning the User/Station ID Function On or Off” on page 118), the ID Screen appears for a few seconds. USER ID MIKE STATION ID PUMP 1 SERIAL NUMBER E2A515 If USER/STATION ID is turned off, only the serial number is shown. 14.
15. The EAGLE 2 is now operating in Leak Check Mode. The pump is on and the following screen appears. LEAK CHECK MODE CH4 0 [ BUZZER ON 0 ppm 5000 ] NOTE: For maximum sensor stability, allow 3-5 minutes for the sensor to warm up. The small increment size in the lower range of a ppm measurement can cause instability if the unit is not properly warmed up.
• 250 ppm increments from 10,000 ppm to 50,000 ppm As the gas concentration increases from 0 ppm, the alarm LEDs begin to blink in unison with the buzzer’s pulsing. The blinking/pulsing rate increases as the gas reading increases. If desired the buzzer can be turned off in Leak Check Mode. In a low-light environment, press and release any of the buttons to turn on the display backlight. See “Updating the Backlight Delay Setting” on page 119 to program backlight duration.
Turning Off the EAGLE 2, Leak Check Mode 1. Press and hold the POWER button. 2. The buzzer will sound and the LCD back light will flash for about five seconds. 3. Release the button when GOODBYE appears on the display. When GOODBYE disappears and the backlight turns off, the unit is off.
Appendix M: Tank Tester Model The EAGLE 2 Tank Tester model is intended for checking tanks or vessels that may contain residual hydrocarbon vapors or water or may have been purged of oxygen. It is supplied as an LEL (catalytic) only unit or an LEL (catalytic)/oxygen unit. You can also use this model as a standard EAGLE 2 gas monitor by connecting the standard hose and probe and selection Normal Mode in the Inert Mode Selection Screen.
Float Probe Assembly The float probe assembly helps prevent liquid from being drawn into the EAGLE 2. The float probe assembly is 12-feet long. The hose between both ends of the float probe has an integral cable that connects a switch in the float end to the plug at the other end. This plug connects to the float probe socket that is adjacent to the EAGLE 2’s IR port. A quick connect fitting at the same end of the assembly connects to the EAGLE 2’s inlet fitting.
assembly and must be installed on the EAGLE 2 inlet fitting before installing the float probe assembly when measuring gas in an inert atmosphere. To attach the dilution fitting: 1. Attach the dilution fitting’s male quick connect fitting to the EAGLE 2’s inlet fitting. 2. Attach the float probe assembly to the opposite end of the dilution fitting. NOTE: When using the dilution fitting, multiply the combustible gas reading (LEL or PPM) by 2 to determine the actual combustible gas concentration.
4. The Battery Voltage Screen appears for a few seconds. BATTERY MIN: 4.3 VOLTS BATTERY NOW: 5.2 VOLTS 5. The Inert Mode Selection Screen will then be displayed. MODE SELECT UP/DOWN THEN ENTER NORMAL MODE Use the AIR ▲ YES or RANGE ▼ SHIFT button to display the mode you wish to enter. To use the EAGLE 2 for tank tester applications, ensure that the float probe is connected and select Inert Mode.
If the instrument is sampling a fresh air environment, the oxygen channel will be in alarm. For more information about Inert Mode, see “Appendix N: Using the EAGLE 2 in Inert Mode” on page 293. 8. When monitoring for combustible gas in an inerted tank, be sure the dilution fitting is installed and multiply the combustible gas reading by 2. When monitoring for oxygen in an inerted tank, remove the dilution fitting or seal the dilution hole with your finger.
Parts List Table 35 lists part numbers for replacement parts and accessories of the EAGLE 2’s Tank Tester model.
Appendix N: Using the EAGLE 2 in Inert Mode Inert Mode is used to measure the combustible gas and/or oxygen level in a purged environment. The oxygen alarms in this mode are both increasing and are generally set at 5.0% and 10.0%. Description Inert Mode is factory activated for instruments that require it. The instrument can still be used in Normal Mode for other applications.
3. The LCD will show the following screen for a few seconds with the “S” in the lower right corner indicating the unit is entering Setup Mode. GAS MONITOR EAGLE 2 S 4. The “S” will then disappear and the following screen will appear for a few seconds. GAS MONITOR EAGLE 2 5.
8. Move the cursor next to the oxygen channel. Press and release POWER ENTER RESET. 9. A mode select screen will appear and prompt you to choose between Normal Mode and Inert Mode. Use the AIR ▲ YES or RANGE ▼ SHIFT button to display Inert Mode and press and release POWER ENTER RESET. MODE SELECT UP/DOWN THEN ENTER NORMAL MODE NOTE: To change the Normal Mode oxygen other alarm points, select Normal Mode and see “Updating the Alarm Point Settings” on page 115. OXY 0- 40.0 vol% INERT ALARM >LO : 5.
13. When you are done making changes, use RANGE ▼ SHIFT to move the cursor next to END. 14. Press and release POWER ENTER RESET to save the new settings and return to the Change Alarm Point Settings Screen. 15. Use RANGE ▼ SHIFT to move the cursor next to END. 16. Press and release POWER ENTER RESET to return to the main menu. Start Up 1. Press and briefly hold down the POWER ENTER RESET button. Release the button when you hear a beep. 2. The LCD will show the following screen for about ten seconds.
WARNING: The EAGLE 2 is not a gas monitoring device until you select a mode and enter it. 5. If you select Normal Mode, the warm up sequence will continue as described in “Start Up” on page 22 unless Leak Check or Bar Hole Mode is active. If Leak Check or Bar Hole Mode is active, the Leak Check/Bar Hole Mode Select Screen will appear immediately after the Normal Mode selection. > NORMAL MODE BAR HOLE MODE LEAK CHECK MODE Use the AIR ▲ YES or RANGE ▼ SHIFT button to display the mode you wish to enter. 6.
Operation See “Measuring Mode, Normal Operation” on page 29 for operating instructions keeping in mind that the Inert Mode oxygen alarm settings are different. You can access the Inert Mode Selection Screen while in Inert Mode or Normal Mode by pressing and holding the RANGE ▼ SHIFT button. MODE SELECT UP/DOWN THEN ENTER NORMAL MODE Use the AIR ▲ YES or RANGE ▼ SHIFT button to display the mode you wish to enter and press and release the POWER ENTER RESET button.
Warranty RKI Instruments, Inc. warrants the EAGLE 2 sold by us to be free from defects in materials, workmanship, and performance for a period of two years from the date of shipment from RKI Instruments, Inc. This includes the instrument and the original sensors. Replacement parts are warranted for 1 year from the date of their shipment from RKI Instruments, Inc. Any parts found defective within their warranty period will be repaired or replaced, at our option, free of charge.