4 1 65-2649RK-HC-04 M2A Transmitter Operator’s Manual
Supported Modbus Functions
The M2A supports Function Code 03: Read Holding Registers and Function Code 16:
Write Registers.
Function Code 03: Read Holding Registers
There are a total of 40 registers available to be read via this Function Code.
Registers 1 - 2
The format (bit & field assignm ents) for registers 1 - 2 is shown below in Table 10 -
Table 11.
Table 10: Register 1, Numerical Reading with Sign & Decimal Point Bit & Field Assignments
Bit(s) Value & Field
[15] Sign of Reading
0 = Positive
1 = Ne gative
[14:13] Format of Rea ding
0 = XXX (no decimal point)
1 = XX.X
2 = X.XX
3 = .XXX
[12:0] Numerical value of display reading (8000 max)
Table 11: Register 2, Operating State, Alarms & Relays Bit & Field Assignments
Bit(s) Value
[15:12] Operat i ng State
0 = Un defined
1 = Initialization
2 = Normal
3 = Warm-up
4 = Fail
5 = Calibration
6 = Post-Calibration
7 = Low Power
8 = Set up
[11] Gas Type Change
0=Not Changed
1=Changed
[10] Configuration Change
0=Not Changed
1=Changed
[9] Calibration Activity Flag
0=No Calibratio n A c tivity
1=Calibration Activity Has Occurred