User guide

16 • 35-3000RK-OXY Sample-Draw Oxygen Detector
Replacing the ferrules
Both the inlet and exhaust ttings include two ferrules that seal the inlet or exhaust tubing
to the tting. Replace the ferrules if the seal is bad or if you replace the sample tubing.
Always replace the ferrules as a pair.
1. Disconnect the sample tubing from the tting, then unscrew the nut from the tting.
2. Verify that the ferrules did not remain in the nut. If necessary, remove the ferrules
from the nut.
3. Position the nut so the threaded end is facing you, then insert the bottom (smaller)
ferrule into the nut. Insert the ferrule so the at side is facing down.
NOTE: Make sure the bottom ferrule is laying at in the nut.
4. Insert the cone-shaped front ferrule on top of the bottom ferrule. Insert the ferrule so
the smaller end of the cone is facing up.
5. Screw the nut onto the tting, then connect the sample tubing to the tting. Make sure
you rmly tighten the tubing to the tting.
Adjusting the Low Flow Setting
NOTE: Adjusting the low ow setting will cause a low ow alarm at the sample-draw
detector and a fail alarm at the controller. Be sure to put the controller into its
calibration program or disable external alarms before performing this test
The factory-set low ow setting is 0.6 SCFH (±0.2). To adjust the low ow setting:
1. Use the ow adjust potentiometer (VR1) to set the ow to 0.6 SCFH.
If the sample-draw detector goes into low ow alarm before you can adjust the ow
down to 0.6 SCFH, adjust the low ow potentiometer 1/4 turn clockwise, then
attempt to set the ow again. Repeat this step until you are able to adjust the ow to
0.6 SCFH.
2. Slowly turn the low ow potentiometer counterclockwise just until the sample-draw
detector goes into low ow alarm.
NOTE: The low ow potentiometer is accessible through a circular cutout in the main
circuit board. The cutout is labeled PS1.
3. Increase the ow using VR1 until the unit is out of low ow alarm.
4. Decrease the ow very slowly and verify that the low ow alarm is 0.6 SCFH (±0.2).
If the low ow alarm is set too low, turn the low ow potentiometer slightly
clockwise. Repeat steps 3 and 4 if necessary.
5. Use the ow adjust potentiometer (VR1) to set the ow to 1.5 SCFH.
6. Make sure the sample draw detectors Fail light is off.