Manual

65-2484RK Multi Point Detector • 23
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the multi-point detector.
NOTE: This troubleshooting guide describes multi-point detector problems only. See the
controller operator’s manual for problems you may encounter with the
controller.
Table 2: Troubleshooting the Multi-Point Detector
Condition Symptom(s) Probable Causes Recommended Action
Fail Condition Controller indicates a
fail condition.
The detector wiring is
disconnected or
misconnected.
•The O
2
, H
2
S, CO, or
ESM-01 plug-in sensor
is not properly plugged
into the sockets in the
detector housing body.
The detector’s zero
reading is low enough
to cause a fail
condition.
The detector is
malfunctioning.
1. Verify that the detector wiring is
correct and secure.
2. Confirm that the O
2
, H
2
S, CO, or
ESM-01 plug-in sensor is installed
properly.
3. Perform a zero adjustment. A full
calibration is recommended.
4. If the fail condition continues, replace
the O
2
, H
2
S, CO, or ESM-01 plug-in
sensor as described later in this
section.
5. If the fail condition continues, contact
RKI for further instruction.
Slow or No
Response/
Difficult or
Unable to
Calibrate
Detector responds
slowly or does not
respond to response
test.
Unable to accurately
set the zero or
response reading
during calibration.
Detector requires
frequent calibration.
Note: Under “normal
circumstances, the
transmitter requires
calibration once every 3
months.
Some applications
may require a more
frequent calibration
schedule.
The calibration cylinder
is low, out-dated, or
defective.
The flame arrestor on
the LEL detector or in
the O
2
, H
2
S, or CO
detector housing cap is
wet or clogged with dirt
or other particulates.
The calibration gas is
not an appropriate
concentration.
The detector is
malfunctioning.
1. Verify that the calibration cylinder
contains an adequate supply of a
fresh test sample.
2. Check the LEL detector or detector
housing cap to determine if the flame
arrestor is wet or dirty. Check the
sensor face of the ESM-01 plug-in
sensor to determine if it is wet or dirty.
Clean if necessary.
3. Verify that the calibration gas
concentration is appropriate for each
detector.
4. If the calibration/response difficulties
continue, replace the LEL detector or
the O
2
, H
2
S, CO, or ESM-01 plug-in
sensor as described later in this
section.
5. If the calibration/response difficulties
continue, contact RKI for further
instruction.