User guide

35-3000RK-OXY Sample-Draw Oxygen Detector • 9
Installation
This section describes procedures to mount the sample-draw oxygen detector in the
monitoring environment and wire the sample-draw detector to a controller.
Mounting the Sample-Draw Detector
1. Select the mounting site. Consider the following when you select the mounting site.
Is there enough room to open the housing door and make wiring connections at
the bottom of the housing and tubing connections at the right of the housing?
Make sure there is sufcient room to perform start-up, maintenance, and
calibration procedures.
Are the owmeter and status lights visible?
Figure 3: Mounting the Sample-Draw Oxygen Detector
6.5 in.
4.0 in.
Note: The housing is 4.25 in. deep.
8.5
in.
9.3
in.
8.9
in.
3/4 in. conduit hub
(total of 2 hubs)
7.3 in.