User Manual

45 • Beacon 200 Rig Monitor Operator’s Manual
Slow or No
Response/
Difficult or
Unable to
Calibrate
•Rig Monitor
responds slowly or
does not respond to
response test.
Unable to
successfully set the
fresh air or span
reading during
calibration.
Rig Monitor requires
frequent calibration.
Note: Under “normal”
circumstances, the
detector requires
calibration once every 3
months.
Some applications may
require a more frequent
calibration schedule.
The calibration
cylinder is low, out-
dated, or defective.
The LEL flame
arrestor is wet or
clogged with dirt or
other particulates.
The detector is
malfunctioning.
The main PCB is
malfunctioning.
1. Verify that the calibration cylinder
contains an adequate supply of a
fresh test sample.
2. Verify that the calibration gas
concentration is appropriate for the
detector. Zero emission air (0 %LEL)
is normally used for a zero adjustment
if the environment is suspect and 50%
LEL is normally used for a response
adjustment.
3. Check the LEL detector to determine
if the flame arrestor is wet or dirty.
Clean if necessary.
4. If the calibration/response difficulties
continue, replace the detector.
5. If the calibration/response difficulties
continue, contact RKI for further
instruction.
Table 9: Troubleshooting the Beacon 200 Rig Monitor (Continued)
Condition Symptom(s) Probable Causes Recommended Action