User guide
35-3001-04  Carbon Monoxide Sample-Draw Detector • 17
4. Reinstall the tubing routed to the sampling area.
Adjusting the Low Flow Setting
NOTE: Adjusting the low flow setting will cause a low flow alarm at the sample-draw detector 
and a fail alarm at the controller. Be sure to put the controller into its calibration program 
or disable external alarms before performing this test.
The factory-set low flow setting is 0.6 SCFH (±0.1). To adjust the low flow setting:
1. Use the sensor flow control valve to set the flow to 0.6 SCFH. Turn the valve’s knob 
counterclockwise to increase the flow and clockwise to decrease the flow.
If the sample-draw detector goes into low flow alarm before you can adjust the flow down to 
0.6 SCFH, adjust the pressure switch adjustment screw 1/4 turn clockwise, then attempt to set 
the flow again. Repeat this step until you are able to adjust the flow to 0.6 SCFH.
NOTE: The pressure switch adjustment screw is accessible through a circular cutout in the 
flowmeter circuit board. 
2. Slowly turn the pressure switch adjustment screw counterclockwise just until the sample-draw 
detector goes into low flow alarm.
3. Turn the sensor flow control valve’s knob counterclockwise to increase the flow until the unit 
is out of low flow alarm.
4. Decrease the flow very slowly by turning the sensor flow control valve’s knob clockwise and 
verify that the low flow alarm is 0.6 SCFH (±0.1). 
If the low flow alarm is set too low, turn the pressure switch adjustment screw slightly 
clockwise. Repeat steps 3 and 4 if necessary.
5. Use the sensor flow control valve to set the flow to 1.0 SCFH.
6. Make sure the sample-draw detector’s Fail LED is off.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a calibration 
frequency of every 3 months is adequate for most carbon monoxide sample draw detector 
applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. 
recommends a calibration frequency of every 3 months for the carbon monoxide sample draw 
detector. 
If an application is not very demanding, for example detection in a clean, temperature controlled 
environment, and calibration adjustments are minimal at calibration, then a calibration frequency of 
every 6 months is adequate. 
If an application is very demanding, for example if the environment is not well controlled, then 
more frequent calibration than every 3 months may be necessary.










