User guide

65-2300RK/65-2301RK Toxic Detector 9
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the toxic detector.
NOTE: This troubleshooting guide describes detector problems only. See the controller
operators manual for problems you may encounter with the controller.
Fail Condition
Symptoms
The controller indicates a fail condition.
Probable causes
The detector wiring is disconnected or misconnected.
The plug-in sensor is not properly plugged into the socket in the detector housing
body.
The detector’s zero reading is low enough to cause a fail condition.
The detector is malfunctioning.
Recommended action
Verify that the detector wiring is correct and secure.
Confirm that the plug-in sensor is installed properly.
Perform a zero (fresh air) adjustment. A full calibration is recommended.
If the fail condition continues, replace the plug-in sensor as described later in this
section.
If the fail condition continues, contact RKI for further instruction.
Slow or No Response/Difficult or Unable to Calibrate
Symptoms
The detector responds slowly or does not respond during the monthly response test.
Unable to accurately set the zero or response reading during the calibration
procedure.
The detector requires frequent calibration.
NOTE: Under “normal” circumstances, the detector requires calibration once every three
months. Some applications may require a more frequent calibration schedule.
See “Calibration Frequency” on page 11 for a discussion of the calibration
frequency.
Probable causes
The calibration cylinder is low, out-dated, or defective.
The incorrect calibration cup or regulator is being used.
The plug-in sensor face is blocked with dirt or some other particulate contamination.
The detector is malfunctioning.