Specifications

2. DETAILS OF PARAMETER SETTING
IMR01H03-E4
31
2.11 F91 Displayed for Maintenance Information
Only displayed for the F91 function block.
(1) ROM version displayed (E261)
Display the version of loading software.
(2) Integrated operating time display (WTH) [upper 3 digits]
Display the integrated operating time (upper 3 digits) of instrument.
Details of display : 0 to 100 (Resolution of display: 10,000 hours)
Up to 1,000,000 from 0 including the upper and lower digits can be displayed.
(3) Integrated operating time display (WTL) [lower 4 digits]
Display the integrated operating time (lower 4 digits) of instrument.
However, as the integral time is increments by 1 when the power is turned on or off.
If the total integrated operating time exceeds 9,999 hours, these digits move to the integrated
operating time display (upper 3 digits).
Details of display : 0 to 9999 (Resolution of display: 1 hours)
(4) Holding peak value ambient temperature display (TCJ)
The maximum ambient temperature on the rear terminal board of the instrument is stored and
displayed on the set value (SV) display.
Details of display : 10 to +100 °C [°F] (Resolution of display: 1°C [°F])