Instruction Manual
37 Beacon 110 Rig Monitor Operator’s Manual
Appendix A: Control Button Quick Reference Guide
The Beacon 110 Rig Monitor’s control buttons and reset switch allow access to operational modes,
resetting of alarms, and display of the Information Screen. Table 9 shows which button
combinations perform these functions and which parameters are available for adjustment while in
the operational modes. While in these modes, display prompts showing a “?” require you to
respond by pressing either the UP/YES (for yes) or DOWN/NO (for no) button. To change
parameter settings, use the UP/YES and DOWN/NO buttons to get the desired setting, then press
and release the ENTER button to accept the setting.
Table 9: Control Button Quick Reference Chart
Mode Entered or
Function Performed
Button
Combination
Adjustable or Viewable Parameters
Alarm Reset Press and Release
the Reset Switch
n/a
Min/Max Press and Hold the
Reset Switch for 3
seconds
Minimum and maximum readings can be viewed
and reset.
Show Information Screen Press and Hold
DOWN/NO
• Operating DC Voltage.
• Beacon 110 Rig Monitor Version/Revision
Information
Calibration Mode Press and Hold
UP/YES for 5
seconds
• Fresh Air Adjust
• Span Adjust with Calibration Gas
Configuration Mode Press and Hold
UP/YES & ENTER
for 5 seconds
• Alarm Levels
• Increasing or Decreasing Alarms
• Normally Energized or Normally De-energized
Relays
• Latching or Self-resetting Alarms
• Strobe operation (if installed)
• Alarm On Delay
• Alarm Off Delay (if self-resetting)
• Zero Suppression
• Noise Filter
• Calibration Time-out Period