RSA5000 Series Real-time Spectrum Analyzer Ultra-Real technology Frequency: up to 6.5 GHz Displayed average noise level (DANL): <-165 dBm (typical) Phase noise: <-108 dBc/Hz (typical) Level measurement uncertainty: <0.8 dB 6.5 GHz tracking generator Min. RBW 1 Hz EMC filter and quasi-peak detector Various measurement functions Multiple measurement modes Up to 40 MHz real-time analysis bandwidth Multiple trigger modes and trigger masks Density, Spectrogram, and other display modes PC software options 10.
RSA5000 Series Real-time Spectrum Analyzer Product Dimensions: Width × Height × Depth = 410 mm × 224 mm × 135 mm Based on the Ultra-Real technology, the high-speed real-time measurement mode allows you to acquire the signals in the analysis bandwidth seamlessly and make data analysis. It also provides various display modes, such as Spectrogram, Density, and PVT. Besides, FMT function is also available.
Specifications Specifications are valid under the following conditions: the instrument is within the calibration period, is stored for at least two hours at 0℃ to 50℃ temperature, and is warmed up for 40 minutes. Unless otherwise noted, the specifications in this manual include the measurement uncertainty. Typical: characteristic performance, which 80 percent of the measurement results will meet at room temperature (approximately 25℃).
GPSA Mode Frequency Frequency Readout Accuracy Marker Frequency Resolution span/(number of sweep points - 1) Marker Frequency Uncertainty ±(marker frequency readout × reference frequency accuracy + 1% × span + 10% × resolution bandwidth + marker frequency resolution) Frequency Counter Resolution 1 Hz Uncertainty ±(marker frequency readout × reference frequency accuracy + counter resolution) Frequency Span Range 0 Hz, 10 Hz to maximum frequency Resolution 2 Hz Uncertainty ±span/(number of sweep
Amplitude Measurement Range fc ≥ 10 MHz Range DANL to +30 dBm Maximum Safe Input Level[1] DC Voltage 50 V +30 dBm, attenuation ≥ 40 dB, preamp off. CW RF Power -10 dBm, attenuation = 20 dB, preamp on.
Measured Displayed Average Noise Level of RSA5065/RSA5065-TG Measured Displayed Average Noise Level of RSA5032/RSA5032-TG Level Display Logarithmic Scale 1 dB to 200 dB Linear Scale 0 to reference level Number of Display Points 801 Number of Traces 6 Trace Detector normal, pos-peak, neg-peak, sample, RMS average, voltage average, and quasi-peak Trace Function clear write, max hold, min hold, average, view, blank Scale Unit dBm, dBmV, dBμV, nV, μV, mV, V, nW, μW, mW, W Frequency Response RSA5
Input Attenuation Switching Uncertainty Setting Range 0 dB to 50 dB, in 1 dB step Switching Uncertainty fc = 50 MHz, relative to 10 dB, preamp off, 20℃ to 30℃ <0.3 dB Measured attenuation switching uncertainty VS temperature of RSA5032/RSA5032-TG/RSA5065/RSA5065-TG Measured attenuation switching uncertainty of RSA5032/RSA5032-TG/RSA5065/RSA5065-TG Absolute Amplitude Accuracy fc = 50 MHz, peak detector, preamp off, attenuation = 10 dB, input signal level = -10 dBm, 20℃ to 30℃ Uncertainty <0.
RF Input VSWR RSA5032 RSA5032-TG RSA5065 RSA5065-TG attenuation ≥10 dB, preamp off VSWR 300 kHz to 3.2 GHz <1.6 (nominal) <1.6 (nominal) 3.2 GHz to 6.5 GHz <1.8 (nominal) Measured VSWR of RSA5032/RSA5032-TG Measured VSWR of RSA5065/RSA5065-TG Distortion Second Harmonic Intercept (SHI) Third-order Intercept (TOI) fc ≥ 50 MHz, input signal level = -20 dBm, attenuation = 0 dB, preamp off. +45 dBm fc ≥ 50 MHz, two -20 dBm tones at input mixer spaced by 200 kHz, attenuation = 0 dB, preamp off.
Sweep Sweep Sweep Time Sweep Time Uncertainty span ≥ 10 Hz 1 ms to 4,000 s zero span 1 μs to 6,000 s span ≥ 10 Hz, RBW ≥ 1 kHz 5% (nominal) zero span (sweep time > 1 ms) 5% (nominal) Sweep Mode continue, single Trigger Trigger Trigger Source Trigger Delay free run, external 1, external 2, video span ≥ 10 Hz 0 to 500 ms zero span 0 to 500 ms Tracking Generator Tracking Generator Output RSA5032 RSA5032-TG RSA5065 RSA5065-TG Frequency Range - 100 kHz to 3.2 GHz - 100 kHz to 6.
RTSA Mode Real-time Analysis Bandwidth 25 MHz 40 MHz (Option RSA5000-B40) Min. Signal Duration for 100% POI at the Full-Scale Accuracy maximum span, default Kaiser window Trace Detector pos-peak, neg-peak, sample, average Number of Traces 6 Window Type Hanning, Blackman-Harris, Rectangular, Flattop, Kaiser, and Gaussian 7.45 μs provides 6 RBWs for each window, except the Rectangular; for Kaiser window Resolution Bandwidth Span Min. bandwidth Max. bandwidth 40 MHz 100 kHz 3.
General Specifications Display Type capacitive multi-touch screen Resolution 1024 × 600 pixels Size 10.1'' Color 24-bit color Printer Supported Protocol network printer Mass Memory Mass Memory Internal Storage 512 MB (nominal) External Storage USB storage device (not supplied) Power Input Voltage Range, AC 100 V to 240 V (nominal) AC Frequency 45 Hz to 440 Hz Power Consumption 55 W (typical), max.
Input/Output Front Panel Connector RF Input TG Output Impedance 50 Ω (nominal) Connector N-type female Impedance 50 Ω (nominal) Connector N-type female Frequency 10 MHz Output Level +3 dBm to +10 dBm, +7 dBm (typical) Impedance 50 Ω (nominal) Connector BNC female Frequency 10 MHz ± 5 ppm Input Level 0 dBm to +10 dBm Impedance 50 Ω (nominal) Connector BNC female Impedance ≥1 kΩ (nominal) Connector BNC female Level 5 V TTL level Internal/External Reference Internal Reference E
Order Information Model Standard Accessories Option Optional Accessories Description Order No . Real-time Spectrum Analyzer, 9 kHz to 3.2 GHz RSA5032 Real-time Spectrum Analyzer, 9 kHz to 6.5 GHz RSA5065 Real-time Spectrum Analyzer, 9 kHz to 3.2 GHz (with TG installed when leaving the factory) RSA5032-TG Real-time Spectrum Analyzer, 9 kHz to 6.
HEADQUARTER EUROPE NORTH AMERICA JAPAN RIGOL TECHNOLOGIES, INC. No.156,Cai He Village, Sha He Town, Chang Ping District, Beijing, 102206 P.R.China Tel:+86-10-80706688 Fax:+86-10-80720067 Electronic Measurement Instrument service and support email:EMD_support@rigol.com RIGOL TECHNOLOGIES EU GmbH Lindbergh str. 4 82178 Puchheim Germany Tel: 0049- 89/89418950 Email: info-europe@rigol.com RIGOL TECHNOLOGIES, USA INC. 8140 SW Nimbus Ave. Beaverton, OR 97008 Tel: 877-4-RIGOL-1 Email: info@rigol.