User Manual
PID Sensor Tab • 24
There are limitations for the full scale and increment values that
depend on the factor for both the high range and the low range. Table 2
and Table 3 below list these limitations.
Table 1: High Range PID
Factor
Full Scale
(ppm)
Increment
0.25-0.49 500 0.2
0.50-0.74 1000 1
0.75-0.99 1500 1
1.00-1.24 2000 1
1.25-1.49 2500 1
1.50-1.99 3000 2
2.00-2.49 4000 2
2.50-4.99 5000 2
5.00-7.49 10000 10
7.50-9.99 15000 10
10.00-14.49 20000 10
15.00-24.49 30000 20
25.00 50000 20
Table 2: Low Range PID
Factor
Full Scale
(ppm)
Increment
0.20-0.29 10.00 0.01
0.30-0.39 15.00 0.01
0.40-0.49 20.00 0.01
0.50-0.59 25.00 0.01
0.60-0.79 30.00 0.02
0.80-0.99 40.00 0.02