Owner's manual

12 • 65-2322RKSS Oxygen Transmitter
11. Connect the wires to the applicable detector/transmitter terminal strip at the controller as
shown in Figure 4.
Figure 4: Wiring the Oxygen Transmitter to a Controller
12. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth)
ground at the controller. RKI controllers typically have a ground stud that can be used to
ground the cable’s drain wire.
J-Box
Install conduit seal within 18
inches (for environments
without acetic acid) or 2
inches (for environments with
acetic acid)
SIG/PWR
S
SPAN
ZERO
TO XI C OXY
Controller Detector/Transmitter
Terminals, Typical Designations
Green
White
Control ler Housing
(installed in ordinary location is
dev ice is non-explosion-proof)
Cable Shield
S
+
Oxygen Detector