Owner's manual

Sample Draw Combustible Gas Detector • 17
5. Connect the tubing from the gas collection bag to the xed ow regulator, then open
the clamp.
6. Screw the xed ow regulator onto the zero air calibration cylinder. The gas collection
bag begins to ll.
7. When the bag is full, unscrew the xed ow regulator from the cylinder.
8. Close the clamp, then disconnect the sample tubing from the xed ow regulator.
Setting the Zero Reading
1. Open the clamp, then connect the sample tubing from the gas collection bag to the
sample draw detector’s inlet line. This step is not necessary if you veried a fresh air
environment earlier in this procedure.
2. Allow the reading to stabilize for approximately 1 minute.
3. Verify a voltmeter reading of 100 mV (± 2 mV).
4. If necessary, use a small at-blade screwdriver to adjust the zero potentiometer until
the voltmeter reading is 100 mV (± 2 mV).
Steps 5 and 6 are not necessary if you veried a fresh air environment earlier in this
procedure.
5. Allow the sample draw pump to draw out any residual gas in the gas collection bag.
6. Disconnect the sample tubing from the inlet line, then close the clamp.
Setting the Response Reading
1. Connect the sample tubing from the gas collection bag to the xed ow regulator,
then open the clamp.
2. Screw the xed ow regulator onto the calibration gas cylinder. The gas collection bag
begins to ll.
3. Unscrew the xed ow regulator from the cylinder when the gas collection bag
appears full.
4. Close the clamp, then disconnect the sample tubing from the xed ow regulator.
5. Open the clamp, then connect the sample tubing from the gas collection bag to the
inlet line at or near the sample draw detector’s INLET tting.
6. Allow the sample draw detector to respond to the calibrating sample for
approximately 1 minute.
7. When the reading on the voltmeter stabilizes, verify that the reading matches the
response reading (±2 mV) you determined earlier.
8. If necessary, use the span potentiometer on the amplier to adjust the reading to
match the correct response reading.
9. Allow the sample draw pump to draw out any residual gas in the gas collection bag.
10. Disconnect the sample tubing from the sample draw detector’s inlet line, then close
the clamp.
11. Reconnect the incoming sample line.
12. Wait 1 to 2 minutes to allow the combustible gas reading to decrease and stabilize.