User guide
 33 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Calibration
This section describes how to calibrate the M2 transmitter. It includes procedures to 
prepare for calibration, enter Calibration Mode, adjust the fresh air (zero) setting, adjust 
the span setting, and return to normal operation. 
WARNING: The M2 is not an active gas monitoring device during the calibration 
procedure. The 4-20 mA output signal will “freeze” at 3.5 mA and all 
relays will remain in their non-alarm state while the M2 is in Calibration 
Mode. The output signal will not indicate current readings and the relays 
will not resume operating normally until the M2 is in normal operation 
again.
NOTE: The following procedure assumes the use of a calibration kit which includes a 
calibration gas cylinder, a zero air cylinder and a fixed flow regulator with an 
on/off knob.
NOTE: While in the calibration program, if there is no switch activity for the 
calibration time-out period the unit will return to normal operation. See 
“Viewing & Changing M2 Parameters” on page 23 for instructions to set the 
calibration time-out.
NOTE: The following procedure assumes that the target gas is present in a high 
enough concentration to affect the fresh air (zero) reading. If a fresh air 
environment can be verified, applying zero air to the detector is not necessary 
when setting the zero reading.
Non-Intrusive Calibration
If the M2 is installed in a classified area and non-intrusive calibration is required, follow 
the instructions below, but do not remove the junction box cover. Use the magnetic wand 
accessory to actuate the magnetic switches instead of pushing the control switch buttons. 
To actuate a magnetic switch and perform the same operation as pressing a control switch 
button, touch the magnet in the end of the magnetic wand to the M2’s glass cover directly 
above the magnetic switch you wish to actuate. Touching the glass and removing the 
wand is the same as pressing and releasing a button. Touching the glass and keeping the 
wand in place is the same as pressing and holding a button.
Calibration Gas Response Memory Feature
The M2 has the capability to “remember” the detector’s response to the calibration gas 
after the gas is removed from the detector during the fresh air and span adjustment 
procedure. This feature enables one person to perform calibration if the detector is 
mounted remotely from the M2. When zero air is applied to the M2 during a fresh air 
adjustment, the M2 will freeze the display reading at the lowest response and the M2 will 
continue to display this reading and retain it in its memory until the fresh air adjustment 
procedure is completed. When calibration gas is applied to the M2 during a span 
adjustment, the M2 will freeze the display reading at the highest response to the 
calibration gas. The calibration gas can then be removed and the M2 will continue to 
display this reading and retain it in its memory until the span adjustment procedure is 
completed.










