User Manual

6 • 65-2516 RK Oxygen Transmitt er
10. Connect the wires to the to the applicable controller transmitter terminal strip as
shown in Figure 3.
Figure 3: Wiring the Oxygen Transmitter to a Controller
11 . If shielded cable is used, connect the cable’s drain wire to an available chassis (earth)
ground at the controller. RKI controllers typically have a ground stud that can be used
to ground the cable’s drain wire.
ZERO
SPAN
4/20
FB
+
24V
POWER/SIG
OXY
W
G
SENSOR
TP
-
TP
+
White
Green
Not
Used
OXYGEN
Not
Used
Controller T ransmitt
er
Terminals,Typical
+ 24 VDC
4 - 20 mA In
(FB or S)
Cable Shield