User Manual

PID Sensor Tab 24
There are limitations for the full scale and increment values that
depend on the factor for both the high range and the low range. Table 2
and Table 3 below list these limitations.
Table 1: High Range PID
Factor
Full Scale
(ppm)
Increment
0.25-0.49 500 0.2
0.50-0.74 1000 1
0.75-0.99 1500 1
1.00-1.24 2000 1
1.25-1.49 2500 1
1.50-1.99 3000 2
2.00-2.49 4000 2
2.50-4.99 5000 2
5.00-7.49 10000 10
7.50-9.99 15000 10
10.00-14.49 20000 10
15.00-24.49 30000 20
25.00 50000 20
Table 2: Low Range PID
Factor
Full Scale
(ppm)
Increment
0.20-0.29 10.00 0.01
0.30-0.39 15.00 0.01
0.40-0.49 20.00 0.01
0.50-0.59 25.00 0.01
0.60-0.79 30.00 0.02
0.80-0.99 40.00 0.02