Beacon 110 Gas Monitor Operator’s Manual Part Number: 71-0110RK Revision: C Released: 1/17/12 RKI Instruments, Inc. www.rkiinstruments.
Product Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
Table of Contents Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 About the Beacon 110 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Specifications . . . .
Chapter 6: Input Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Selecting the Detector Head Input and Gas Type . . . . . . . . . . . . . . . . . . . . . . . 29 Chapter 7: Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Overview . . . . . . . . . . . . . . . . . . . . . . .
Chapter 1: Introduction Overview This chapter briefly describes the Beacon 110 Gas Monitor. This chapter also describes the Beacon 110 Operator’s Manual (this document). Table 1 at the end of this chapter lists the specifications for the Beacon 110. About the Beacon 110 The Beacon 110 is a fixed mount, single channel continuous-monitoring gas monitoring controller. All user adjustable parameters may be accessed using push button switches.
Specifications Table 1 lists specifications for the Beacon 110. Table 1: Beacon 110 Specifications Input Power 100/115/220V ~ ±10%, 50/60Hz, 0.2/0.2/0.1A or 24 V ± 10%, 0.5A VDC Construction (housing) Fiberglass/polyester with lexan window (NEMA 4X) Dimensions 8.5 in. H x 6.5 in. W x 4.5 in. D (216 mm H x 165 mm W x 108 mm D) Weight 4.2 lbs.
Chapter 2: Description Overview This chapter describes external and internal components of the Beacon 110 gas monitor. External Description This section describes the housing and all external components of the Beacon 110. For the purposes of this description, the housing door is considered the front of the monitor. Housing The Beacon 110’s fiberglass housing is weather- and corrosion-resistant. It is suitable for installation where general purpose equipment is in use.
Internal Description This section describes the internal components of the Beacon 110. Control PCB Main PCB Power Supply Relay (3X) DC Power Polyfuse Power Switch AC Fuses Detector/ Transmitter Terminal Strips AC IN Terminal Strip Control Terminal Strip Alarm Terminal Strip Buzzer Reset Switch Conduit Hubs (2X) Figure 1: Beacon 110 Component Location Main Printed Circuit Board (PCB) The main PCB is mounted inside the housing.
Terminal Strips The Beacon 110 includes terminal strips for external and factory wiring connections. See “Wiring the Beacon 110” on page 13 for detailed wiring procedures. • AC In Terminal Strip. The AC in terminal strip is a 3-point terminal strip located in the lower left corner of the main PCB. It facilitates wiring from the AC power source. Table 2 lists the function of each terminal.
Ground Stud The threaded ground stud is used for making connections to earth ground. It is connected through the main PCB to the G (ground) terminal on the AC in terminal strip. A kep nut on the stud may be removed for installation of one or more lugs to make wiring connections. This stud is typically used to connect the shield drain wire of shielded cable to earth ground at the Beacon 110. Relays The Beacon 110 includes three alarm relays located above the alarm terminal strip.
Alarm 1 LED F Fail LED A1 A2 Alarm 2 LED LCD Display Contrast Potentiometer UP/YES DOWN/NO ENTER Control Switches Figure 2: Control PCB Component Location LCD Display The LCD display is located at the top of the control PCB. It indicates the current gas reading and displays messages and parameters in the Beacon 110’s operating modes. Contrast Potentiometer The contrast potentiometer is located to the right of the LCD display. It is used to adjust the contrast of the LCD.
Table 4: Beacon 110 Control Button Functions Switch Function ENTER • • • Initiates operations Enters Configuration Mode (press with UP/YES button) Accepts displayed parameters Status LEDs The Beacon 110 includes three active status LEDs that are located above the display (see Figure 1). Two LEDs, labelled RX and TX, to the right of those described below, are not active. • Fail LED The fail LED turns on when the Beacon 110 is experiencing a fail condition.
Optional Accessories This section describes the optional accessories available for the Beacon 110. Both optional accessories are wired to the Strobe Terminal Strip as shown below.
Alarm Strobe The Beacon 110 can be ordered with a red alarm strobe light installed on the top of the housing. The Beacon 110 retains its NEMA 4X rating with the strobe installed. Strobe operation can be programmed in Configuration Mode (see “Chapter 5: Configuration Mode” on page 26). The outline and mounting dimensions of the Beacon 110 with the optional alarm strobe mounted are the same as the standard Beacon 110 without the strobe. See Figure 6 for outline and mounting dimensions.
NOTE: See “Chapter 3: Installation & Startup” on page 12 for complete Beacon 110 installation instructions. CAUTION: Do not adjust the strobe brightness or the horn volume at the strobe/horn, as this may overload the Beacon 110 strobe control circuit. The horn/strobe does not come factory installed to the Beacon 110. To install the horn/ strobe: 1. Mount the horn/strobe in the desired location. 2.
Chapter 3: Installation & Startup Overview This chapter describes procedures to mount the Beacon 110 Gas Monitor, make wiring connections to the monitor, and start up the monitor. WARNING: Perform all installation and start-up procedures in a “fresh air” environment (environment known to be free of combustible gas, toxic gas, and of normal oxygen content). The Beacon 110 is not in operation as a gas monitoring system until the start-up procedure is complete.
• Is an AC or DC power source available? • Is there enough room to open the housing door and make wiring connections through the conduit hubs at the bottom of the housing? • Are the display screen and status lights visible? 2. If necessary, close and latch the housing door. 3. Position the monitor on a vertical surface at eye level (4 1/2 to 5 feet from the floor). 4. Insert 1/4 in. or 5/16 in.
CAUTION: Do not route power and detector head wiring through the same conduit hub. The power wiring may disrupt the transmission of the detector head signal to the monitor. 6. Connect the AC wires to the AC power terminal strip as shown in Figure 7 below.
6. Connect the DC wires to the controller terminal strip as shown in Figure 8. EXT DC (24V BATT) + - External DC In Terminals on Controller Terminal Strip 24 VDC + - Figure 8: DC Power Wiring NOTE: The Beacon 110 will operate from the DC input down to 18.5 volts. If a self contained backup battery is used, see its operator’s manual for a description of its recharging characteristics.
6. Connect the leads from the external alarm to the appropriate terminals on the alarm terminal strip as shown in Figure 9 to obtain the desired operation. FAIL Beacon 110 Alarm Terminal Strip ALARM-1 ALARM-2 NO NC C NO NC C NO NC C (+) H (-) N External Power Source External Alarm Device Figure 9: Typical Alarm Device Wiring 7. Repeat steps 5 and 6 for additional alarm devices. Connecting Recorders Perform the following procedure to connect an analog signal recording device to the Beacon 110.
Connecting RKI Detector Heads Perform the following procedure to connect an RKI detector head to the Beacon 110. 1. Turn off or unplug power to the Beacon 110. 2. Open the Beacon 110 door and place the power switch in the off position. 3. See the detector head operator’s manual for instructions on how to connect the detector head to a controller. 4. Install an appropriately rated cable bushing or conduit in the right conduit hub on the bottom of the Beacon 110 housing. 5.
be setup for “Pressure” with units of “PSI” and a full scale of “100” PSI. Perform the following procedure to connect a 4 to 20 mA transmitter that you supply to the Beacon 110. 1. Turn off or unplug power to the Beacon 110. 2. Open the Beacon 110 door and turn off the power switch. 3. See the transmitter’s instruction manual for instructions on how to connect wires to the transmitter. 4.
Start Up Introducing Incoming Power Perform the following procedure to place the Beacon 110 into normal operation. 1. Complete the mounting and wiring procedures described earlier in this chapter. 2. Complete all installation procedures described in the detector head or user supplied 4 - 20 mA transmitter operator’s manual. 3. Verify that all wiring connections are correct and secure and that the Beacon 110’s power switch is in the OFF position. 4. Plug in or turn on the incoming power source (AC or DC). 5.
Chapter 4: Operation Overview This chapter describes the Beacon 110 in normal operation. This chapter also describes the Beacon 110 in alarm 1, alarm 2, and fail conditions and suggests response to these conditions. Normal Operation Normal operation is defined as follows: • The start-up procedure is complete. • The Beacon 110 is not indicating an alarm 1, alarm 2, or fail condition. • The Beacon 110 is not in Calibration, Configuration, or Input Mode.
• If the Beacon 110 goes into a fail condition, the signal output is fixed below 2.4 mA until the fail alarm is cleared. Viewing and Resetting Min/Max Readings The reset switch may be used to view and reset the minimum and maximum gas readings. NOTE: Minimum and maximum readings are reset if the instrument is turned off. 1. While the Beacon 110 is in normal operation, press and hold the reset switch button for 5 seconds. 2.
Alarm Indications NOTE: The Beacon 110 includes alarm on and alarm off delay settings for alarm 1 and alarm 2. The alarm indications described in this section operate according to the factory set alarm settings. See Table 7 on page 27 for all the factory settings.
Alarm 1 Condition Alarm 1 Condition Indications When the gas reading reaches the alarm 1 setpoint, the Beacon 110 senses an alarm 1 condition. The Beacon 110 alerts you to an alarm 1 condition as follows: • The alarm 1 LED turns on. • The gas reading alternates with the ALARM-1 message. • The buzzer sounds a pulsing tone. • The alarm 1 relay energizes. • If the optional strobe is installed and set to operate for alarm 1, it flashes. Responding to an Alarm 1 Condition 1.
Responding to an Alarm 2 Condition 1. Follow your established procedure for a high level combustible or toxic gas condition or an increasing oxygen concentration condition. 2. Oxygen alarms are factory set as self-resetting and will automatically clear when the oxygen reading rises above the alarm 2 setpoint. 3. Alarms for all other gas types are factory set as latching. After the gas reading falls below the alarm 2 setpoint, press the reset switch to reset the alarm circuit.
The Beacon 110 senses a DC low power condition when the DC power source is 18.5 volts or less. WARNING: While in a low power condition, the Beacon 110 is not an active gas monitor. When the Beacon 110 senses a low DC power condition, it alerts you as follows: • The fail LED turns on. • The message LowPower is indicated on the top line of the LCD and the input voltage is displayed on the bottom line of the LCD. NOTE: The low DC power alarm cannot be cleared using the reset switch.
Chapter 5: Configuration Mode Overview This chapter describes how to view and change Beacon 110 parameters using Configuration Mode. It is accessed using the program buttons. Configuration Mode includes a 5-minute time-out feature. If you do not press a control button for 5 minutes, the Beacon 110 automatically returns to normal operation. NOTE: If the Beacon 110 returns to normal operation because of a time-out, it enters a warm-up period just as it does when it is first turned on.
Table 7: Configuration Parameters Parameter (Factory Set Value) Description ALARM-1 (level) See the Beacon 110 Detector Head Specification Sheet for the detector head installed The gas reading at which the Beacon 110 initiates an alarm 1 condition. ALARM-1 (activation) (Decrease for oxygen, Increase for all other types) Indicates if the alarm 1 circuit is activated by gas readings increasing (Increase) or decreasing (Decrease) to the ALARM-1 Level. ALARM-1 (relay action) (N. DE-EN) If set as N.
Table 7: Configuration Parameters (Continued) Parameter (Factory Set Value) Description A2Strobe (None if no strobe is installed, Non Reset if a strobe is installed) If set as None, there is no strobe operation. If set as NonReset, it is not possible to turn off the strobe with the reset switch while the Beacon 110 is in an alarm 2 condition. If set as CanReset, the strobe can be turned off while the Beacon 110 is in an alarm 2 condition by pressing the reset switch.
Chapter 6: Input Mode Overview This chapter describes how to use Input Mode to select the Beacon 110’s detector head input type and the gas type. The detector head input type determines whether a direct connect type or a 4 - 20 mA type of detector head will be used and the gas type determines the target gas and detection range. Input Mode includes a 5-minute time-out feature. If you do not press a button for 5 minutes, the Beacon 110 automatically returns to normal operation.
Table 8: Beacon 110 Detector Head Input Types Detector Head Input Type Description OXY DIR An OXY DIR detector head is an oxygen detector head in which the oxygen detector is wired to the Beacon 110 with 2 wires using the OXY GRN and WHT terminals from the detector/ transmitter terminal strips. All calibration adjustments are made at the Beacon 110. TOX DIR A TOX DIR detector head is connected to the Beacon 110 with 2 wires using the AMP + and S terminals from the detector/ transmitter terminal strips.
6. If you changed the input type above, you must select a gas type. Proceed to step 7. If you did not change the input type, you can decide if you want to select the gas type. If you do not want to select the gas type, press and release the DOWN/NO button. The display will indicate NO CHANGE and the Beacon 110 will return to normal operation with the previous setup.
10. While in Configuration Mode press and release the ENTER button to accept the displayed parameter setting and move to the next one. Use the UP/YES and DOWN/ NO buttons to adjust a setting. See “Chapter 5: Configuration Mode” on page 26 for a complete description of Configuration Mode. 11. When you have reviewed and accepted the last item in Configuration Mode by pressing the ENTER button, display will ask SAVE IT? YES/NO. 12. To save the configuration settings, press and release the UP/YES button.
Chapter 7: Maintenance Overview This chapter describes how to calibrate the Beacon 110 using Calibration Mode, replace the AC fuses, and preventive maintenance procedures for the Beacon 110. It includes a troubleshooting guide for problems you may encounter with the Beacon 110. Calibration Frequency Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 3 to 6 months is adequate for most Beacon 110 applications.
When calibration gas is applied to the Beacon 110 during a span (zero for an oxygen detector) adjustment, the Beacon 110 will freeze the display reading at the highest (lowest for an oxygen detector) response to the calibration gas. The calibration gas can then be removed and the Beacon 110 will continue to display this reading and retain it in its memory until the span (zero for an oxygen detector) adjustment procedure is completed.
No Normal Operation Press UP/YES Calibrate? Yes/No Yes Calibrate at Detector Head Press ENTER When Done Figure 13: 4 - 20 mA Transmitter Detector Head Entering Calibration Mode WARNING: The Beacon 110 is not an active gas monitoring device during the calibration procedure. The 4-20 mA output signal will “freeze” at 3.5 mA (17.4 mA if an oxygen detector head is installed) and all relays will remain in their non-alarm state while the Beacon 110 is in Calibration Mode.
Calibrating a 4 - 20 mA Detector Head If a 4-20 mA detector head is installed on the Beacon 110, proceed as follows: 1. The display will be alternating between CAL AT AMP and the current gas reading as described above. 2. Verify that the detector head is in a fresh-air environment. (If necessary, use a zeroemission air cylinder, also known as zero air, to introduce a fresh-air sample when adjusting the fresh air reading below.) 3.
4. If you want to skip adjusting the span (zero for an oxygen detector head) setting, press and release the DOWN/NO button. The display will indicate Leaving Cal Mode and the Beacon 110 will return to normal operation. If you want to continue with adjusting the span setting, press and release the UP/YES button. APPLY will alternate with SPAN Gas (ZERO Gas for an oxygen detector head) on the top display line and the current gas reading will be on the bottom display line. 5.
Replacing the Fuses The Beacon 110 has two replaceable AC fuses. NOTE: To replace other components of the Beacon 110, contact RKI Instruments, Inc. for further information. 1. Turn off or unplug all incoming power to the Beacon 110. 2. Open the housing door of the Beacon 110, then place the power switch in the OFF position. 3. Locate the vertical fuse holders above the AC In Terminal Strip. The AC fuses are labelled F1 and F2 on the main PCB silkscreen. 4.
Troubleshooting The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the Beacon 110. NOTE: This troubleshooting guide describes controller problems only. See the detector head operator’s manual for troubleshooting procedures that apply to the detector head supplied with your Beacon 110. Table 9: Troubleshooting the Beacon 110 Condition No Power Symptom(s) • The display backlight is off and the display screen is blank.
Table 9: Troubleshooting the Beacon 110 (Continued) Condition Reset Switch Not Working Symptom(s) • • If Installed, Option Strobe Not Working • • Probable Causes The buzzer does not silence when you press the reset switch. The alarm 1 or alarm 2 circuit does not reset when you press the reset switch after the alarm condition passes. • The strobe does not flash when an alarm 1 or alarm 2 condition occurs. The strobe does not flash when a fail condition occurs.
Parts List Table 10 lists replacement parts and accessories for the Beacon 110. Table 10: Parts List Part Number Description 18-0107RK Conduit Hub (3/4 in.
Appendix A: Control Button Quick Reference Guide The Beacon 110’s control buttons and reset switch allow access to operational modes, resetting of alarms, and display of the Information Screen. Table 11 shows which button combinations perform these functions and which parameters are available for adjustment while in the operational modes. While in these modes, display prompts showing a “?” require you to respond by pressing either the UP/YES (for yes) or DOWN/NO (for no) button.