12 • Sample-Draw Hydrogen Sulfide Gas Detection
Recommended action
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. If necessary, set the correct ow rate with the bypass valve or ow adjust
potentiometer.
3. If you cannot set the correct ow rate, check the sample line for obstructions or kinks.
4. If the calibration/response difculties continue, replace the sensor as described later
in this section.
5. If the calibration/response difculties continue, contact RKI Instruments, Inc., for
further instruction.
Replacing Components of the Hydrogen Sulfide Gas
Sample-draw Detector
This section includes procedures to replace the sensor, lter, and ferrules.
Replacing the hydrogen sulfide gas sensor
1. Turn off incomming power.
2. Open the housing door of the sample-draw detector.
3. Unplug the sensor connector from the main circuit board.
4. Grasp the sensor rmly and pull it out of the ow chamber.
5. Install the new sensor in the ow chamber. Make sure it is pushed all the way down.
6. Connect the sensor connector to the main circuit board.
7. Turn on incomming power.
CAUTION:
Allow the replacement sensor to warm up for 15 minutes before you continue.
8. Calibrate the replacement sensor as described in the Calibration section on page 14.
Replacing the filter
1. Open the housing door of the sample-draw detector.
2. Note the direction of the arrow on the lter. The arrow indicates the direction of the
sample ow.
3. Disconnect the lter from the elbows on each end of the lter, then remove the lter
from the sample-draw detector.
4. Make sure the arrow is pointing in the same direction as the arrow on the lter you
removed, then connect each end of the replacement lter to the elbows.
5. Verify that the ow rate is approximately 1.5 SCFH, then close the housing door.
Replacing the ferrules
The INLET and EXHAUST ttings each includes two ferrules that seal the incoming or
exhaust tubing to the tting. Replace the ferrules if the seal is bad or if you replace the
sample tubing. Always replace the ferrules as a pair.
1. Disconnect the sample tubing from the tting, then unscrew the nut from the tting.
2. Verify that the ferrules did not remain in the nut. If necessary, remove the ferrules
from the nut.
3. Position the nut so the threaded end is facing you, then insert the back (smaller)