Appendix E: Performance Verification
E–12
TDS3000 Series User Manual
24. Verify that the peak-to-peak measurement is ≥425 mV.
25. Push the waveform off button.
26. Repeat steps 11 through 25 for each channel of the oscilloscope
(not including the external trigger input on some models).