Instruction manual
8. SPECIFICATIONS
IMSRM15-E6
172
8.10 Common Specifications
Control unit
Power supply voltage: 90 to 132 V AC (50/60 Hz)
[Including power supply voltage variation]
(Rating: 100 to 120 V AC)
180 to 264 V AC (50/60 Hz)
[Including power supply voltage variation]
(Rating: 200 to 240 V AC)
21.6 to 26.4 V DC
[Including power supply voltage variation]
(Rating: 24 V DC)
Insulation resistance: Between power and ground terminals:
20 M
or more at 500 V DC
Between input/output and ground terminals:
20 M
or more at 500 V DC
Withstand voltage: Between power and ground terminals:
1500 V AC for 1 minute
Between input/output and ground terminals:
1000 V AC for 1 minute
Withstand noise: 1500 V (peak to peak)
Pulse width: 1
s
Rise time: 1 ns
By noise simulator
Withstand vibration: Frequency: 5 to 9 Hz
Amplitude: 1.5 mm
Frequency: 9 to 150 Hz
Acceleration: 5.0 m/s
Sweep speed: 10 Hz/min
Vibration director: Front and back, Right and left, Up and down
(Three directions)
Vibration time: 1 hour, all directions
Power failure effect: No influence even under power failure of 20 ms or less.
Ambient temperature range: 0 to 50
C
Ambient humidity range: 45 to 85 %RH
(Absolute humidity: MAX.W.C 29.3 g/m
3
dry air at 101.3 kPa)
Operating environment: No corrosive gases, no large amounts of dust or particulates.
Storage temperature range: -20 to +70
C
Storage humidity range: 95 %RH or less (Non condensing)
Grounding resistance: 100
or less
Cooling method: Natural cooling