Fuse Replacement 
If you need to replace the fuse, use only the specified fuse (AC 250V, T3.15A) and 
perform the following operations: 
1.  Turn off the instrument, cut off the power, and remove the power cord. 
2.  Use a small straight slotted screwdriver to pry out the fuse holder. 
3.  Take out the fuse holder. 
4.  Replace the old fuse with a specified fuse. 
5.  Install the fuse holder. 
Figure 1-15 To Replace the Fuse 
WARNING 
To avoid electric shock, please ensure that the instrument has been 
turned off, the power source has been cut off, and the fuse to be used 
conforms to the fuse rating. 
Fuse Holder 
Fuse 