User Manual

22 35-3010RKA-02 Sample-Draw Detector
Preparing for Calibration
CAUTION: This proce dure may cause alarms at the controller. Take appropriate action to avoid
this, such as entering the calibration mode at the controller.
1. Open the housing door.
2. Set a voltmeter to measure in the millivolt (mV) range.
3. For the oxyg en channel, plug the voltmeter into the test points on the oxygen
transmitter. Plug the positive lead into the test point labeled TP+; plug the negative
lead into the test point labeled TP-.
4. Use the following formula to determine the correct test points output for the
calibrating sample.
Output (mV) = (calibratin g sample/fullscale) X 400 + 100
For example, with a calibrating sample of 12% and a fullscale setting of 25%, the
correct output for the oxygen test points is 292 mV.
292 (mV) = (12/25) X 400 +100
Setting the Zero Reading
NOTE: The 4-gas calibration cylinder included in the calibration kit may be used to zero
the oxygen channel. Keep in mind that the oxygen concentration in that cylinder
is 12%, not 0%. Be sure t o set the controller appropriately.
1. Screw the regulator into the 4-gas calibration cylin der.
2. Connect the calibration kit sa mple tubing to the regulator.
3. Allow the samp le-draw detector to respond to the calibrating sample for
approximately 1 minute.
4. After one minute, check the mV output on the oxygen transmitter test po ints and
verify that the reading matches the response reading (±2 mV) you determined earlier
(292 mV for 12% oxygen).
5. If necessary, use the span pot on the tra nsmitter to adjust the reading to match the
correct response reading.
6. Disconnect th e sample tubing from the inlet line.
7. Unscrew the regulator from the 4-gas ca libration cylind er.
8. For convenience, lea v e the regulator attached to the sample tubing.
Setting the Fresh Air Reading
CAUTION: This proce dure may cause alarms at the controller. Take appropriate action to avoid
this, such as entering the calibration mode at the controller.
NOTE: If you can verify a fresh air environment, it is not necessary to use the zero air
calibration cylinder to set the zero reading.
1. Screw the regulator into the zero air calibration cylinder.
2. Connect the calibration kit sa mple tubing to the regulator.