User Manual

6 • 35-3000RK-LEL Combustible Gas Sample-Draw Detector
Flow System
The sample-draw detector’s ow system consists of the INLET tting, hydrophobic lter,
pump, owmeter, bypass valve, status lights, pressure switch, and EXHAUST tting (see
Figure 1). Figure 2 illustrates how the gas sample moves through the ow system.
Figure 2: Combustible Gas Sample-draw Detector Flow Diagram
INLET fitting
The INLET tting on the right side of the housing allows the gas sample to enter the
sample-draw detector. The INLET tting accepts 1/4 in. rigid tubing. See the Installation
section on page 9 to connect tubing to the INLET tting.
Hydrophobic Filter
The hydrophobic lter is below the main circuit board. The lter prevents particulates and
water in the incoming gas sample from damaging the ow and detection systems. Replace
the lter when it appears dirty, discolored, or clogged.
Pump
The pump is behind the main circuit board near the top of the sample-draw detector.
The pump pulls the gas sample into the sample-draw detector. The pump operates on
24 VAC, which is generated from the 24 VDC supplied by the controller.
Flowmeter
The owmeter is attached to the main circuit board near the top left corner (see Figure 1.)
You can see it through the window in the door. A ball in the owmeter column indicates
the ow rate of the sample-draw detector. The owmeter measures the ow in the range
0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The optimum ow rate is 1.5 SCFH.
Bypass valve
The bypass valve is to the left of the owmeter. The bypass valve adjusts the ow rate to
the detector. Use a at-blade screwdriver to adjust the bypass valve.
NOTE:
The bypass valve allows ne adjustments of the ow rate. For a wider range of
adjustment, use the ow adjust potentiometer (see Figure 1.)
Inlet Filter Pump Sensor
Flowmeter
Pressure Switch
Exhaust
To
Valve
Bypass