User manual
Appendix E: Performance Verification
E–14
TDS3000 Series User Manual
11. Move the output cable of the leveled sine wave generator to the
oscilloscope channel you want to check.
12. Push the channel button (CH 1, CH 2, CH 3, or CH4) for the
channel you want to check.
13. Push the vertical MENU button.
14. Push the Coupling screen button and select 50 input resistance.
15. Push the MEASURE menu button.
16. Push the Select Measurement screen button.
17. Push the –more– screen button until you can select the Pk–Pk
measurement.
18. Set the vertical SCALE to 500 mV/div.
19. Set the output frequency of the leveled sine wave generator to the
frequency shown in the table below.
Oscilloscope model Frequency
TDS3012, TDS3014 100 MHz
TDS3032, TDS3034 300 MHz
TDS3052, TDS3054 500 MHz