User Manual
65-2649RK-HC-04 M2A Tran smitter Op erator’s Manual 42
Register 3
Register 3 is the supply vo ltage (0.1 volt per count).
Registers 4 - 11
The 16 bytes in these registers contain the same ASCII text information that is displayed
on the M2A’s OLED display (not NULL Terminated).
Register 12
Register 12 is the range (full scale readout).
Register 13
Regist er 13 is the alarm 1 se t poin t .
The decimal poin t location is the same a s specif ied in Register 1.
Register 14
Regist er 14 is the alarm 2 se t poin t .
The decimal poin t location is the same a s specif ied in Register 1.
[8] Fail Status
0 = Fail Not Ass erte d
1 = Fail Asserted
[7:6] Alarm 2 Status
0 = No Alarm
1 = Unacknowledged Alarm
2 = Acknowledged Alarm
3 = Un used Code
[5:4] Alarm 1 Status
0 = No Alarm
1 = Unacknowledged Alarm
2 = Acknowledged Alarm
3 = Un used Code
[3] Overscale Status
0 = Not Overscale
1 = Ove rscale
[2] Fail Relay Status
0 = Not Energized
1 = Energized
[1] Alarm 2 Relay Status
0 = Not Energized
1 = Energized
[0] Alarm 1 Relay Status
0 = Not Energized
1 = Energized
Table 11: Register 2, Operating State, Alarms & Relays Bit & Field Assignments (Continued)