User Manual
4 3 65-2649RK-HC-04 M2A Transmitter Operator’s Manual
Registers 15 - 19
Registers 15 - 19 are the Gas Name ASCII String (NULL Terminated).
Registers 20 - 22
Registers 20 - 22 are the Gas Units String (NULL Terminated).
Register 23
Regist er 23 is the alarm 1 trigger.
0=Decreasing
1=Increasing
Register 24
Regi ster 24 is the alar m 1 normal relay s t ate.
0=Norma lly De-Energized
1=Norma lly Energized
Register 25
Regi st er 25 is the alarm 1 relay rese t.
0=Latching
1=Self-Resetting
Register 26
Regi ster 26 is the alarm 1 ON del ay.
Integer value s are in seconds.
Register 27
Register 27 is the alarm 1 O FF d e lay.
Integer value s are in seconds.
Register 28
Regist er 28 is the alarm 2 trigger.
0=Decreasing
1=Increasing
Register 29
Regi ster 29 is the alar m 2 normal relay s t ate.
0=Norma lly De-Energized
1=Norma lly Energized
Register 30
Regi st er 30 is the alarm 2 relay rese t.
0=Latching
1=Self-Resetting
Register 31
Regi ster 31 is the alarm 2 ON del ay.
Integer value s are in seconds.
Register 32
Register 32 is the alarm 2 O FF d e lay.
Integer value s are in seconds.