Instructions
Chapter 5 Functions of the Front Panel of VSA  RIGOL 
RSA3000E User Guide  5-13 
Trigger 
Selects the trigger source and sets trigger-related parameters. 
Trigger Source 
Sets "Free Run", "External 1", "External 2", "Power (Time)", or "FMT" to be the 
trigger source. 
Free Run 
The trigger conditions are met at any time, that is, the analyzer generates trigger 
signals continuously. 
External 1 
In this mode, an external signal is input via the [TRIGGER IN] connector on the 
rear panel. When the signal meets the set trigger conditions, trigger signals are 
generated. 
1.  Slope 
Sets the trigger polarity for External Trigger 1. It is set positive to trigger on a 
rising edge and negative to trigger on a falling edge. 
2.  Delay State 
Enables or disables the trigger delay function. After the trigger delay function is 
enabled, you can set the trigger delay time. 
3.  Delay Time 
Sets the time interval during which the instrument waits to start the sweep 
operation after the trigger signal that meets the trigger conditions is generated. 
You can use the numeric keys, the knob, or arrow keys to modify this parameter; 
you can also use the touch screen to modify the parameter. 
Table 5-13 Trigger Delay Time 
Parameter 
Remarks 
Default 
1 μs 
Range 
-16.384 s
[1]
 to 500 ms 
Unit 
s, ms, μs, ns, ps 
Knob Step 
trigger delay/100, Min = 1 μs 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
at 1-1.5-2-3-5-7.5 step 
Note
[1]
: A negative trigger delay time indicates a pre-trigger. 