Instruction Manual
8 • 65-2428 H
2
S Detector
Quarterly
Calibrate the H
2
S detector as described in “Calibration” on page 11.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended 
action for problems you may encounter with the H
2
S detector.
NOTE: This troubleshooting guide describes detector problems only. See the controller 
operator’s manual for problems you may encounter with the controller.
Fail Condition
Symptoms
• The controller indicates a fail condition.
Probable causes
• The detector wiring is disconnected or misconnected.
• The detector’s zero reading is low enough to cause a fail condition.
• The detector is malfunctioning.
Recommended action
• Verify that the detector wiring is correct and secure.
• Calibrate the detector.
• If the fail condition continues, replace the H
2
S sensor as described later in this section.
• If the fail condition continues, contact RKI for further instruction.
Slow or No Response/Difficult or Unable to Calibrate
Symptoms
• The detector responds slowly or does not respond during the monthly response test.
• Unable to accurately set the zero or response reading during the calibration 
procedure.
• The detector requires frequent calibration.
NOTE: Under “normal” circumstances, the detector requires calibration once every three 
months. Some applications may require a more frequent calibration schedule.
Probable causes
• The calibration cylinder is low, out-dated, or defective.
• The detector is malfunctioning.
Recommended action
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. If the calibration/response difficulties continue, replace the H
2
S sensor as described 
later in this section.
3. If the calibration/response difficulties continue, contact RKI Instruments, Inc. for 
further instruction.










