4 7 65-2649RK-HC-04 M2A Transmitter Operator’s Manual
Appendix A: Con trol Button Quick Reference Guide
The M2A’s control buttons allow access to operational modes, resetting of alarms, and
display of the Information Screen. Table 14 shows which button combinations perform
these functions and which parameters are available for adjustment while in the
operational modes. While in these modes, display prompts showing a “?” require you to
respond by pressing either the UP/YES (for yes) or DOWN/NO (for no) button. To
change parameter settings, use the UP/YES and DOWN/NO buttons to get the desired
setting, then press and release the ENTER button to a ccept the setting.
Table 14: Control Button Quick Reference Chart
Mode Entered or
Function Performed
Button
Combination
Adjustable or Viewable Parameters
Alar m Reset Press an d
Release ENTER
n/a
Show Information
Screen
Press and Hold
DOWN/NO
•Input Voltage
M2A Version/Revision Information
Calibration Mod e Press and Hold
UP/YES for 5
seconds
Fresh Air Adjust
Span (or Zero fo r Oxygen) Adjus t with
Calibration Gas
Configuration Mode Press and Hold
UP/YES &
ENTER for 5
seconds
Alarm Levels
Increas ing or Dec reasing Alarms
Normally En ergized or Norma lly De-
energized Relays
Latching or Self-resetting Alar ms
Alarm On Delay
Alarm Off Delay (if self-resetting)
Zero Suppression
Noise Filter
Calibration Time-out Period
Gas Type Mode Press and Hold
UP/YES and
DOWN/NO for
5 seconds
Gas Type
Modbus Mode Press and Hold
DOWN/NO and
ENTER for 5
seconds
Enable or Disable Modbus
•Slave ID
•Baud Rate
•Parity
Response Delay