Hydrogen Sulfide Beacon 200 Rig Monitor Operator’s Manual Part Number: 71-0236RK Revision: B Released: 9/30/14 www.rkiinstruments.
WARNING Read and understand this instruction manual before operating instrument. Improper use of the rig monitor could result in bodily harm or death. Periodic calibration and maintenance of the rig monitor is essential for proper operation and correct readings. Please calibrate and maintain this rig monitor regularly! Frequency of calibration depends upon the type of use you have and the sensor types.
Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
Table of Contents Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 About the Beacon 200 Gas Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Specifications . . . . .
Chapter 6: Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Calibration Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Adjusting Strobe/Horn Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Replacing Components of the Rig Monitor . . . .
Chapter 1: Introduction Overview This chapter briefly describes the Beacon 200 Rig Monitor. This chapter also describes the Beacon 200 Rig Monitor Operator’s Manual (this document). Table 1 at the end of this chapter lists the specifications for the Beacon 200 Rig Monitor. About the Beacon 200 Rig Monitor The Beacon 200 Rig Monitor is a fixed-mounted, continuous-monitoring gas detection instrument. This gas monitor is capable of detecting gas at two locations.
About this Manual The Beacon 200 Gas Monitor Operator’s Manual uses the following conventions for notes, cautions, and warnings. NOTE: Describes additional or critical information. CAUTION: Describes potential damage to equipment. WARNING: Describes potential danger that can result in injury or death.
Specifications Table 1 lists specifications for the Beacon 200 Rig Monitor. Table 1: Beacon 200 Rig Monitor Specifications Description Specification Input Power 10.8 - 14.5 VDC, 1.2A VDC Construction (housing) Fiberglass/polyester with lexan window (NEMA 4X) Controller Dimensions 21.0 in. H x 8.5 in. W x 7.00 in. D (533 mm H x 216 mm W x 178 mm D) Weight 20 lbs. Environmental Conditions • • -20°C to 50°C (-4°F to 122°F) max.
Chapter 2: Description Overview This chapter describes external and internal components of the Beacon 200 Rig Monitor.
External Description This section describes the housing and all external components of the Beacon 200 Rig Monitor. For the purposes of this description, the housing door is considered the front of the monitor. Housing The Rig Monitor’s fiberglass housing is weather- and corrosion-resistant. It is suitable for installation where general purpose equipment is in use. The housing door is hinged on the left side and is secured by two latches on the right side.
H2S Detector Junction Boxes and Cables Each of the two H2S detectors consists of a junction box with a mounting bracket, the detector, and the calibration cup/splash guard. Mounting Bracket To Beacon 200 Cable Bushing Terminal Strip H2S Detector (Factory Wired) Calibration Cup/ Splash Guard 1 of 2 H2S Detectors Figure 2.
there is enough room to perform a calibration. A cover on the front of the junction box allows access to the interior of the junction box. The detector housing body protects the sensing components within the housing body. Use the removable cap near the bottom of the housing body to access the sensor for maintenance or replacement. The cap protects the sensor from damage and includes a flame arrestor which contains any sparks which may occur within the detector housing body.
Internal Description This section describes the internal components of the Rig Monitor. Display Printed Circuit Board (PCB) The display PCB is mounted to the power supply mounting plate which is in turn mounted to the main PCB. The power supply mounting plate and main PCB are described below. The display PCB includes the display, the status lights, and the program buttons.
Status Lights The Rig Monitor includes four status lights that indicate the current status of the monitor. The status lights are to the left and right of the display (see Figure 4). • Pilot Light. The pilot light is on when the Rig Monitor is receiving incoming power. • Fail Light. The fail light turns on when the Rig Monitor is experiencing a fail condition. A fail condition can be caused by a failure within the Rig Monitor or detector head(s) wired to the Rig Monitor.
Main PCB The main PCB is mounted inside the housing. The power supply mounting plate is mounted to the main PCB with four standoffs and the display PCB is mounted to the power supply mounting plate with four standoffs. The main PCB includes the terminal strips, relays, fuses, and power switch. Terminal Strips The Rig Monitor includes four terminal strips for external wiring connections. See “Wiring the Beacon 200 Rig Monitor” on page 16 for detailed wiring procedures.
• DC In Terminal Strip. The DC in terminal strip is a 3-point terminal strip located above the controller terminal strip (see Figure 1). It facilitates wiring from the 12 VDC battery. The 50 foot battery cable is factory wired to this terminal strip. Table 4 lists the function of each terminal.
NOTE: The alarm 2 channel relays may be set to operate as individual channel fail relays. See “Configure Channel Settings Menu” on page 29 for instructions. • Common relays. The three common relays, alarm 1, alarm 2, and fail, are to the left of the controller terminal strip (see Figure 1). These relays are common for both channels. For example, the alarm 1 common relay energizes if either channel 1 or channel 2 recognizes an alarm 1 condition.
Chapter 3: Installation and Start Up Overview This chapter describes procedures to mount the Beacon 200 Rig Monitor, make wiring connections to the monitor, and start up the monitor. WARNING: Perform all installation and start-up procedures in a “fresh air” environment (known to be free of combustible gas, toxic gas, and of normal oxygen content). The Beacon 200 Rig Monitor is not in operation as a gas monitoring system until the start-up procedure is complete.
8.50 7.00 4.83 8.0 Max Mounting Feet, 4X Ø .31 x .50 slot, 4X .41 FAIL AL AR M 2 6.45 10.50 PILOT AL AR M 1 10.94 BEACON 200 GAS MONITOR RESET Buzzer 3/4 Conduit Hubs (4X) Reset Switch 3/4 Cable Bushings (3X) 6.00 Figure 6. Rig Monitor Outline and Mounting Dimensions 1. Select the mounting site.
• Are the display screen and status lights visible? 2. Close and latch the housing door. 3. Position the monitor on a vertical surface at eye level (4 1/2 to 5 feet from the floor). 4. The Rig Monitor is shipped with the mounting feet positioned behind the housing. Loosen the screws that secure the feet to the housing, rotate the feet to their mounting position (as shown in Figure 6), then tighten the screws. 5. Insert 1/4 in. or 5/16 in.
Wiring the Beacon 200 Rig Monitor The detector leads, junction box connections, power cable wiring, and strobe/horn wiring are all factory wired.
WARNING: Make all connections to the Beacon 200 Rig Monitor before you connect the DC power source. Before you make any wiring adjustments, always verify that all power sources are not live. Connecting the DC Power Source Perform the following functions to connect the Rig Monitor to DC power. One end of the battery cable is already factory wired to the appropriate terminals at the Rig Monitor housing. Use the other end to connect a 12 VDC battery.
bushing. 4. Guide the wiring of the external alarm device through the new conduit hub on the bottom of the Rig Monitor housing. 5. Connect the leads from the external alarm device and power to the alarm terminals as shown in Figure 9. Beacon 200 Alarm Terminal Strip CH1-A1 NO NC C + (H) External Alarm Device - (N) External Power Source Figure 9. Typical External Alarm Wiring 6. Repeat steps 4 and 5 for additional external alarm devices.
5. Connect the wires form the recording device to the recorder output terminals as shown in Figure 10. Recording Device #1, 1Kohm Max Impedance Recording Device #2, 1Kohm Max Impedance + + + CH1 OUT + CH2 OUT Figure 10. Recorder Output Wiring Starting Up the Beacon 200 Rig Monitor Perform the following procedure to place the Beacon 200 Rig Monitor into normal operation. 1. Complete the mounting and wiring procedures described earlier in this chapter. 2.
Chapter 4: Operation Overview This chapter describes the Beacon 200 Rig Monitor in normal operation. This chapter also describes the Rig Monitor in alarm 1, alarm 2, and fail conditions and suggests response to these conditions. Normal Operation Normal operation is defined as follows: • The start-up procedure is complete. • The Rig Monitor is not indicating an alarm 1, alarm 2, or fail condition. • The Rig Monitor is not running the Channel Control & Setup or Calibration Programs.
Recorder Output Operation The output at the recorder output terminals is a 4 - 20 mA signal for each active channel that is proportional to the detection range of the channel. A channel that is set as CHANNEL NOT USED or CHANNEL DISABLED in the Channel Control & Setup Program (see “Chapter 5: Channel Control and Setup Program” on page 27) has an output of 0 mA.
Alarm Indications This section describes the Rig Monitor in alarm 1, alarm 2, and fail conditions and suggests response to these conditions. Table 5 below lists the alarm indications for each condition. NOTE: The Beacon 200 Rig Monitor includes alarm on and alarm off delay settings for each channel and level of gas alarm. The alarm indications described in this section operate according to the factory set delay settings. See “Configure Channel Settings Menu” on page 29 for all the factory settings.
NOTE: You can select normally energized (NE) or normally de-energized (NDE) channel relay settings in the Channel Control & Setup menu. The following sections describe the default setting for the channel relays which is NDE. Common alarm 1 and alarm 2 relays are factory-set as NDE, and the common fail relay is factory set as NE. The common relays’ NE/NDE settings are not user-adjustable.
Alarm 2 Condition This section describes the audible and visual indications for an alarm 2 condition and suggests response to an alarm 2 condition. Alarm 2 Condition Indications When the gas reading of an active channel reaches the alarm 2 setpoint, the Rig Monitor senses an alarm 2 condition. The Rig Monitor alerts you to an alarm 2 condition as follows: • The ALARM 2 light turns on. • The gas reading in alarm 2 condition continues to flash and alternates with the ALARM-2 messages.
• The display reading is -10% of full scale or lower. • The Rig Monitor or detector head is malfunctioning. When the Rig Monitor senses a fail condition, it alerts you as follows: • The FAIL light turns on. • The gas reading for the failing channel is replaced by the FAIL message. • The buzzer sounds a steady tone. • The common fail relay de-energizes.
Viewing & Resetting Min/Max Readings The Reset switch may be used to view and reset the minimum and maximum gas readings for the active channel(s). 1. While the Rig Monitor is in normal operation, press and hold the Reset switch button for 3 seconds. 2. The display will indicate MIN / MAX Display Press RESET when done viewing . . . for 5 seconds before displaying the minimum and maximum readings for the active channel(s).
Chapter 5: Channel Control and Setup Program Overview The Channel Control & Setup Program allows viewing of and changes to instrument setup parameters. It is accessed using the program buttons. The Channel Control & Setup Program includes three menus as described in Table 6.
Enable/Disable Channel(s) Menu 1. From normal operation, simultaneously press and hold the ESCAPE and ENTER buttons for approximately 5 seconds to enter the Channel Control & Setup Program. Release the buttons when the Control & Setup Program Proceed? [YES] or [NO] message appears on the display screen. 2. Press the UP/YES button to continue. 3. Press the UP/YES or DOWN/NO button until the 1) Enable/Disable Channel(s) message appears on the display screen, then press the ENTER button. 4.
Configure Channel Settings Menu This section describes how to view and change channel parameters for the installed gas channels. 1. Simultaneously press and hold the ESCAPE and ENTER buttons for approximately 5 seconds to enter the Channel Control & Setup Program. Release the buttons when the Control & Setup Program Proceed? [YES] or [NO] message appears on the display screen. 2. Press the UP/YES button to continue. 3.
10. Press the UP/YES button to save the configuration. The screen will then return to the Channel Control & Setup menu. 11. To view or change the Channel 2 settings, scroll to the Configure Channels Menu and repeat steps 4 through 10. 12. To exit the Channel Control & Setup menu, press ESCAPE to return to the screen which asks Control & Setup Program Proceed? [YES] or [NO]. 13. Press the DOWN/NO button to return to normal operation.
Table 8: Channel Setting Parameters (Continued) Parameter (Factory-Set Value) Description ALARM-2 OFF DELAY (0 sec) The amount of time the Rig Monitor delays turning off the alarm 2 circuit once an alarm 2 condition passes. ALARM-2 (activation) (INCREASING) Indicates if the alarm 2 circuit is activated by gas readings INCREASING or DECREASING to the ALARM-2 Level.
Chapter 6: Maintenance Overview This chapter describes use of the Calibration Program and corrective maintenance procedures for the Beacon 200 Rig Monitor. It includes a troubleshooting guide for problems you may encounter with the Beacon 200 Rig Monitor. Procedures to replace components of the Rig Monitor are at the end of this chapter. Calibration Program The Calibration Program is used to calibrate the Rig Monitor’s active channel(s).
ESCAPE Calibration Program Enter ENTER/ESCAPE Normal Operation ENTER Calibration Timeout Selection Calibrate Channel 1 Y/N? Span Channel 1 Y/N? Calibrate Channel 2 Y/N? (If Installed) Air Adjust Channel 1 Air Adjust Channel 2 (If Installed) Span Channel 2 Y/N? (If Installed) Select Cal. Gas Concentration for Channel 1 Select Cal. Gas Concentration for Channel 2 Apply Gas to Ch. 1 & Ch. 2 Detectors Press Enter to Adjust Span Figure 11.
NOTE: While in the Calibration Program, the alarm status of the Beacon 200 Rig Monitor will be locked in the state it was when the Calibration Program was entered. 2. Press the ENTER button to continue and display the Calibration Timeout setting. The Rig Monitor will remain in the Calibration Program for the amount of time indicated by the Calibration Time-out setting or until you exit the program. If necessary, adjust the setting using the UP/YES and DOWN/NO buttons.
4. Screw the fixed flow regulator into the zero air calibration cylinder. 5. Use the calibration kit sample tubing to connect the fixed flow regulator to the calibration cup/splash guard of the first detector. 6. Turn the regulator knob counterclockwise to open the regulator. 7. Allow zero air to flow for two minutes. 8. Turn the regulator knob clockwise to close the regulator. The Rig Monitor will continue to display the minimum gas response on the display and retain the response level in its memory. 9.
If you press the DOWN/NO button, the span of channel 1 will continue if you selected it or if you did not select channel 1, the unit will return to the first calibration program screen which asks if you want to continue or escape from the Calibration Program. 3. If you pressed the UP/YES button for either channel, the display will prompt you for the span gas concentration that will be used for the first selected channel. The following instructions assume you pressed the UP/YES button for each channel. 4.
If the Rig Monitor is able to successfully make the calibration adjustment, it will display the message Cal Passed for: Channel 1 Channel 2. It will then return to the first calibration program screen which asks if you want to continue or escape from the program. Press ESCAPE to return to normal operation.
4. Grasp the top and bottom of the strobe/horn and push up and out in order to remove the cover. Base Top Captive Screw Push Out Push Up Here Figure 12. Top Case Removal 5. Turn the cover over and locate the Audio Select switch at the top of the cover as shown in the figure below. Audio Select Switch 4 5 6 3 2 1 AUDIO SELECT CANDELLA S ELECT Strobe Brightness Switch (DO NOT ADJUST) Figure 13. Strobe/Horn Cover 6. The Audio Select switch can be set at any number between 1 and 6.
factory setting is 2. 7. Turn the Audio Select switch so that the selection arrow is pointing to the desired number. 8. Place the cover over the base and push in and down in order to reinstall the cover. Make sure that the cover is sealed to the base by the gasket. 9. Screw the captive screw at the bottom front of the strobe/horn back in. Replacing Components of the Rig Monitor This section describes how to replace the fuses, the plug-in H2S sensor, and the entire H2S detector. Replacing the Fuses 1.
10. Close and secure the housing door. Replacing the Plug-In H2S Sensor 1. Open the housing door of the Rig Monitor, then place the power switch in the OFF position. 2. Disconnect the battery clamps from the + and - terminals of the 12 VDC battery. 3. Unscrew the calibration cup/splash guard from the detector. 4. Unscrew the detector housing cap from the housing body. Make sure not to lose the cap gasket. See Figure 3 for an illustration of the H2S detector. 5.
VDC battery. 3. Remove the H2S detector junction box cover. 4. Disconnect the detector leads from the terminal block in the junction box. Note the position of the color-coded leads as you remove them. 5. Unscrew the calibration cup/splash guard from the detector housing cap. 6. Unscrew the detector from the junction box hub. 7. Guide the detector leads of the replacement detector through the junction box hub then screw the mounting threads of the detector into the hub. 8.
Monthly This procedure describes a test to verify that the H2S detectors respond properly to hydrogen sulfide. It describes a test using a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator. NOTE: Performing a response test on the H2S detectors may cause alarms. Be sure to put the Rig Monitor into its calibration program or disable external alarms before performing this test. Preparing for the response test 1.
NOTE: If the reading is not within ± 20% of the correct response reading, calibrate the detector as described in the Calibration section of this manual. 7. Turn the regulator’s on/off knob clockwise to close the regulator. 8. Detach the sample tubing from the calibration cup/splash guard. 9. Unscrew the regulator from the calibration cylinder. 10. When the Rig Monitor display reading falls below the alarm setpoints, return it to normal operation or enable alarms. 11.
Table 9: Troubleshooting the Beacon 200 Rig Monitor (Continued) Condition Flickering Display Symptom(s) • The display readings flicker often. Probable Causes • • • • Buzzer Not Working • • Reset Switch Not Working • • Strobe/Horn Not Working Fail Condition • • • The buzzer does not sound an audible alarm during alarm conditions. The buzzer sounds weak or broken. • The buzzer does not silence when you press the reset switch.
Table 9: Troubleshooting the Beacon 200 Rig Monitor (Continued) Condition Slow or No Response/ Difficult or Unable to Calibrate Symptom(s) • Rig Monitor responds slowly or does not respond to response test. • Unable to successfully set the fresh air or span reading during calibration. • Rig Monitor requires frequent calibration. Note: Under “normal” circumstances, the detector requires calibration once every 3 months. Some applications may require a more frequent calibration schedule.
Parts List Table 10 lists the part numbers and descriptions for replacement parts and accessories offered for the Beacon 200 Rig Monitor. Table 10: Parts List, Beacon 200 Rig Monitor Part No. Description 06-1248RK-03 Sample tubing for calibration, 3 foot length 07-0033RK Detector housing cap gasket 07-0203RK Rubber retaining boot 14-2101RK Spacer between H2S sensor and rubber boot 18-0061RK Cable bushing 18-0107RK Conduit Hub (3/4 in.